P.F.C. SIGMA 96-0911Recalls & reported problems
Depuy (Ireland) · P.F.C. SIGMA DISTAL AUGMENT COMBO SIZE 5 16mm LEFT · Class 3 · Implant
All models of P.F.C. SIGMA (487)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10603295234548), received between Feb 24, 2022 and Feb 24, 2022 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Tear, Rip or Hole in Device Packaging14
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 (B)(4) PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4). ON (B)(6) 2022, AN ADDITIONAL (B)(4) PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4). LATER ON (B)(6) 2022, AN ADDITIONAL (B)(4) PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593824 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-001042377). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593823 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6)2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593814 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593812 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593767 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593753 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)). ON 10-JAN-2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13593740 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592493 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592491 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592489 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592488 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592485 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN (B)(4)¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592484 ↗MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13592472 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page