Skip to content

P.F.C. SIGMA 96-2145Recalls & reported problems

Depuy (Ireland) · P.F.C. SIGMA TIBIAL INSERT ROTATING PLATFORM STABILIZED 4 (STAB) 20mm GVF · Class 3 · Implant

All models of P.F.C. SIGMA (487)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295241355), received between Sep 4, 2019 and Nov 13, 2025 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Pain2
  • Joint Disorder1
  • Joint Laxity1
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT HAD DEVELOPED PAINFUL LAXITY OVER TIME. THE COMPONENTS APPEARED TO BE WELL FIXED. A REVISION KNEE WAS SCHEDULED TO ADDRESS LAXITY. DURING THE OPERATION, IT WAS NOTED THAT THERE WAS METAL DEBRIS BEING GENERATED BY THE PROSTHESIS AS THERE WAS STAINING THROUGHOUT THE CAPSULE. THE FEMORAL COMPONENT WAS REMOVED, AND IT APPEARED THAT THE DEBRIS WAS BEING GENERATED AT THE TAPER BETWEEN THE FEMORAL ADAPTER AND THE FEMORAL SLEEVE; BOTH OF WHICH WERE WELL FIXED. THE WELL-FIXED TIBIAL TRAY WAS REMOVED. TRIALS WERE PLACED AND MOTION AND STABILITY WAS CONFIRMED. FINAL COMPONENTS WERE PLACED. NO PATIENT HARM NOR DELAY WAS NOTED. PATIENT APPEARED HAVE HAD THICKER INSERTS PLACED ON (B)(6) 2020 AND (B)(6) 2021. DOI: (B)(6) 2017. DOR: (B)(6) 2025. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 23543500 ↗
  • InjurySep 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED WITH A PS PFC / MBT REVISION IN SITU FROM A PREVIOUS REVISION OF A HOWMEDICA PRIMARY KNEE (PRIMARY FROM 1998, 1ST REVISION NOVEMBER 2007), SYMPTOMS OF PAIN, FIXED FLEXION DEFORMITY AND PATELLA BAHA. DR INGLIS PERFORMED SYNOVECTOMY, REMOVED CALCIFICATION FORM PATELLA TENDON AND EXCHANGED POLY BEARING, DOWNSIZED FROM 20MM IMPLANTED IN 2007 TO A 17.5MM.

    Patient problems: Joint Disorder; No Code Available; Pain

    MAUDE report 8958820 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page