Skip to content

SIGMA 96-2355Recalls & reported problems

Depuy (Ireland) · SIGMA TIBIAL INSERT ROTATING PLATFORM TC3 SIZE 4 20mm GVF · Class 3 · Implant

Quick summary

This DePuy SIGMA rotating-platform knee implant (tibial insert, size 4) is used in total knee replacement. The 31 reports mostly describe pain, infection, and knee instability or looseness that led to revision surgery, with the insert or femoral component being removed or exchanged; there are 31 injuries and no deaths. Most reports name no specific device problem. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
31
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 10603295242413), received between Jan 10, 2018 and Aug 4, 2026 — classified as 31 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem27
  • Insufficient Device Problem Information2
  • Device Dislodged or Dislocated1
  • Malposition of Device1

Patient problems

  • Pain11
  • Unspecified Infection9
  • No Code Available7
  • Joint Laxity5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Discomfort2
  • Muscular Rigidity2
  • Swelling2

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that there was ligamentuos instability.DOI: (b)(6) 2024. DOR: (b)(6) 2026.Affected Side: Left.

    Patient problems: Joint Laxity

    MAUDE report 50133391 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2025. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 25457474 ↗
  • InjuryDec 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS POLY EXCHANGES FOR INFECTION. NO HARM TO PATIENT. NO BROKEN INSTRUMENT. NO DELAYS. DOI: (B)(6) 2020. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT.

    Patient problems: Unspecified Infection

    MAUDE report 20912015 ↗
  • InjuryMar 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED: 1. THERE WAS NO SURGICAL DELAY. 2. THERE WERE NO ADVERSE CONSEQUENCE/S THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 18970481 ↗
  • InjuryDec 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO RANGE OF MOTION. IT WAS UNKNOWN ABOUT SURGICAL DELAY. DOI: UNKNOWN; DOR: (B)(6) 2023; AFFECTED SIDE: UNKNOWN.

    Patient problems: Loss of Range of Motion

    MAUDE report 18372679 ↗
  • InjuryJun 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL NOTIFICATION RECEIVED FOR REVISION DUE TO INSTABILITY. DATE OF IMPLANT: (B)(6) 2017. DATE OF REVISION: (B)(6) 2023. (LEFT KNEE). TREATMENT: FEMORAL COMPONENT AND INSERT WERE REVISED.

    Patient problems: Joint Laxity; Pain; Wound Dehiscence

    MAUDE report 17058197 ↗
  • InjuryJun 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO KNEE INSTABILITY. DOI: (B)(6) 2017; DOR: (B)(6) 2023; AFFECTED SIDE: LEFT KNEE.

    Patient problems: Joint Laxity

    MAUDE report 17044206 ↗
  • InjuryMay 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE INSERT WAS TAKEN OUT DURING REVISION SURGERY FOR INFECTION. ALL PIECES RETRIEVED. SURGEON DOES NOT WISH TO PROVIDE ADDITIONAL INFORMATION. DOI: (B)(6) 2022. DOR: (B)(6) 2023. AFFECTED SIDE: UNKNOWN KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 16938005 ↗
  • InjuryMar 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD PAIN AND STERLING AND WAS TESTED FOR NICKEL ALLERGY. THE TEST CAME BACK POSITIVE SO THE SURGEON TOOK IT ALL OUT AND PUT IN AN AESCULAP REVISION CONSTRUCT. DOI: UNKNOWN. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT KNEE.

    Patient problems: Pain

    MAUDE report 16585077 ↗
  • InjuryJan 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SWELLING AND REDNESS IN KNEE AREA. SURGERY SCHEDULED (I AND D) WITH A POLY SWAP. DOI: (B)(6) 2018, DOR: (B)(6) 2022, AFFECTED SIDE: LEFT KNEE.

