SIGMA LCS 9505-02-806Recalls & reported problems

Depuy (Ireland) · SIGMA HIGH PERFORMANCE LCS HIGH PERFORMANCE MBT TIBIA PREP CASE · Class 1

All models of SIGMA LCS (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295229346), received between Jan 10, 2019 and Dec 19, 2019 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Peeled/Delaminated10
  • Material Twisted/Bent6
  • Break4
  • Naturally Worn4
  • Adverse Event Without Identified Device or Use Problem2
  • Crack2

Patient problems

  • No Consequences Or Impact To Patient10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2019Crack; Peeled/Delaminated; Material Twisted/Bent; Crack; Peeled/Delaminated; Material Twisted/Bent

    IT WAS REPORTED THAT THE TRAYS ARE WORN AND OLD. NEED TO BE REPLACED. PLASTIC INSTRUMENT HOLDERS IN TRAYS ARE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9493392 ↗
  • MalfunctionDec 19, 2019Crack; Peeled/Delaminated; Material Twisted/Bent; Crack; Peeled/Delaminated; Material Twisted/Bent

    IT WAS REPORTED THAT THE TRAYS ARE WORN AND OLD. NEED TO BE REPLACED. PLASTIC INSTRUMENT HOLDERS IN TRAYS ARE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9493385 ↗
  • MalfunctionDec 6, 2019Peeled/Delaminated; Peeled/Delaminated

    THE VARNISHING OF TRAY IS EXFOLIATING. SEEN IN CSSD. NO PATIENT INVOLVEMENT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9432399 ↗
  • MalfunctionDec 6, 2019Peeled/Delaminated; Peeled/Delaminated

    THE VARNISHING OF TRAY IS EXFOLIATING. SEEN IN CSSD. NO PATIENT INVOLVEMENT. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9432327 ↗
  • MalfunctionOct 30, 2019Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn; Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9257639 ↗
  • MalfunctionOct 30, 2019Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn; Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9257636 ↗
  • MalfunctionOct 30, 2019Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn; Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9257632 ↗
  • MalfunctionOct 30, 2019Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn; Break; Peeled/Delaminated; Material Twisted/Bent; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9257631 ↗
  • MalfunctionJan 10, 2019Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8235757 ↗
  • MalfunctionJan 10, 2019Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8235722 ↗

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →