SIGMA 9505-02-808Recalls & reported problems

Depuy (Ireland) · SIGMA HIGH PERFORMANCE PATELLA & INSERTION CASE · Class 1

All models of SIGMA (379)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10603295229360), received between Mar 22, 2018 and Jun 17, 2022 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Peeled/Delaminated9
  • Naturally Worn6
  • Break4
  • Adverse Event Without Identified Device or Use Problem3
  • Crack2
  • Delamination2
  • Material Twisted/Bent2

Patient problems

  • No Consequences Or Impact To Patient11
  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 17, 2022Peeled/Delaminated; Naturally Worn; Peeled/Delaminated; Naturally Worn

    IT WAS REPORTED THAT INSIDE OF THE INSTRUMENT PANS WERE WORN AND FLAKING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14734201 ↗
  • MalfunctionApr 29, 2022Peeled/Delaminated; Naturally Worn; Peeled/Delaminated; Naturally Worn

    IT WAS REPORTED THAT DEVICE IS LOOKING WORN AND TIRED, COATING HAS STARTED TO PEEL AWAY ON THE INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14256294 ↗
  • MalfunctionMar 21, 2022Peeled/Delaminated; Peeled/Delaminated

    ACCOUNT COMPLAINT THAT THE HP SIGMA PANS WERE WORN OUT. ACCOUNT WAS REQUESTING TO REPLACE THESE WORN PANS BECAUSE ACCOUNT HAD A CONCERN ON THE FLAKING OF THE INTERIORS OF THE PAN TO POSSIBLY GET INTO WOUND OF PATIENT. THEREFORE ACCOUNT REQUESTED NEW PANS. SALES CONSULTANT WAS RETURNING THESE PANS FOR A CREDIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13840508 ↗
  • MalfunctionMar 21, 2022Peeled/Delaminated; Peeled/Delaminated

    ACCOUNT COMPLAINT THAT THE HP SIGMA PANS WERE WORN OUT. ACCOUNT WAS REQUESTING TO REPLACE THESE WORN PANS BECAUSE ACCOUNT HAD A CONCERN ON THE FLAKING OF THE INTERIORS OF THE PAN TO POSSIBLY GET INTO WOUND OF PATIENT. THEREFORE ACCOUNT REQUESTED NEW PANS. SALES CONSULTANT WAS RETURNING THESE PANS FOR A CREDIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13840502 ↗
  • MalfunctionDec 20, 2019Crack; Peeled/Delaminated; Material Twisted/Bent; Crack; Peeled/Delaminated; Material Twisted/Bent

    IT WAS REPORTED THAT THE TRAYS ARE WORN AND OLD. NEED TO BE REPLACED. PLASTIC INSTRUMENT HOLDERS IN TRAYS ARE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9506264 ↗
  • MalfunctionDec 20, 2019Crack; Peeled/Delaminated; Material Twisted/Bent; Crack; Peeled/Delaminated; Material Twisted/Bent

    IT WAS REPORTED THAT THE TRAYS ARE WORN AND OLD. NEED TO BE REPLACED. PLASTIC INSTRUMENT HOLDERS IN TRAYS ARE CRACKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9506258 ↗
  • MalfunctionOct 31, 2019Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9261983 ↗
  • MalfunctionOct 31, 2019Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9261967 ↗
  • MalfunctionOct 31, 2019Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9261965 ↗
  • MalfunctionOct 31, 2019Break; Naturally Worn; Break; Naturally Worn

    IT WAS REPORTED THAT THE HP SIGMA KNEE TRAYS ARE VERY WORN. THEY HAVE COATING ON RACKS DELAMINATING WHICH COULD CAUSE STERILIZATION ISSUES. DUE TO USE AND THE TIME SPENT IN THE AUTOCLAVE, THESE INSTRUMENTS TRAYS HAVE BECOME VERY RAGGED LOOKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9261963 ↗
  • MalfunctionNov 9, 2018Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8054925 ↗
  • MalfunctionNov 9, 2018Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8054920 ↗
  • MalfunctionNov 9, 2018Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem

    THE INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8054880 ↗
  • MalfunctionMar 22, 2018Delamination; Delamination

    IT WAS REPORTED THAT THE SIGMA TRAY WAS REPLACED DUE TO PEELING. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7363624 ↗
  • MalfunctionMar 22, 2018Delamination; Delamination

    IT WAS REPORTED THAT THE SIGMA TRAY WAS REPLACED DUE TO PEELING. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7363602 ↗

FDA clearance and classification

The FDA regulates product code FSM (tray, surgical, instrument) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SIGMA 9505-02-808

Has the SIGMA 9505-02-808 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295229360), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SIGMA 9505-02-808?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between Mar 22, 2018 and Jun 17, 2022. In those reports, the most common specific device problems are “Peeled/Delaminated” (9), “Naturally Worn” (6) and “Break” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SIGMA 9505-02-808?

None of the reports are classified as deaths, 0 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SIGMA 9505-02-808 FDA approved?

Devices of this type (product code FSM) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the SIGMA 9505-02-808, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand SIGMA, model 9505-02-808). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page