SUMMIT 2570-00-005Recalls & reported problems

Depuy (Ireland) · SUMMIT LATERALIZER · Class 1

Quick summary

This DePuy Summit lateralizer is a reamer tool used in hip surgery. The 30 reports mostly describe the tool becoming dull or worn with use so it no longer cuts or attaches properly, without delay or harm mentioned in most cases. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10603295142867), received between Sep 19, 2018 and Oct 29, 2024 — classified as 30 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut17
  • Naturally Worn8
  • Dull, Blunt7
  • Device-Device Incompatibility3
  • Material Twisted/Bent3
  • Illegible Information2
  • Electrical Power Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Consequences Or Impact To Patient12
  • Insufficient Information3
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2024Failure to Cut; Failure to Cut

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAYS OR NEGATIVE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20556539 ↗
  • MalfunctionOct 11, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP SURGERY ON (B)(6) 2024. THE HOSPITAL STAFF DROPPED THE BIPOLAR RETAINING CLIP ON THE FLOOR AND THREW IT IN THE TRASH. ALSO, THE SUMMIT REAMERS HAVE BECOME DULL. THIS DID NOT DELAY SURGERY. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20429763 ↗
  • MalfunctionOct 11, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP SURGERY ON (B)(6) 2024. THE HOSPITAL STAFF DROPPED THE BIPOLAR RETAINING CLIP ON THE FLOOR AND THREW IT IN THE TRASH. ALSO, THE SUMMIT REAMERS HAVE BECOME DULL. THIS DID NOT DELAY SURGERY. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20429592 ↗
  • MalfunctionOct 11, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP SURGERY ON (B)(6) 2024 . THE HOSPITAL STAFF DROPPED THE BIPOLAR RETAINING CLIP ON THE FLOOR AND THREW IT IN THE TRASH. ALSO, THE SUMMIT REAMERS HAVE BECOME DULL. THIS DID NOT DELAY SURGERY. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20429552 ↗
  • MalfunctionOct 20, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT A BHA. DURING SURGERY, THE T-HANDLE IN QUESTION AND A REAMER DID NOT MATCH, SO THE T-HANDLE IN QUESTION COULD NOT BE USED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17974305 ↗
  • MalfunctionAug 25, 2023Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT A SUMMIT LATERALIZER WILL NO LONGER ATTACH TO THE CHUCK ATTACHMENT AND IS NO LONGER USABLE DUE TO WEAR AND TEAR. THIS OCCURRED INTRA OP BUT SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAYS OR PATIENT ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17627817 ↗
  • MalfunctionSep 14, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE LATERALIZER IS DULL. THERE WAS A 10 MINUTE SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15419076 ↗
  • MalfunctionAug 3, 2022Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE LATERALIZER IS DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15157018 ↗
  • MalfunctionJan 14, 2022Failure to Cut; Failure to Cut

    INSTRUMENT IS DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13260417 ↗
  • MalfunctionJan 14, 2022Failure to Cut; Failure to Cut

    INSTRUMENTS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13258839 ↗
  • MalfunctionJan 14, 2022Failure to Cut; Failure to Cut

    INSTRUMENTS ARE DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13258837 ↗
  • MalfunctionDec 17, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn

    LATERALIZER REAMER FROM TWO SUMMIT TRAYS WERE DULL. THERE WERE KICKS AND BURR MARKS ON TWO BROACH HANDLE ALIGNMENT RODS. NO SURGICAL DELAY OR ADVERSE AFFECT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13026482 ↗
  • MalfunctionDec 17, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn

    LATERALIZER REAMER FROM TWO SUMMIT TRAYS WERE DULL. THERE WERE KICKS AND BURR MARKS ON TWO BROACH HANDLE ALIGNMENT RODS. NO SURGICAL DELAY OR ADVERSE AFFECT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13026479 ↗
  • MalfunctionDec 7, 2021Failure to Cut; Naturally Worn; Failure to Cut; Naturally Worn

    IT WAS REPORTED THAT THE INSTRUMENT IS BLUNT AND WORN DUE TO USAGE. SECOND KIT WAS OPENED SO THE SURGERY WAS CONTINUED. THERE WAS A THREE MINUTE SURGICAL DELAY. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12950782 ↗
  • MalfunctionMay 10, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE REAMER IS DULL AND NEEDS REPLACEMENT. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11797707 ↗
  • MalfunctionMay 10, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT THE REAMER IS DULL AND NEEDS TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11797626 ↗
  • MalfunctionJan 22, 2021Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT REAMER IS DULL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11214593 ↗
  • MalfunctionAug 26, 2020Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE HUDSON CONNECTION ON REAMER WAS WORN. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10456582 ↗
  • MalfunctionMar 20, 2020Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information; Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information

    SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9859627 ↗
  • MalfunctionMar 20, 2020Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information; Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information

    SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9859626 ↗
  • MalfunctionFeb 5, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SUMMIT STRAIGHT REAMER SIZES 0-1,2-3,4-5,6-7,8-9, AND LATERALIZER ARE STRIPPED AT THE TOP AND WILL NOT WORK PROPERLY IN THE HUDSON ADAPTOR. NO PIECES ARE BROKEN AND ALL PARTS WERE RETRIEVED AND SENT TO THE OFFICE. THE SCRUB TECHNIQUE MADE ME AWARE. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9672564 ↗
  • MalfunctionJul 15, 2019Dull, Blunt; Dull, Blunt

    THE SUMMIT TAPERED REAMERS NEED REPLACEMENTS. THE SURGEON FEELS THAT THEY AREN¿T SHARP ENOUGH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8790257 ↗
  • MalfunctionJul 15, 2019Dull, Blunt; Dull, Blunt

    THE SUMMIT TAPERED REAMERS NEED REPLACEMENTS. THE SURGEON FEELS THAT THEY AREN¿T SHARP ENOUGH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8790255 ↗
  • MalfunctionJul 15, 2019Dull, Blunt; Dull, Blunt

    THE SUMMIT TAPERED REAMERS NEED REPLACEMENTS. THE SURGEON FEELS THAT THEY AREN¿T SHARP ENOUGH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8790254 ↗
  • MalfunctionMay 17, 2019Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE LATERALIZING REAMER IS DULL, MAKING THE REMOVING OF BONE DIFFICULT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8619355 ↗

Other REAMER devices

All REAMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.