Quick summary

This DePuy Sigma femoral notch impactor is a surgical tool used to seat knee implant parts. Its 184 reports include no deaths and 1 injury. Most describe the tool's tip or other parts breaking during knee surgery, with pieces removed and no harm reported. Reports in the last 12 months (3) are down from 9 in the prior 12 months, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
183
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 183 adverse event reports that list this exact device (DI 10603295225256), received between Jul 7, 2017 and Apr 16, 2026 — classified as 1 injury, 182 malfunctions.

The most frequently reported specific device problem is “Break” (148 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 6 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break148
  • Scratched Material9
  • Material Deformation8
  • Material Integrity Problem6
  • Crack5
  • Naturally Worn4
  • Device-Device Incompatibility3
  • Loose or Intermittent Connection3

Patient problems

  • No Consequences Or Impact To Patient89
  • No Clinical Signs, Symptoms or Conditions80
  • No Information9
  • Insufficient Information6
  • No Code Available4
  • Not Applicable2

Recent adverse event reports

The 25 most recent of 183 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Break

    IT WAS REPORTED THAT DURING A TOTAL KNEE ARTHROPLASTY (TKA) PROCEDURE, THE TIBIAL JIG STYLUS PLASTIC END BROKE OFF. ALL PIECES WERE REMOVED AND NO HARM WAS DONE TO THE PATIENT. WHEN USING SECONDARY IMPACTOR SIGMA NOTCH IMPACTOR TIP BROKE. ALL PIECES WERE REMOVED AND NO HARM WAS DONE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24897111 ↗
  • MalfunctionDec 19, 2025Break; Break

    IT WAS REPORTED THAT THE SIGMA IMPACTOR TIP BROKE WHILE IMPACTING FEMUR. ALL PIECES WERE REMOVED AND NO HARM WAS DONE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23854090 ↗
  • MalfunctionOct 27, 2025Break; Break

    IT WAS REPORTED THAT THE SIGMA FEMORAL IMPACTOR BROKE INTO TWO PIECES. ALL PIECES WERE RETRIEVED. NO DELAY OR ADVERSE EVENT WAS EXPERIENCED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393552 ↗
  • MalfunctionOct 24, 2025Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL KNEE ARTHROPLASTY (TKA). WHEN IMPACTING FEMUR SIGMA IMPACTOR TIP BROKE. ALL PIECES WERE REMOVED AND NO HARM WAS DONE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383381 ↗
  • MalfunctionOct 23, 2025Break; Break

    IT WAS REPORTED THAT WHEN IMPACTING FEMUR IMPLANT INITIALLY RED KNOB BROKE OFF. ALL PIECES WERE RECOVERED AND NO HARM WAS DONE TO THE PATIENT. THEN WITH THE SECONDARY IMPACTOR THE SIGMA IMPACTOR TIP BROKE. ALL PIECES WERE RECOVERED AND NO HARM WAS DONE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368102 ↗
  • MalfunctionOct 15, 2025No Apparent Adverse Event; No Apparent Adverse Event

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT: "VISUAL INSPECTION REVEALED ONE PORTION BROKEN AT THE LATERAL SIDE OF THE RADEL HEAD. DEVICE HAD SIGNS OF HEAVY USAGE ON ITS OVERALL SURFACE AND THE BROKEN PORTION WAS NOT RETURNED FOR EVALUATION".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23279967 ↗
  • MalfunctionOct 15, 2025No Apparent Adverse Event; No Apparent Adverse Event

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT: "VISUAL INSPECTION REVEALED ONE PORTION BROKEN AT THE LATERAL SIDE OF THE RADEL HEAD. DEVICE HAD SIGNS OF HEAVY USAGE ON ITS OVERALL SURFACE AND THE BROKEN PORTION WAS NOT RETURNED FOR EVALUATION".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23279958 ↗
  • MalfunctionOct 15, 2025No Apparent Adverse Event; No Apparent Adverse Event

