Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10603295234104), received between Nov 30, 2017 and Dec 14, 2022 — classified as 9 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Tear, Rip or Hole in Device Packaging10
- Adverse Event Without Identified Device or Use Problem6
- Loose or Intermittent Connection2
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Unspecified Infection4
- Pain3
- No Code Available2
- Not Applicable2
- No Consequences Or Impact To Patient1
- Tissue Damage1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
InjuryDec 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT TOTAL KNEE EXPLANT WAS PERFORMED DUE TO KNEE JOINT INFECTION. NO ADDITIONAL INFORMATION. DOI: (B)(6) 2014, DOR: (B)(6) 2022, AFFECTED SIDE : LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 15983796 ↗InjuryDec 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT TOTAL KNEE EXPLANT WAS PERFORMED DUE TO KNEE JOINT INFECTION. NO ADDITIONAL INFORMATION. DOI: (B)(6) 2014. DOR: (B)(6) 2022. AFFECTED SIDE : LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 15983792 ↗InjuryMar 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO AN INFECTED LEFT KNEE. EXTRACTION OF COMPONENTS AND INSERTION OF ANTIBIOTIC SPACER WERE DONE. NO ADDITIONAL INFORMATION. NO SURGICAL DELAY. DOI: (B)(6) 2018. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT KNEE AFFECTED.
Patient problems: Unspecified Infection
MAUDE report 13925417 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4). LATER ON (B)(6) 2022 , AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593846 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593844 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593831 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593819 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593805 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. THIS PC WAS CREATED TO CAPTURE THE PRODUCTS THAT ARE ALSO INVOLVE IN THE COMPLAINT REPORTED IN MAIN PC: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592305 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. THIS PC WAS CREATED TO CAPTURE THE PRODUCTS THAT ARE ALSO INVOLVE IN THE COMPLAINT REPORTED IN MAIN PC - (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592302 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. THIS PC WAS CREATED TO CAPTURE THE PRODUCTS THAT ARE ALSO INVOLVE IN THE COMPLAINT REPORTED IN MAIN (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592301 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. THIS PC WAS CREATED TO CAPTURE THE PRODUCTS THAT ARE ALSO INVOLVE IN THE COMPLAINT REPORTED IN MAIN PC - (B)(4) AND (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592299 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. THIS PC WAS CREATED TO CAPTURE THE PRODUCTS THAT ARE ALSO INVOLVE IN THE COMPLAINT REPORTED IN MAIN PC - (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592288 ↗InjuryOct 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED TO ADDRESS INFECTION. NO LOOSENING REPORTED. DOI: (B)(6) 2020; DOR: (B)(6) 2021; LEFT KNEE.
Patient problems: Unspecified Infection
MAUDE report 12719448 ↗InjuryDec 10, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryDec 10, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
InjuryOct 5, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
PATIENT WAS REVISED TO ADDRESS FALL, PAIN AND LOOSENING OF THE FEMUR TO FEMORAL COMPONENT AND BOLT TO FEMORAL ADAPTOR AT THE BONE TO CEMENT AND CEMENT TO IMPLANT.
Patient problems: Not Applicable; Pain
MAUDE report 7940257 ↗InjuryAug 7, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
PATIENT WAS REVISED TO ADDRESS METALLOSIS AND LOOSENING OF THE STEM BOLT AND FEMUR AT THE CEMENT TO IMPLANT INTERFACE. UNKNOWN CEMENT WAS USED. IT WAS ALSO REPORTED THAT THE PATIENT UNDERWENT POLY SWAP SINCE DOING THE REVISION. DOI: (B)(6) 2001; DOR: (B)(6) 2018; RIGHT KNEE.
Patient problems: Not Applicable; Tissue Damage
MAUDE report 7756905 ↗InjuryNov 30, 2017Malposition of Device; Malposition of Device
PATIENT INITIATED COMPLAINT; MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR MDR REPORTABILITY, THE PATIENT WAS REVISED FOR TIBIAL MALPOSITIONING, POLY WEAR, METALLOSIS, AND FEMORAL LOOSENING AT AN UNKNOWN INTERFACE/CEMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7071445 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page