Skip to content

SUMMIT 2570-00-070Recalls & reported problems

Depuy (Ireland) · SUMMIT BROACH SIZE 1 · Class 1

All models of SUMMIT (43)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295142881), received between Mar 23, 2020 and Mar 22, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Illegible Information” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Illegible Information4
  • Material Twisted/Bent4
  • Dull, Blunt2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 22, 2021Material Twisted/Bent; Illegible Information; Material Twisted/Bent; Illegible Information

    IT WAS REPORTED THAT THE BROACHES WERE WORN FROM NORMAL USE THEY WERE NOT REMOVING ENOUGH BONE THEN CAUSING THE IMPLANT TO SIT PROUD A BIT. SURGERY WASN¿T DELAYED AND NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11541695 ↗
  • MalfunctionMar 22, 2021Material Twisted/Bent; Illegible Information; Material Twisted/Bent; Illegible Information

    IT WAS REPORTED THAT THE BROACHES WERE WORN FROM NORMAL USE THEY WERE NOT REMOVING ENOUGH BONE THEN CAUSING THE IMPLANT TO SIT PROUD A BIT. SURGERY WASN¿T DELAYED AND NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11541589 ↗
  • MalfunctionMar 23, 2020Dull, Blunt; Material Twisted/Bent; Illegible Information; Dull, Blunt; Material Twisted/Bent; Illegible Information

    SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9867889 ↗
  • MalfunctionMar 23, 2020Dull, Blunt; Material Twisted/Bent; Illegible Information; Dull, Blunt; Material Twisted/Bent; Illegible Information

    SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9867885 ↗

Other BROACH devices

All BROACH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page