ATTUNE 2545-00-979Recalls & reported problems

Depuy (Ireland) · ATTUNE FB PS ARTICULATION SURFACE SIZE 9 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10603295135562), received between Jun 7, 2016 and Sep 2, 2025 — classified as 19 malfunctions.

The most frequently reported specific device problem is “Break” (15 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 12, 2015Class IITerminated

    Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

    Z-2142-2015 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break15
  • Material Deformation3
  • Detachment of Device or Device Component2
  • Crack1
  • Scratched Material1

Patient problems

  • No Consequences Or Impact To Patient10
  • No Clinical Signs, Symptoms or Conditions6
  • No Information2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2025Material Deformation; Scratched Material; Material Deformation; Scratched Material

    DATE J&J BECAME AWARE: 14.7.25. DATE OF EVENT: 12.7.24. PRODUCT NAME: ATTUNE ARTICULATE SURFACE SIZE 9 FB . PRODUCT CODE: 254500979. LOT NUMBER: BFA0V5Y. DESCRIPTION OF DAMAGE SPRING IS LOOSE AND DAMAGE PLASTIC. EVENT OCCURRED DURING A PROCEDURE. NO SURGICAL DELAY. NO PATIENT CONSEQUENCE OR ADVERSE OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22936227 ↗
  • MalfunctionJul 17, 2023Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT THERE ARE SEVERAL ARTICULATING SURFACE COMPONENTS THAT HAVE PRONGS BENT, OR WARPED, AND SPRINGS MISSING. THIS DID NOT CAUSE A DELAY, AND NO ONE WAS INJURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17334335 ↗
  • MalfunctionJul 17, 2023Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT THERE ARE SEVERAL ARTICULATING SURFACE COMPONENTS THAT HAVE PRONGS BENT, OR WARPED, AND SPRINGS MISSING. THIS DID NOT CAUSE A DELAY, AND NO ONE WAS INJURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17334307 ↗
  • MalfunctionFeb 22, 2023Break; Break

    IT WAS REPORTED THAT THE SURGEON HIT TRIAL WITH A MALLET AND THE SIDE OF SURFACE BROKE OFF. ALL PARTS RETRIEVED. IT WAS UNKNOWN ABOUT THE SURGERY DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16419188 ↗
  • MalfunctionApr 1, 2022Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WAS BROKEN OR NOT IN USABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13977020 ↗
  • MalfunctionDec 23, 2021Break; Break

    BOTTOM PEG IS SHEERED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13076817 ↗
  • MalfunctionSep 21, 2020Break; Break

    IT WAS REPORTED THAT THE INSERT TRIAL WAS BROKEN DURING PROCEDURE. ALL PIECES GATHERED. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10563585 ↗
  • MalfunctionJul 16, 2020Break; Break

    IT WAS REPORTED THAT WHILE REMOVING THE INSERT TRIAL, THE PLASTIC INSTRUMENT SNAPPED IN TWO. ALL PARTS WERE RETRIEVED. NOTHING LEFT IN PATIENT. THE INSERT TRIAL INSTRUMENT WAS THROWN IN TRASH. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10281254 ↗
  • MalfunctionDec 28, 2018Crack; Crack

    IT WAS REPORTED THAT THE SIZE 9 FB PS INSERT ARTICULATE SURFACE WAS BEING TRIALED AND WHEN TAKEN, IT CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8202439 ↗
  • MalfunctionDec 26, 2018Break; Break

    INSERTS BROKE DURING TRIAL REDUCTION. PATIENT CONSEQUENCE?: NO. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT?: YES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8196832 ↗
  • MalfunctionNov 2, 2018Break; Break

    IT WAS REPORTED THAT THE PEG ON ATTUNE TRIALS BROKE DURING INSERTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8033608 ↗
  • MalfunctionNov 2, 2018Break; Break

    IT WAS REPORTED THAT THE PEG ON ATTUNE TRIALS BROKE DURING INSERTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8033557 ↗
  • MalfunctionJun 11, 2018Break; Break

    IT WAS REPORTED THAT THE POLY TRIAL BROKE AT INSERTION WHILE CEMENTING. NO SURGICAL DELAY. ALL PIECES RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7586482 ↗
  • MalfunctionJun 11, 2018Break; Break

    IT WAS REPORTED THAT INSTRUMENT WAS FOUND TO BE DAMAGED/ BROKEN PRIOR TO THE PROCEDURE- ALL BROKEN PARTS WERE RETRIEVED FROM THE FIELD PRIOR TO INCISION. SURGERY TIME WAS NOT DELAYED DUE TO THIS INCIDENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7586312 ↗
  • MalfunctionMay 31, 2018Break; Break

    IT WAS REPORTED THAT ATTUNE POLY TRIAL HAS A BROKEN PEG FROM DOCTOR EXTRACTING IT FROM THE TRAIL. ALL PIECES RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7556299 ↗
  • MalfunctionJan 10, 2018Break; Break

    TOTAL KNEE REPLACEMENT - DR. (B)(6), (B)(6) ON (B)(6) 2017. WHILE PERFORMING A TOTAL KNEE REPLACEMENT, IT WAS NOTICED THAT THERE WAS DAMAGE TO AN ATTUNE ARTICULATING SURFACE AS WELL AS 2 ATTUNE SHIMS. THIS DAMAGE INCLUDES A SILVER O RING BROKEN FROM A SHIM AND STUCK TO THE ARTICULATING SURFACE AS EVIDENCED BY PHOTOS ATTACHED TO THIS EMAIL. THE ADDITIONAL SHIM HAS A SMALL CRACK IN IT. THESE WERE NOTICED BEFORE BEING USED ON THE PATIENT. NO DELAY OR ADVERSE RESULT WAS REPORTED. (B)(6). NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7178108 ↗
  • MalfunctionNov 16, 2017Break; Break

    IT WAS REPORTED THAT THE TRIALS BROKE WHILE TRIALING.

    Patient problems: No Information

    MAUDE report 7036915 ↗
  • MalfunctionAug 15, 2017Break; Break

    IT WAS REPORTED THAT THE MACRO TRIAL HAD A SPRING COME OFF AFTER TRAILLING. IT WAS STATED THAT THE X-RAY CONFIRMED THAT THE SPRING WAS NOT ON PATIENTS KNEE.

    Patient problems: No Information

    MAUDE report 6794619 ↗
  • MalfunctionJun 7, 2016Break; Break

    THE ARTICULAR SURFACE TRIAL BROKE IN TWO PIECES WHEN REMOVING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5702754 ↗