P.F.C. SIGMA 96-0911Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA DISTAL AUGMENT COMBO SIZE 5 16mm LEFT · Class 3 · Implant

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10603295234548), received between Feb 24, 2022 and Feb 24, 2022 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Tear, Rip or Hole in Device Packaging14

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 (B)(4) PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4). ON (B)(6) 2022, AN ADDITIONAL (B)(4) PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4). LATER ON (B)(6) 2022, AN ADDITIONAL (B)(4) PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593824 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-001042377). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593823 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6)2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593814 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593812 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593767 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). ON (B)(6) 2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON (B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593753 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT ON (B)(6) 2022, A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS. ON (B)(6) 2021 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)). ON 10-JAN-2022, AN ADDITIONAL 12 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)). LATER ON (B)(6) 2022, AN ADDITIONAL 29 PARTS WERE DISCOVERED WITH DEFECTS (CAPTURED ON PC-(B)(4)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13593740 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592493 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592491 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592489 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592488 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592485 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN (B)(4)¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592484 ↗
  • MalfunctionFeb 24, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A CAST PRODUCT CLEANROOM (CPCR) 2ND SHIFT FINISH OPERATOR WAS PERFORMING THE WI-0897 ¿REPACKAGING / RESTERILIZATION OF STERILE PRODUCT IN RAYNHAM¿ FOR A PFC SIGMA AUGMENT REPACKAGING ORDER. DURING THIS PROCESS, THE OPERATOR IDENTIFIED MULTIPLE PARTS DISPLAYED A VISUAL COSMETIC DEFECT ON THE TOP SURFACE. THE OPERATOR NOTIFIED THE FINISH LEAD, WHO CONTACTED PROCESS AND QUALITY ENGINEERING FOR FURTHER EVALUATION. THE IMAGE ATTACHED SHOWS A AUGMENT RETURNED FROM EXPIRY WITH FOAM PARTICLES TRANSFERRED FROM THE PACKAGING COMPONENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592472 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →