S-ROM NOILES 63-3805ARecalls & reported problems

Depuy (Ireland) · S-ROM NOILES IMPACTION CAP · Class 1

All models of S-ROM NOILES (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295180692), received between Jan 8, 2019 and Feb 12, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Naturally Worn3
  • Component Missing1
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1
  • Entrapment of Device1
  • Material Twisted/Bent1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Detachment of Device or Device Component; No Apparent Adverse Event

    ADDITIONAL INFORMATION STATES: CAN YOU PLEASE CLARIFY THE EXACT ALLEGATIONS AGAINST THE REPORTED PRODUCT? DID IT BREAK INTO TWO OR MORE PIECES? IS IT BENT, CRACKED, STRIPPED, DULL, WORN OR CROSS THREADED? MOST OF THEM HAVE JUST WEAR AND TEAR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24321444 ↗
  • MalfunctionSep 10, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT WHILE ASSEMBLING THE FEMORAL SLEEVE AND STEM TO THE ADAPTOR AND DISTAL FEMUR THE S-ROM IMPACTOR CAP GOT STUCK ON THE STEM. SURGEON HAD TO USE A SAW TO GET THE IMPACTOR LOOSE FROM THE STEM. THE IMPLANTS WERE NOT DAMAGED. SURGERY WAS NOT DELAYED. THE IMPACTOR CAP WAS DAMAGED AND NEEDS TO BE REPLACED. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20182150 ↗
  • MalfunctionJul 11, 2024Naturally Worn; Scratched Material; Naturally Worn; Scratched Material

    IT WAS REPORTED THAT THE IMPACTION CUP HAD DAMAGE ON THE INNER RIM THAT IS FLAKING OFF. THE LPS INSERTER/EXTRACTOR'S INTERNAL ROD IS BENT AND WEAR AND TEAR SCRATCHES ON IT MAKING IT HARD TO INSERT INTO THE HOLDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19720493 ↗
  • MalfunctionJul 9, 2024Break; Naturally Worn; Break; Naturally Worn

    ADDITIONAL INFORMATION WAS RECEIVED: THE INSTRUMENT IS A BLUE PLASTIC HOLLOW IMPACTOR. THE INTERNAL SURFACE WAS ROUGHENED AND WHEN IT CAME INTO CONTACT WITH THE IMPLANT DURING IMPACTION, THE PLASTIC RUBBED OFF THE IMPACTOR AND ONTO THE SURFACE OF THE IMPLANT. THIS ITEM IS USED ON THE BACK TABLE TO PUT THE IMPLANTS TOGETHER SO FRAGMENTS DO NOT FLAKE OFF INTO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19706521 ↗
  • MalfunctionJan 18, 2023Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE IMPACTOR CAP, THE PLASTIC IS CHIPPED FROM IMPACTION OF THE CEMENTED STEM AND DISTAL FEMUR MATING TOGETHER. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16196679 ↗
  • MalfunctionMar 9, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE LPS STEM INSERTER IS BENT AND WILL NOT WORK, AND NEEDS TO BE REPLACED. THE BLUE SROM IMPACTION PIECE IS DAMAGED ON THE INSIDE DUE TO EXCESSIVE WEAR AND NEEDS TO BE REPLACED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11446296 ↗
  • MalfunctionNov 19, 2020Component Missing; Naturally Worn; Component Missing; Naturally Worn

    THE INSTRUMENT IS PLASTIC, SO IT APPEARS TO HAVE BEEN DAMAGED DURING USAGE AND HAS A PIECE MISSING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10867878 ↗
  • MalfunctionSep 27, 2019Break; Break

    IT WAS REPORTED THAT THE IMPACTOR LPS WAS DAMAGED DUE TO BREAKAGE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9129581 ↗
  • MalfunctionJan 8, 2019Break; Break

    THE SROM IMPACTOR IS CHIPPED/TORN UP ON THE EDGES/INSIDE AND NEEDS TO BE REPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8227328 ↗