SUMMIT 2570-02-000Recalls & reported problems
Depuy (Ireland) · SUMMIT TAPERED REAMER SIZE 0/1 · Class 1
All models of SUMMIT (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295142997), received between Dec 4, 2018 and Oct 29, 2024 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut11
- Dull, Blunt9
- Material Twisted/Bent3
- Device-Device Incompatibility2
- Illegible Information2
- Naturally Worn1
Patient problems
- No Consequences Or Impact To Patient12
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2024Failure to Cut; Failure to Cut
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAYS OR NEGATIVE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556602 ↗MalfunctionOct 29, 2024Failure to Cut; Failure to Cut
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAYS OR NEGATIVE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556529 ↗MalfunctionOct 11, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP SURGERY ON OCTOBER 4, 2024. THE HOSPITAL STAFF DROPPED THE BIPOLAR RETAINING CLIP ON THE FLOOR AND THREW IT IN THE TRASH. ALSO, THE SUMMIT REAMERS HAVE BECOME DULL. THIS DID NOT DELAY SURGERY. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 20429581 ↗MalfunctionOct 11, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP SURGERY ON (B)(6) 2024. THE HOSPITAL STAFF DROPPED THE BIPOLAR RETAINING CLIP ON THE FLOOR AND THREW IT IN THE TRASH. ALSO, THE SUMMIT REAMERS HAVE BECOME DULL. THIS DID NOT DELAY SURGERY. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 20429576 ↗MalfunctionOct 11, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE PATIENT UNDERWENT A HIP SURGERY ON (B)(6) 2024. THE HOSPITAL STAFF DROPPED THE BIPOLAR RETAINING CLIP ON THE FLOOR AND THREW IT IN THE TRASH. ALSO, THE SUMMIT REAMERS HAVE BECOME DULL. THIS DID NOT DELAY SURGERY. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 20429568 ↗MalfunctionMar 20, 2020Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information; Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information
SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9859644 ↗MalfunctionMar 20, 2020Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information; Dull, Blunt; Failure to Cut; Material Twisted/Bent; Illegible Information
SIZE ETCHINGS ARE WORN AND NOT LEGIBLE, AND EDGES ARE DULL AND NOT ADEQUATE IN CUTTING BONE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9859624 ↗MalfunctionFeb 5, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SUMMIT STRAIGHT REAMER SIZES 0-1,2-3,4-5,6-7,8-9, AND LATERALIZER ARE STRIPPED AT THE TOP AND WILL NOT WORK PROPERLY IN THE HUDSON ADAPTOR. NO PIECES ARE BROKEN AND ALL PARTS WERE RETRIEVED AND SENT TO THE OFFICE. THE SCRUB TECHNIQUE MADE ME AWARE. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9672567 ↗MalfunctionJan 23, 2020Device-Device Incompatibility; Device-Device Incompatibility
SUMMIT REAMERS WERE NOT PROPERLY ENGAGING INTO THE ADAPTER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9619768 ↗MalfunctionOct 30, 2019Dull, Blunt; Failure to Cut; Dull, Blunt; Failure to Cut
IT WAS REPORTED THAT THE SUMMIT SZ0/1, 2/3, 4/5, 6/7, 8/9 TAPERED REAMER ARE ALL DULL. NEEDS REPLACEMENT X'S 4.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9257716 ↗MalfunctionOct 30, 2019Dull, Blunt; Failure to Cut; Dull, Blunt; Failure to Cut
IT WAS REPORTED THAT THE SUMMIT SZ0/1, 2/3, 4/5, 6/7, 8/9 TAPERED REAMER ARE ALL DULL. NEEDS REPLACEMENT X'S 4.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9257714 ↗MalfunctionOct 30, 2019Dull, Blunt; Failure to Cut; Dull, Blunt; Failure to Cut
IT WAS REPORTED THAT THE SUMMIT SZ0/1, 2/3, 4/5, 6/7, 8/9 TAPERED REAMER ARE ALL DULL. NEEDS REPLACEMENT X'S 4.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9257713 ↗MalfunctionOct 30, 2019Dull, Blunt; Failure to Cut; Dull, Blunt; Failure to Cut
IT WAS REPORTED THAT THE SUMMIT SZ0/1, 2/3, 4/5, 6/7, 8/9 TAPERED REAMER ARE ALL DULL. NEEDS REPLACEMENT X'S 4.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9257711 ↗MalfunctionJul 15, 2019Dull, Blunt; Dull, Blunt
THE SUMMIT TAPERED REAMERS NEED REPLACEMENTS. THE SURGEON FEELS THAT THEY AREN¿T SHARP ENOUGH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8790256 ↗MalfunctionJul 15, 2019Dull, Blunt; Dull, Blunt
THE SUMMIT TAPERED REAMERS NEED REPLACEMENTS. THE SURGEON FEELS THAT THEY AREN¿T SHARP ENOUGH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8790226 ↗MalfunctionJul 15, 2019Dull, Blunt; Dull, Blunt
THE SUMMIT TAPERED REAMERS NEED REPLACEMENTS. THE SURGEON FEELS THAT THEY AREN¿T SHARP ENOUGH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8790224 ↗MalfunctionDec 4, 2018Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS REPORTED THAT THE SURGEON WENT TO REAM THE FEMORAL CANAL AND THE REAMERS DID NOT CATCH IN THE HUDSON ADAPTER. THE HUDSON ADAPTER WAS SPINNING AND THE REAMERS WERE NOT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8128921 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.