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT: "VISUAL INSPECTION REVEALED ONE PORTION BROKEN AT THE LATERAL SIDE OF THE RADEL HEAD. DEVICE HAD SIGNS OF HEAVY USAGE ON ITS OVERALL SURFACE AND THE BROKEN PORTION WAS NOT RETURNED FOR EVALUATION".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23279754 ↗
  • MalfunctionSep 3, 2025Break; Break

    IT WAS REPORTED THAT SIGMA FEMORAL NOTCH IMPACTOR BROKE WHILE IMPACTING FEMUR IMPLANT IN. ALL PIECES WERE RECOVERED AND NO HARM WAS DONE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22952398 ↗
  • MalfunctionNov 7, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT IS BROKEN. THE EVENT DID NOT HAPPEN DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20632594 ↗
  • MalfunctionNov 7, 2024Break; Break

    IT WAS REPORTED THAT THE PLASTIC BROKE IN HALF. SURGICAL DELAY WAS UNKNOWN. NO ADDITIONAL INFORMATION PROVIDED TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20632194 ↗
  • MalfunctionSep 6, 2024Break; Break

    IT WAS REPORTED THAT THE FEMORAL IMPACTOR BROKE ON IMPACTION OF COMPONENT.

    Patient problems: Insufficient Information

    MAUDE report 20155544 ↗
  • MalfunctionJul 31, 2024Break; Break

    IT WAS REPORTED IMPACTOR PLASTIC BROKE INTO TWO PIECES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19874066 ↗
  • MalfunctionJul 1, 2024Device-Device Incompatibility; Device-Device Incompatibility

    THE FEMORAL IMPACTOR AND HANDLE HAVE FUSED TOGETHER AND COULD NOT BE PULLED APART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643833 ↗
  • MalfunctionMay 16, 2024Break; Break

    IT WAS REPORTED THAT THE FEMORAL NOTCH IMPACTOR BROKE WHILE IMPLANTING FINAL FEMORAL COMPONENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19330453 ↗
  • MalfunctionApr 30, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING KNEE SURGERY. IT DID NOT CAUSE A DELAY, NO PARTS WERE LEFT IN THE PATIENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19211998 ↗
  • MalfunctionMar 11, 2024Break; Break

    IT WAS REPORTED THAT THE IMPACTOR BROKE WHILE IMPACTING FEM TRIAL. UNKNOWN SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18873510 ↗
  • MalfunctionFeb 22, 2024Break; Break

    IT WAS REPORTED THAT THE SIGMA NOTCH IMPACTOR BROKE INTO TWO PIECES DURING IMPACTION. SLIGHT DELAY IN CASE WHILE OPENING BACK UP INSTRUMENT TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18756882 ↗
  • MalfunctionJan 16, 2024Break; Break

    IT WAS REPORTED THAT SIGMA FEMORAL IMPACTOR BROKE. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18520333 ↗
  • MalfunctionDec 4, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT UNIVERSAL HANDLE PIN HAS BROKEN AND THE FEMORAL IMPACTOR IS STUCK IN PLACE. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18263784 ↗
  • MalfunctionNov 16, 2023Break; Break

    WHILE IMPACTING THE FEMORAL IMPLANT, THE SIGMA HP FEMORAL NOTCH IMPACTOR BROKE INTO TWO PIECES ON THE BLACK RUBBER PART OF IMPACTOR. ALL PIECES WERE RETRIEVED FROM THE STERILE FIELD AND DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18153863 ↗
  • MalfunctionOct 13, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ITEMS JAMMED TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17929918 ↗
  • MalfunctionOct 5, 2023Break; Break

    IT WAS REPORTED THAT DURING FEMUR IMPACTION THE SIGMA HP FEM NOTCH IMPACTOR BROKE. BLACK PLASTIC BROKE IN HALF OFF OF METAL PART OF INSTRUMENT. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17883101 ↗
  • MalfunctionAug 29, 2023Break; Break

    IT WAS REPORTED THAT SIGMA IMPACTOR BROKE AT BLACK PLASTIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17643160 ↗
  • MalfunctionAug 29, 2023Break; Break

    IT WAS REPORTED THAT THE SIGMA IMPACTOR BROKE AT BLACK PLASTIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17643128 ↗

Other IMPACTOR devices

All IMPACTOR devices →