Skip to content

P.F.C. SIGMA 96-0886Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA POSTERIOR AUGMENT COMBO CEMENTED SIZE 4 4mm · Class 3 · Implant

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy P.F.C. Sigma posterior augment is a cemented spacer used in revision knee replacement surgery. The 50 reports mostly describe knee pain, swelling, looseness of the implant, and further surgery to remove or swap components, with metal wear debris and infection also mentioned. There are 49 injuries and no deaths.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
48
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 10603295234401), received between Dec 14, 2017 and Jul 1, 2026 — classified as 48 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem34
  • Loss of or Failure to Bond10
  • Insufficient Device Problem Information2
  • Loss of Osseointegration2
  • Difficult to Insert1

Patient problems

  • Pain19
  • No Code Available13
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available7
  • Edema7
  • Joint Laxity6
  • Insufficient Information5
  • Unspecified Infection5
  • Adhesion(s)4

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT DEVELOPED A SEVERE INFECTION. SO, ALL COMPONENTS WERE REMOVED AND AN ANTIBIOTIC NAIL WAS PUT IN. DOI: (B)(6) 2019. DOR: (B)(6) 2026 AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 25678188 ↗
  • InjuryJan 12, 2026Loss of or Failure to Bond

    IT WAS REPORTED THAT A FEMORAL REVISION PERFORMED DUE TO LOOSENING OF THE FEMORAL COMPARTMENT. LOOSING OCCURRED AT THE FEMORAL ADAPTER AND SLEEVE JUNCTION. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2016. DOE: (B)(6) 2025. AFFECTED SIDE: RIGHT.

    Patient problems: Insufficient Information

    MAUDE report 24039354 ↗
  • InjuryNov 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT HAD DEVELOPED PAINFUL LAXITY OVER TIME. THE COMPONENTS APPEARED TO BE WELL FIXED. A REVISION KNEE WAS SCHEDULED TO ADDRESS LAXITY. DURING THE OPERATION, IT WAS NOTED THAT THERE WAS METAL DEBRIS BEING GENERATED BY THE PROSTHESIS AS THERE WAS STAINING THROUGHOUT THE CAPSULE. THE FEMORAL COMPONENT WAS REMOVED, AND IT APPEARED THAT THE DEBRIS WAS BEING GENERATED AT THE TAPER BETWEEN THE FEMORAL ADAPTER AND THE FEMORAL SLEEVE; BOTH OF WHICH WERE WELL FIXED. THE WELL-FIXED TIBIAL TRAY WAS REMOVED. TRIALS WERE PLACED AND MOTION AND STABILITY WAS CONFIRMED. FINAL COMPONENTS WERE PLACED. NO PATIENT HARM NOR DELAY WAS NOTED. PATIENT APPEARED HAVE HAD THICKER INSERTS PLACED ON (B)(6) 2020 AND (B)(6) 2021. DOI: (B)(6) 2017. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 23543615 ↗
  • InjuryNov 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT HAD DEVELOPED PAINFUL LAXITY OVER TIME. THE COMPONENTS APPEARED TO BE WELL FIXED. A REVISION KNEE WAS SCHEDULED TO ADDRESS LAXITY. DURING THE OPERATION, IT WAS NOTED THAT THERE WAS METAL DEBRIS BEING GENERATED BY THE PROSTHESIS AS THERE WAS STAINING THROUGHOUT THE CAPSULE. THE FEMORAL COMPONENT WAS REMOVED, AND IT APPEARED THAT THE DEBRIS WAS BEING GENERATED AT THE TAPER BETWEEN THE FEMORAL ADAPTER AND THE FEMORAL SLEEVE; BOTH OF WHICH WERE WELL FIXED. THE WELL-FIXED TIBIAL TRAY WAS REMOVED. TRIALS WERE PLACED AND MOTION AND STABILITY WAS CONFIRMED. FINAL COMPONENTS WERE PLACED. NO PATIENT HARM NOR DELAY WAS NOTED. PATIENT APPEARED HAVE HAD THICKER INSERTS PLACED ON (B)(6) 2020 AND (B)(6) 2021. DOI: (B)(6) 2017. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 23543421 ↗
  • InjuryJul 11, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    PATIENT WAS REVISED DUE TO LOOSENING OF FEMORAL COMPONENT WAS DUE TO BONE LOSS. CONVERTED TO A DFR. DOI: UNKNOWN, DOR: (B)(6) 2025, AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Bone Notching

