P.F.C. SIGMA 95-1032Recalls & reported problems
Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT ROTATING PLATFORM RP-F SIZE 3 15mm GVF · Class 3 · Implant
All models of P.F.C. SIGMA (487)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295230397), received between Jun 28, 2017 and Jan 19, 2022 — classified as 5 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information2
- Naturally Worn1
- Unintended Movement1
Patient problems
- No Code Available2
- Pain2
- Adhesion(s)1
- Joint Dislocation1
- Joint Disorder1
- Loss of Range of Motion1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJan 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CLINICAL ADVERSE EVENT RECEIVED FOR DECREASED RANGE OF MOTION : ADHESIONS EVENT IS NOT SERIOUS AND IS CONSIDERED MILD EVENT IS DEFINITELY NOT RELATED TO DEVICE AND IS DEFINITELY RELATED TO PROCEDURE. DATE OF IMPLANTATION: (B)(6) 2016. DATE OF EVENT (ONSET): (B)(6) 2016. (LEFT KNEE) TREATMENT: KNEE MANIPULATION UNDER ANESTHESIA.
Patient problems: Adhesion(s); Loss of Range of Motion
MAUDE report 13290581 ↗InjuryOct 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN DOR: (B)(6) 2020 LEFT KNEE.
Patient problems: No Code Available; Pain
MAUDE report 10621489 ↗InjuryJun 28, 2019Naturally Worn; Unintended Movement; Naturally Worn; Unintended Movement
THE PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2011 FOR THE PATIENT HAS PARKINSON'S DISEASE. IT WAS REPORTED THAT DISLOCATION OCCURRED AFTER 2 WEEKS FROM THE PRIMARY SURGERY. AFTER THAT, DISLOCATION HAD OCCURRED 1-2 TIMES A MONTH AND FREQUENT DISLOCATION HAD RECURRED RECENTLY. THE FEMUR WAS DISLOCATED TO THE ANTERIOR WHEN THE PATIENT STOOD UP, IT WAS CONFIRMED THAT THERE WAS A SUSPECTED THE TIBIAL INSERT SPINOUT AT X-RAY. THUS, THE REVISION SURGERY WAS PERFORMED BY REPLACING THE TIBIAL INSERT (P/N: 951032) TO ONE SIZE BIGGER 17.5MM TIBIAL INSERT ON (B)(6) 2019. THE SURGERY WAS COMPLETED WITHOUT A SURGICAL DELAY. IT WAS CONFIRMED THAT WEAR OF THE TIBIAL INSERT ANTERIOR LIP AND THE TIBIAL INSERT POST. POSSIBLE CAUSE: INSUFFICIENT REMOVAL OF TISSUE SUCH AS OUTER MENISCUS, INSUFFICIENT GAP MANAGEMENT AT THE PRIMARY SURGERY (BECAUSE THE EXTENSION GAP WAS LOOSE THAN FLEXION GAP). NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Joint Dislocation; No Code Available
MAUDE report 8743967 ↗InjuryApr 19, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
SUBJECT: (B)(6). CLINICAL ADVERSE EVENT RECEIVED FOR INSTABILITY. EVENT IS NOT SERIOUS AND IS CONSIDERED MODERATE. EVENT IS DEFINITELY RELATED TO DEVICE AND DEFINITELY NOT RELATED TO PROCEDURE. REVISION OF INSERT. DOI: (B)(6) 2005. DOE: (B)(6) 2017. DOR: (B)(6) 2018, (LEFT KNEE).
Patient problems: Joint Disorder
MAUDE report 7442582 ↗InjuryJun 28, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
DER STATED THAT THE PATIENT EXPERIENCING INSTABILITY AND PAIN. PATIENT HAD A FALL AND DAMAGED TO PATELLAR TENDON WHICH HAD BEEN SURGICALLY REPAIRED. UPON EXAM IN OR, THE MCL WAS FOUND INCOMPETENT. ALL PARTS WERE REMOVED AND PATIENT CONVERTED TO A HINGE. NO INSTRUMENT WERE BROKEN.
Patient problems: Pain
MAUDE report 6674995 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page