SUMMIT 2570-00-150Recalls & reported problems
Depuy (Ireland) · SUMMIT BROACH SIZE 8 · Class 1
All models of SUMMIT (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295142959), received between Feb 23, 2018 and Nov 21, 2023 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Dull, Blunt2
- Illegible Information2
- Material Twisted/Bent2
- Adverse Event Without Identified Device or Use Problem1
- Device-Device Incompatibility1
- Material Deformation1
Patient problems
- No Consequences Or Impact To Patient4
- Bone Fracture(s)1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Not Applicable1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2023Break; Break
IT WAS REPORTED THAT THE POST ON SIZE 8 BROACH BROKE IN HALF WHILE BROACHING WITH KINCISE. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 18181077 ↗MalfunctionSep 30, 2022Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT WHILE REMOVING THE SUMMIT BROACH FROM THE FEMORAL CANAL, THE BROACH ATTACHMENT POST BECAME BENT. THIS MAKES THE BROACH UNABLE TO ATTACH TO THE BROACH HANDLE. ALL PARTS RETURNING. SURGERY WAS NOT DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15509397 ↗InjuryNov 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING BROACHING PROXIMAL FEMUR CRACKED. CERCLAGE WIRE WAS PLACED, CASE COMPLETED SUCCESSFULLY. DOI: UNKNOWN, DOR: (B)(6) 2021, UNKNOWN SIDE.
Patient problems: Bone Fracture(s)
MAUDE report 12752556 ↗MalfunctionMar 23, 2020Dull, Blunt; Material Twisted/Bent; Illegible Information; Dull, Blunt; Material Twisted/Bent; Illegible Information
SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9867994 ↗MalfunctionMar 23, 2020Dull, Blunt; Material Twisted/Bent; Illegible Information; Dull, Blunt; Material Twisted/Bent; Illegible Information
SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9867987 ↗InjuryOct 18, 2019Break; Break
IT WAS REPORTED THAT THE BROKEN INSTRUMENTATION REQUIRING OSTEOTOMY. PATIENT WAS BROUGHT TO THE OPERATING ROOM FOR A LEFT ANTERIOR TOTAL HIP REPLACEMENT. DURING THE PROCEDURE, THERE WAS MALFUNCTION OF THE DEPUY FEMORAL TRIAL BROACH (FRACTURE OF THE BROACH) WHICH REQUIRED OSTEOMY OF THE FEMUR FOR THE REMOVAL OF THE TRIAL BROACH.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 9209054 ↗InjuryMar 18, 2019Break; Break
SIZE 8 SUMMIT BROACH BECAME INCARCERATED IN FEMUR DURING DAA THR. SIZE 8 SUMMIT BROACH EXTRACTION TRUNION BROKE WHILE ATTEMPTING THE EXTRACT BROACH. SURGERY WAS DELAYED OF 180 MINUTES. FEMORAL OSTEOTOMY TO EXTRACT BROACH WAS PERFORMED WHEN EVENT OCCURRED TO COMPLETE THE SURGERY SUCCESSFULLY. THERE WERE FRAGMENTS GENERATED BUT WERE REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION.
Patient problems: Not Applicable
MAUDE report 8429881 ↗MalfunctionFeb 23, 2018Break; Break
IT WAS REPORTED THAT DURING BROACHING OF THE FEMUR THE MALE END OF THE BROACH BROKE OFF IN THE BROACH HANDLE. ALL PIECES WERE RETRIEVED. SURGICAL DELAY OF 10 MINUTES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7291956 ↗
Other BROACH devices
- NA 2598-07-550Depuy (Ireland) · BROACH323 reports
- SUMMIT 2570-00-000Depuy (Ireland) · BROACH269 reports
- NA 952212500FDepuy (Ireland) · BROACH245 reports
- ACTIS 2010-01-050Depuy (Ireland) · BROACH235 reports
- CORAIL 9522-11-500Depuy (Ireland) · BROACH210 reports
- ACTIS 2010-01-060Depuy (Ireland) · BROACH186 reports
- ACTIS 2010-01-040Depuy (Ireland) · BROACH172 reports
- ACTIS 2010-01-070Depuy (Ireland) · BROACH162 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page