    MAUDE report 22484502 ↗
  • InjuryApr 23, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO POSSIBLE ASEPTIC LOOSENING. REMOVED ALL IMPLANTS AND CONVERTED TO ATTUNE REVISION KNEE IMPLANTS. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 21892748 ↗
  • InjuryNov 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED, THAT PATIENT HAD LOSS OF MOBILITY, DUE TO SUSPECTED HETEROTOPIC OSSIFICATION. THE PATIENT HAD A DEPUY TC3 RP REVISION KNEE. THE SZ 4 TC3 RIGHT FEMUR, 5 DEGREE BOLT, +2/-2 FEMORAL ADAPTOR, TC3 SZ 4 X 4 DISTAL AUGMENTS X2, TC3 SZ 4 X 4 POSTERIOR AUGMENT, 115 X 16 STEM AND TC3 RP SZ 4 X 12.5 INSERT WERE EXPLANTED. DOI: (B)(6) 2016. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Extraskeletal ossification; Loss of Range of Motion

    MAUDE report 20689266 ↗
  • InjuryJan 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO NICKEL ALLERGY. DOI: UNKNOWN. DOR: (B)(6), 2024. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 18570289 ↗
  • InjuryJan 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO NICKEL ALLERGY. DOI: UNKNOWN. DOR: JAN 17, 2024. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 18570213 ↗
  • InjuryNov 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD AN UNDISCOVERED NICKEL ALLERGY. THEY HAD PAIN THROUGHOUT THEIR LEG, SO THE SURGEON REPLACED EVERYTHING WITH COMPETITOR IMPLANTS. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE : UNKNOWN.

    Patient problems: Hypersensitivity/Allergic reaction; Pain

    MAUDE report 18184873 ↗
  • InjuryOct 25, 2023Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT WAS REVISED DUE TO LOOSENING OF THE FEMORAL COMPONENT AT THE UNKNOWN INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. REVISED FEMUR AND TIBIAL COMPONENTS. DOI: (B)(6) 2018. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Insufficient Information

    MAUDE report 18002541 ↗
  • InjuryAug 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS EXPLANTED TC3 CONSTRUCT DUE TO NICKEL ALLERGY. DOI: (B)(6), 2020, DOR: (B)(6), 2023, AFFECTED SIDE: LEFT KNEE.

    Patient problems: Hypersensitivity/Allergic reaction

    MAUDE report 17488032 ↗
  • InjuryJul 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENT RECEIVED FOR SEVERE PAIN. EVENT IS NOT SERIOUS. DATE OF IMPLANT: (B)(6) 2010. DATE OF REVISION: (B)(6) 2013. DATE OF EVENT: (B)(6) 2011. (LEFT KNEE). TREATMENT: REVISION; FEMORAL COMPONENT, TIBIAL BASE, AND INSERT WERE REVISED.

    Patient problems: Pain

    MAUDE report 17412630 ↗
  • InjuryJul 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENT RECEIVED FOR SEVERE PAIN. EVENT IS NOT SERIOUS. DATE OF IMPLANT: (B)(6) 2010, DATE OF REVISION: (B)(6) 2013, DATE OF EVENT: (B)(6) 2011. (LEFT KNEE). TREATMENT: REVISION; FEMORAL COMPONENT, TIBIAL BASE, AND INSERT WERE REVISED.

    Patient problems: Pain

    MAUDE report 17412587 ↗
  • InjuryJun 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL NOTIFICATION RECEIVED FOR REVISION DUE TO INSTABILITY. DATE OF IMPLANT: ON (B)(6) 2017. DATE OF REVISION: ON (B)(6) 2023 (LEFT KNEE). TREATMENT: FEMORAL COMPONENT AND INSERT WERE REVISED.

    Patient problems: Joint Laxity; Pain; Wound Dehiscence

    MAUDE report 17180746 ↗
  • InjuryJun 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO KNEE INSTABILITY. DOI: (B)(6) 2017. DOR: (B)(6) 2023. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Joint Laxity

    MAUDE report 17048111 ↗
  • InjuryDec 22, 2022Loss of or Failure to Bond; Loss of or Failure to Bond

    THE PATIENT WAS REVISED DUE TO LOOSENING OF THE UNKNOWN COMPONENT AT THE UNKNOWN INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. IT WAS ALSO REPORTED THAT THE PATIENT BROKE THE ADAPTOR AND ADAPTOR BOLT FOR THE SECOND TIME. BONE HAS RESORBED AROUND THE IMPLANT CREATING ALMOST A DISTAL FEMORAL REPLACEMENT SITUATION. ADAPTOR BOLT, ADAPTOR, FEMORAL COMPONENT AND INSERT WERE REMOVED LEAVING THE TIBIAL COMPONENT AND FEMORAL SLEEVE AND STEM WELL FIXED IN THE PATIENT. A HINGED COMPONENT WAS THEN PREPPED FOR AND IMPACTED INTO THE SLEEVE THAT WAS WELL FIXED IN THE BONE AND A LPS INSERT WAS INSERTED. DOI: (B)(6) 2017. DOR: (B)(6) 2022. RIGHT KNEE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16033314 ↗
  • InjuryDec 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SURGEON STATED PATIENT WAS UNSTABLE. THE SURGEON REMOVED THE FEMORAL COMPONENTS/POLY AND REPLACED FEMORAL COMPONENTS/POLY. THERE WAS NO REPORTED LOOSENING. THERE WAS NO REPORTED SURGICAL DELAY. DOI: (B)(6) 2018. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Joint Laxity

