The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295217992), received between Sep 15, 2017 and Oct 7, 2022 — classified as 7 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture3
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information1
- Osseointegration Problem1
Patient problems
- No Code Available3
- Pain3
- Fall1
- Inadequate Osseointegration1
- Inflammation1
- Insufficient Information1
- Joint Disorder1
- Joint Laxity1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryOct 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT UNDERWENT THE INITIAL PROCEDURE ON THE (B)(6) 2022 (RECEIVED AN S-ROM NOILES ROTATING HINGE KNEE). THE PATIENT FELT HIS KNEE "GIVE WAY", FOLLOWED BY SIGNIFICANT PAIN. UPON INVESTIGATION IT WAS DISCOVERED THAT THE "BOSS" FOR THE SLEEVE ATTACHMENT ON THE FEMORAL COMPONENT HAD BROKEN RESULTING IN FAILURE OF THE CONSTRUCT. THE PATIENT UNDERWENT REVISION SURGERY AND THE DAMAGED FEMORAL COMPONENTS WERE REMOVED AND REVISED TO AN LPS DISTAL FEMORAL REPLACEMENT. THE RETRIEVED IMPLANTS WERE SENT TO THE BIO-ENGINEERING DEPARTMENT AT ROYAL PERTH HOSPITAL AS PER STANDARD PROTOCOL HERE IN WA. IF OTHER, DESCRIBE --> N/A, IF OTHER, DESCRIBE --> REVISION KNEE SURGERY., PATIENT STATUS/ OUTCOME / CONSEQUENCES --> YES, PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)?--> PAIN, INFLAMMATION, WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: --> YES, IF YES, DESCRIBE --> X-RAYS, REVISION SURGERY, OTHER INFORMATION --> N/A, IS THE PATIENT PART OF A CLINICAL STUDY --> UNKNOWN, IP-(B)(6). DEVICE PROPERTY OF -->NONE, DEVICE IN POSSESSION OF -->NONE, IP-(B)(6). DEVICE PROPE
Patient problems: Fall; Inflammation; Joint Laxity; Pain
MAUDE report 15568441 ↗InjuryApr 12, 2022Osseointegration Problem; Osseointegration Problem
REVISION COMPLETED DUE TO LOOSENING OF THE FEMORAL COMPONENT AT THE BONE TO IMPLANT INTERFACE. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT KNEE.
Patient problems: Inadequate Osseointegration
MAUDE report 14084472 ↗InjurySep 10, 2021Fracture; Fracture
(B)(4). THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE FEMORAL COMPONENT AND THE FEMORAL SLEEVE AT THE BONE TO IMPLANT INTERFACE. PATIENT HAD A LEFT DISTAL FEMORAL REPLACEMENT ON (B)(6) 2019 AFTER HISTORY OF FAILED HARDWARE PLATE AND SCREWS OF THE FEMUR. IT WAS REVISED FOR LOOSE SLEEVE COMPONENT AND FRACTURE OF THE STEM SLEEVE JUNCTION. THE STEM WAS WELL FIXED WITH FRACTURE AT THE THREADED PORTION OF THE STEM, LEAVING BEHIND THE THREADS OF THE STEM IN THE SLEEVE. THE SLEEVE AND FEMORAL CONSTRUCT WERE LOOSE AND REMOVED AS WELL WAS THE POLY. THE TIBIA WAS RETAINED AS WELL AS THE PATELLA COMPONENT OF THE LEFT KNEE. NO DELAY IN SURGERY. DOI:(B)(6) 2019. DOR:(B)(6) 2021. LEFT KNEE.
Patient problems: Insufficient Information
MAUDE report 12454022 ↗InjuryJul 30, 2019Fracture; Fracture
IT WAS REPORTED THAT THE SCREW THAT HOLDS THE STEM AND FEMORAL SLEEVE TOGETHER BROKE DUE TO LACK OF GOOD BONE SURROUNDING INTERFACE. THE SURGEON STATED THAT THE PATIENT DID NOT ACHIEVE BONE GROWTH INTO THE SLEEVE CAUSING SUPERIOR MIGRATION OF THE SLEEVE. THE SURGEON CEMENTED AN LPS STEM IN THE REVISED CONSTRUCT. DOI: (B)(6) 2017; DOR: (B)(6) 2019, LEFT KNEE.
Patient problems: No Code Available
MAUDE report 8840890 ↗InjuryJul 2, 2019Fracture; Fracture
THE PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE PRESSFIT SLEEVE COMPONENT AT THE BONE TO IMPLANT INTERFACE. IT WAS ALSO REPORTED THAT THE THREADS ON THE UNIVERSAL STEM BROKE OFF INSIDE THE FEMORAL SLEEVE. THE PRESSFIT SLEEVE CAME OUT EASILY AND THEY USED A TREPHINE TO REMOVE THE BROKEN STEM. DOI: (B)(6) 2018; DOR: (B)(6) 2019, LEFT KNEE.
Patient problems: No Code Available; Pain
MAUDE report 8754312 ↗InjuryNov 13, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS FEMUR FRACTURE.
Patient problems: No Code Available
MAUDE report 8064373 ↗InjurySep 15, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
THE PATIENT WAS REVISED TO ADDRESS PAIN AND INSTABILITY.
Patient problems: Joint Disorder; Pain
MAUDE report 6868533 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page