RECLAIM 2976-14-140Recalls & reported problems
Depuy (Ireland) · RECLAIM TAPERED DISTAL REAMER 14mm X 140mm · Class 1
All models of RECLAIM (149)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10603295156475), received between Aug 28, 2018 and Nov 10, 2022 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Cut4
- Detachment of Device or Device Component1
- Device-Device Incompatibility1
- Material Twisted/Bent1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
- No Code Available1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2022Failure to Cut; Failure to Cut
IT WAS REPORTED THE REAMER IS DULL. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15775669 ↗MalfunctionAug 4, 2022Failure to Cut; Failure to Cut
DISTAL REAMERS ARE DULL AND NEED REPLACED. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. NO PATIENT CONSEQUENCES OR ADVERSE EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15163024 ↗MalfunctionAug 4, 2022Failure to Cut; Failure to Cut
DISTAL REAMERS ARE DULL AND NEED REPLACED. THE PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS. NO PATIENT CONSEQUENCES OR ADVERSE EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15162829 ↗MalfunctionSep 30, 2021Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE 14 MM REAMER IS CROSS THREADED AT THE TOP THREADS AND THE MARKINGS FOR THE SIZE ARE UNABLE TO BE READ ANYMORE. NOT USED IN A CASE SO NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12554389 ↗MalfunctionOct 15, 2020Failure to Cut; Failure to Cut
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO CONSEQUENCE OF DELAY, THE DELAY WAS ABOUT 5 MINUTES TRYING TO GET THE REAMERS TO ENTER THE CANAL.
Patient problems: No Code Available
MAUDE report 10685748 ↗MalfunctionOct 12, 2020Material Twisted/Bent; Material Twisted/Bent
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WERE NO PIECES BROKE OFF FROM THE INSTRUMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10667189 ↗MalfunctionJul 10, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE RECLAIM 14X140MM REAMERS IS NOT CONNECTING TO THE REAMER SHAFT PROPERLY AND NEEDS TO BE REPLACED. THE REAMERS LOCKS TO THE REAMERS SHAFT BUT IS LOOSES AND WOLVES AROUND. THIS PART NEED TO BE REPLACED AND IS BEING SENT BACK TO DEPUY. NO PARTS BROKE OFF THE INSTRUMENT. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8775450 ↗MalfunctionAug 28, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE RECLAIM 14X140 REAMER WAS MALFUNCTIONING AND NOT ATTACHING PROPERLY TO THE REAMER GUIDE. IT NEEDS TO BE REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7823930 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page