    MAUDE report 15908975 ↗
  • InjurySep 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT'S KNEE WAS REVISED DUE TO INFECTION AND LOOSENING OF THE FEMORAL & TIBIAL COMPONENT AT THE BONE TO CEMENT INTERFACE. UNKNOWN CEMENT WAS USED. REMOVAL OF ALL IMPLANTS DUE TO CEMENT COULDN'T SEE MOST NUMBERS. ONLY ABLE TO OBTAIN THE NUMBERS FROM FEMORAL STEM, TC3 RP INSERT AND THE THICK TRAY. ALL OTHER PARTS WERE COVERED BY CEMENT. NO FURTHER INFORMATION IS AVAILABLE. DOI: (B)(6) 2015. DOR: (B)(6) 2021. RIGHT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12517254 ↗
  • InjurySep 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT'S KNEE WAS REVISED DUE TO INFECTION AND LOOSENING OF THE FEMORAL & TIBIAL COMPONENT AT THE BONE TO CEMENT INTERFACE. UNKNOWN CEMENT WAS USED. REMOVAL OF ALL IMPLANTS DUE TO CEMENT COULDN'T SEE MOST NUMBERS. ONLY ABLE TO OBTAIN THE NUMBERS FROM FEMORAL STEM, TC3 RP INSERT AND THE THICK TRAY. ALL OTHER PARTS WERE COVERED BY CEMENT. NO FURTHER INFORMATION IS AVAILABLE. DOI: (B)(6) 2015 DOR: (B)(6) 2021 RIGHT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12517248 ↗
  • InjurySep 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENT RECEIVED FOR BONE FRACTURE OF MEDIAL AND LATERAL FEMORAL CONDLYLES: ABNORMAL RADIOGRAPHIC EVALUATION - LIKELY FEMORAL COMPONENT LOOSENING (UNCONFIRMED), WITH FRACTURE OF THE MEDIAL AND LATERAL CONDYLES, AND INSTABILITY EVENT IS SERIOUS AND IS CONSIDERED SEVERE EVENT IS DEFINITELY RELATED TO DEVICE AND IS DEFINITELY NOT RELATED TO PROCEDURE. DATE OF IMPLANTATION: (B)(6) 2014. DATE OF REVISION: 03/31/2017 (ALL COMPONENTS). DATE OF EVENT (ONSET): (B)(6) 2021. (LEFT KNEE). TREATMENT: REVISION OF THE FEMORAL AND INSERT COMPONENTS OCCURRED ON 17/AUG/2021

    Patient problems: Fall; Joint Laxity; Limb Fracture; Pain; Swelling/ Edema

    MAUDE report 12415386 ↗
  • InjuryAug 18, 2021Loss of or Failure to Bond; Loss of or Failure to Bond

    ORIGINAL OPERATION WAS PERFORMED (B)(6) 2017. TIBIA WAS LEFT IN THE PATIENT. SURGEON REPLACED THE IMPLANTS WITH ANOTHER TC3 FEMUR, SLEEVE AND A BIGGER PRESSFIT STEM. THE PROCEDURE TOOK 3 ½ HOURS. AFFECTED SIDE: LEFT.

    Patient problems: Insufficient Information

    MAUDE report 12337181 ↗
  • InjuryAug 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTED TKA. REMOVED IMPLANTS THAT WERE IMPLANTED (B)(6) 2020. PERFORMED WASHOUT, I&D AND IMPLANTED A ANTIBIOTIC CEMENT SPACER. DOI: (B)(6) 2020; DOR: (B)(6) 2021 LEFT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12296636 ↗
  • InjuryJun 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A TOTAL IMPLANT EXPLANTED, CONVERT TO ANTIBIOTIC SPACER DUE TO INFECTION. POLY INSERT CHANGED ON (B)(6) 2019 FROM PREVIOUS INFECTION. DOI: (B)(6) 2019 ALL EXPLANTED PRODUCTS, DOI: (B)(6) 2019 TIBIAL INSERT, DOR: (B)(6) 2021, RIGHT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11917903 ↗
  • InjuryJun 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A TOTAL IMPLANT EXPLANTED, CONVERT TO ANTIBIOTIC SPACER DUE TO INFECTION. POLY INSERT CHANGED ON (B)(6) 2019 FROM PREVIOUS INFECTION. DOI: (B)(6) 2019 ALL EXPLANTED PRODUCTS, DOI: (B)(6) 2019 TIBIAL INSERT, DOR: (B)(6) 2021, RIGHT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11917900 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →