Skip to content

P.F.C. SIGMA 96-2043Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT FIXED BEARING CURVED 4 (CVD) 15mm GVF · Class 3 · Implant

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295240839), received between Aug 30, 2017 and Oct 13, 2023 — classified as 9 injuries.

The most frequently reported specific device problem is “Naturally Worn” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Naturally Worn3
  • Fracture2

Patient problems

  • No Code Available2
  • Pain2
  • Unspecified Infection2
  • Fall1
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • Joint Laxity1
  • Loss of Range of Motion1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 13, 2023Naturally Worn; Naturally Worn

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: Insufficient Information

    MAUDE report 17931646 ↗
  • InjuryApr 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE POLY SWAP DUE TO INFECTION OF KNEE JOINT. NO ADDITIONAL INFORMATION. DOI: (B)(6) 2013, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 14083286 ↗
  • InjuryApr 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO INFECTION OF KNEE JOINT AND LOOSE FEMUR. DOI: (B)(6) 2013. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 14046263 ↗
  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT'S KNEE WAS REVISED TO ADDRESS PAIN. EXPLANTED 4X12.5 CR/RP TIBIAL INSERT AND 38 MM OVAL DOME PATELLA. NO LOOSENING ON FEMUR OR TIBIAL COMPONENT. I HAVE NO FURTHER INFORMATION. NO LOT NUMBERS . (B)(6) OFFICE HAS PAPER WORK FROM 2016. DOI: (B)(6) 2016. DOR: (B)(6) 2021; LEFT KNEE.

    Patient problems: Pain

    MAUDE report 12194049 ↗
  • InjuryJun 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6)2013, THE PATIENT UNDERWENT BILATERAL KNEE ARTHROPLASTIES TO ADDRESS OSTEOARTHRITIS. THE PATELLA WAS RESURFACED AND DEPUY PRODUCTS WERE IMPLANTED BILATERALLY. THERE WERE NO INDICATED INTRA-OPERATIVE COMPLICATIONS. ON (B)(6) 2021, THE PATIENT UNDERWENT A LEFT KNEE REVISION TO ADDRESS PAIN, SWELLING, AND INSTABILITY. THE PATIENT EXPERIENCED A FALL PRIOR TO EXPERIENCING INSTABILITY. THE SURGEON INDICATED EVIDENCE OF RECURRENT BLEEDS AS THERE WAS HEMOSIDERIN STAINED SYNOVITIS WHICH WAS RESECTED. THERE WAS A LITTLE WEAR ON THE RETAINED PATELLA, BUT NO DELAMINATION. THERE WAS MENISCUS AROUND THE PATELLA THAT WAS EXCISED. THE TIBIAL INSERT WAS THE ONLY PRODUCT REVISED. THE TIBIAL INSERT, FEMORAL COMPONENT, AND PATELLAR COMPONENT WERE RETAINED. THERE WERE NO INDICATED INTRA-OPERATIVE COMPLICATIONS.

    Patient problems: Fall; Hemorrhage/Blood Loss/Bleeding; Joint Laxity; Pain; Swelling/ Edema; Synovitis

    MAUDE report 12010318 ↗
  • InjuryOct 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INSTABILITY. DOI: UNKNOWN; DOR: (B)(6) 2019; LEFT KNEE.

    Patient problems: No Code Available

    MAUDE report 9193943 ↗
  • InjuryMay 8, 2018Naturally Worn; Naturally Worn

    PATIENT WAS REVISED TO ADDRESS POLY WEAR. IT WAS ALSO REPORTED THAT THE SURGEON REPLACED WITH A THICKER POLY. DOI: (B)(6) 2002; DOR: (B)(6) 2018; LEFT KNEE.

    Patient problems: Not Applicable

    MAUDE report 7493782 ↗
  • InjuryNov 3, 2017Fracture; Naturally Worn; Fracture; Naturally Worn

    CLINICAL DER STATES THAT THE PATIENT WAS REVISED TO ADDRESS INSTABILITY AND FRACTURED DEPUY INSERT.

    Patient problems: No Code Available

    MAUDE report 6998797 ↗
  • InjuryAug 30, 2017Fracture; Fracture

    IT WAS INDICATED THAT POLYETHYLENE FRACTURE OF 962043, 4X15 RP CR INSERT SECONDARY TO INSTABILITY REPLACED WITH A 4X17.5 RP CR TIBIAL INSERT. UPDATE 8-1-2017: MEDICAL RECORDS HAVE BEEN REVIEWED. OPERATIVE NOTE (B)(6) 2009 INDICATES THE PATIENT RECEIVED A RIGHT TOTAL KNEE REPLACEMENT DUE TO DEGENERATIVE JOINT DISEASE OF THE RIGHT KNEE. THE PROCEDURE WAS COMPLETED WITHOUT ANY COMPLICATION NOTED BY THE SURGEON.

    Patient problems: Loss of Range of Motion

    MAUDE report 6830980 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →

Sources and updates