This DePuy impactor tip is a surgical instrument used to seat hip replacement liners during hip replacement surgery. Most of the 50 reports describe the tip breaking, often into two pieces, while the surgeon was impacting the implant; in the reports reviewed, all pieces were retrieved and surgery continued without delay or harm. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 10603295221883), received between Oct 11, 2017 and Feb 3, 2026 — classified as 50 malfunctions.

The most frequently reported specific device problem is “Break” (49 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break49
  • Device-Device Incompatibility2
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions34
  • No Consequences Or Impact To Patient10
  • Insufficient Information3
  • No Information2
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2026Break

    IT WAS REPORTED THAT THE 36 MM IMPACTOR TIP BROKE INTO TWO PIECES. ALL PIECES WERE RETRIEVED. NO DELAY OR ADVERSE EVENT OCCURRED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24248044 ↗
  • MalfunctionOct 22, 2025Break; Break

    IT WAS REPORTED THAT THE 36 MM LINER IMPACTOR TIP WAS BROKEN INTO TWO PIECES AND NEEDS TO BE REPLACED. ALL PIECES WERE RETRIEVED. THERE WAS NO DELAY OR ADVERSE EVENT WAS EXPERIENCES. THERE WAS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23357273 ↗
  • MalfunctionMar 6, 2025Break; Break

    IT WAS REPORTED THAT THE LINER IMPACTOR TIP BROKE DURING IMPACTING. SURGERY WASN'T DELAYED DUE TO THE REPORTED EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 21539606 ↗
  • MalfunctionOct 7, 2024Break; Break

    IT WAS REPORTED THAT THE IMPACTOR TIP BROKE AT BASE OF INSERTION PART. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: Insufficient Information

    MAUDE report 20385670 ↗
  • MalfunctionJun 7, 2024Break; Break

    IT WAS REPORTED THAT THE PINNACLE LINER IMPACTOR BROKE INTO TWO PIECES WHEN IMPACTING IMPLANT INTO PATIENT. ALL PIECES REMOVED. NO DELAY IN SURGERY NO HARM TO PT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19484485 ↗
  • MalfunctionDec 29, 2023Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP BROKE INTO TWO PIECES. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18416512 ↗
  • MalfunctionNov 9, 2023Break; Break

    IT WAS REPORTED THAT THE SURGEON WAS INSERTING AN ALTRX LINER, AND THE ORANGE IMPACTOR TIP BROKE INTO TWO PIECES. BOTH PIECES WERE RETRIEVED AND MATCHED TOGETHER TO MAKE SURE THAT NO SMALLER PIECES WERE GENERATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18103811 ↗
  • MalfunctionOct 13, 2023Break; Break

    IT WAS REPORTED THAT THE IMPACTOR TIP 36MM FROM OPTIMUS TRAY WAS DAMAGED. THE PLASTIC IMPACTOR TIP WAS BROKEN INTO TWO PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17930043 ↗
  • MalfunctionOct 12, 2023Break; Break

    DEFECTIVE DEVICE: LINER IMPACTOR HEAD (THE OLD VERSION) - THE PLASTIC STEM OF THE HEAD BROKE. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17919649 ↗
  • MalfunctionJul 25, 2023Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP BROKE INTO TWO PIECES. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390080 ↗
  • MalfunctionMay 12, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT TWO DIFFERENT OFFSET IMPACTORS BREAK DURING A TOTAL HIP. ON ONE HANDLE THE HANDLE ITSELF COMPLETELY BROKE. ON THE OTHER ONE THE THREADED END PIECE BECAME STUCK AND DISASSOCIATED FROM THE POSITIONING ROD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16919570 ↗
  • MalfunctionMar 28, 2023Break; Break

    IT WAS REPORTED THE OPTIMUS IMPACTOR TIP IS BROKEN WITH SMALL PART BROKEN OFF INSIDE THE IMPACTOR HANDLE ON OPENING THE STERILE TRAY. PASSED OFF BEFORE CASE. NO ADVERSE AFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16632778 ↗
  • MalfunctionMar 6, 2023Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP IS CRACKED AND NEEDS TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16488068 ↗
  • MalfunctionJan 6, 2023Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP IS BROKEN INTO TWO PIECES. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16114199 ↗
  • MalfunctionDec 15, 2022Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP BROKE WHILE IMPACTING THE LINER. ALL PIECES WERE RETRIEVED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15993098 ↗
  • MalfunctionNov 28, 2022Break; Break

    OPTIMUS LINER INSERTER TIP (SIZE 36- ORANGE) FRACTURES UPON IMPACTION OF LINER. THE ORANGE TIP SNAPPED WHEN BEING IMPACTED BY A MALLET. THERE DID NOT SEEM TO BE UNNECESSARY FORCE APPLIED. ALL PIECES REMOVED. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15868487 ↗
  • MalfunctionNov 1, 2022Break; Break

    THE INSERTION END OF THE 36MM LINER IMPACTOR TIP SNAPPED OFF WHILE IMPACTING A 36MM ACETABULAR LINER. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15708060 ↗
  • MalfunctionOct 12, 2022Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    STAFF HAVE INDICATED ALL ITEMS REQUIRE REPLACEMENT DUE TO WEAR AND TEAR OVER TIME. THE FEMORAL HEAD TRIALS ARE WORN AND SCRATCHED. THE FEMORAL HEAD IMPACTOR TIP IS ALSO VERY WORN AWAY FROM SEVERAL USES. THE IMPACTOR TIPS ARE CROSS THREADED AND NO LONGER THREAD ONTO THE WHITE HANDLE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15591074 ↗
  • MalfunctionOct 4, 2022Break; Break

    IT WAS REPORTED THAT THE STEM OF IMPACTOR BROKE OFF. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15536920 ↗
  • MalfunctionAug 22, 2022Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP WAS BROKEN INTO 2 PIECES. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15265519 ↗
  • MalfunctionAug 19, 2022Break; Break

    IT WAS REPORTED THAT THE MI IMPACTOR TIP IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15258697 ↗
  • MalfunctionJul 20, 2022Break; Break

    IT WAS REPORTED THAT THE 36MM IMPACTOR TIP BROKE WHILE IMPACTING THE LINER. NO ADDITIONAL INFORMATION IS AVAILABLE. THERE WAS NO LOOSENING AND NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15064532 ↗
  • MalfunctionMay 7, 2022Break; Break

    IT WAS REPORTED THAT WHILE INSERTING AN ALTRX LINER, AND THE IMPACTOR TIP BROKE INTO TWO PIECES. BOTH PIECES WERE RETRIEVED AND MATCHED TOGETHER TO MAKE SURE THAT NO SMALLER PIECES WERE CREATED. THE BROKEN PIECES WERE ASKED TO BE WASHED SO WE COULD RETURN THEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14316880 ↗
  • MalfunctionApr 19, 2022Break; Break

    IT WAS REPORTED THAT 36MM ACETABULAR LINER IMPACTOR TIP BROKE WHILE IMPACTING THE LINER INTO THE SHELL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14142574 ↗
  • MalfunctionMar 14, 2022Break; Break

    IT WAS REPORTED THAT THE IMPACTOR TIP WAS BROKEN. OPERATION COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13757841 ↗