    Patient problems: Erythema; Swelling/ Edema

    MAUDE report 16140817 ↗
  • InjuryDec 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    TC3 REVISION KNEE? INFECTED - UNSTABLE SIGNASURE CAME BACK NEGATIVE FOR INFECTION PREVIOUS IV DRUG ABUSER, HIGH FUNCTIONING KNEE PLAN A- ATTUNE REVISION, PLAN B - ATTUNE HINGE PLAN C - TC3 FEMUR REVISION IF TIBIA WELL FIXED IN SITU - TC3FEMUR SIZE 4 PLUS DISTAL AND POSTERIOR AUGMENTS, STEM AND ADAPTOR MBT TIBIA REVISION SIZE 3, WEDGE, STEM SIZE 4 20MM INSERT TC3 RP PATELLA 41MM 3 PEG.

    Patient problems: Joint Laxity

    MAUDE report 16017385 ↗
  • InjuryMay 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    EXPLORATION RIGHT TC3 REVISION TKR FOR PAIN. PATIENT UNDERWENT REVISION TO TC3 TKR ON (B)(6) 2011. PRESENTED BACK TO SURGEON COMPLAINING OF PAIN. SURGEON UNSURE OF CAUSE OF PAIN, SCAN HAD SHOWN A 'HOTSPOT' AROUND THE PATELLA. PROCEDURE TODAY WAS REVISION OF PATELLA COMPONENT AND POLY EXCHANGE RIGHT TC3 REVISION TOTAL KNEE REPLACEMENT. SURGEON STATED HE THOUGHT THE PATELLA COMPONENT 'MAY HAVE BEEN LOOSE'. POLY WAS UPSIZED FROM 20MM TO 22.5MM.-- REVISION OF PATELLA COMPONENT AND POLY EXCHANGE RIGHT TC3 TKR, DOI: (B)(6), 2011. DOR: (B)(6), 2022. SIDE: RIGHT KNEE.

    Patient problems: Pain

    MAUDE report 14315312 ↗
  • InjuryMar 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A POLY SWAP AND I&D DUE TO INFECTION. NO SURGICAL DELAY. DOI: (B)(6) 2021; DOR: (B)(6) 2022; AFFECTED SIDE: LEFT KNEE.

    Patient problems: Unspecified Infection

    MAUDE report 13652067 ↗
  • InjuryFeb 18, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    REVISION COMPLETED ON RIGHT KNEE SECONDARY TO SEVERAL DISLOCATIONS. SURGEON DISCOVERED THE TIBIAL INSERT WAS ROTATED 180 DEGREES POSTERIORLY. SURGEON NOTED LIGAMENTOUS DEFICIENCIES AND ELECTED TO CONVERT THE KNEE CONSTRUCT TO A HINGED FEMORAL COMPONENT. THE FEMORAL HINGE COMPONENT AND TIBIAL INSERT WERE EXCHANGED. THE INDEX FEMORAL SLEEVE AND STEM, THE TIBIAL CONSTRUCT, AND THE PATELLA WERE KEPT. SURGEON STATED HE DID NOT THINK THE COMPONENTS FAILED TO MEET EXPECTATIONS.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 13559683 ↗
  • InjuryNov 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(6). CLINICAL NOTIFICATION RECEIVED FOR REVISION DUE TO INFECTION. TIBIAL INSERT EXCHANGE ONLY. DATE OF IMPLANT: (B)(6) 2012, DATE OF REVISION: (B)(6) 2021, (LEFT KNEE).

    Patient problems: Joint Contracture; Pain; Sepsis; Swelling/ Edema; Unspecified Infection; Unspecified Musculoskeletal problem

    MAUDE report 12770083 ↗
  • InjuryAug 31, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTION. DOI: (B)(6) 2021, DOR: (B)(6) 2021, LEFT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12397167 ↗
  • InjuryAug 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO INFECTED TKA. REMOVED IMPLANTS THAT WERE IMPLANTED (B)(6) 2020. PERFORMED WASHOUT, I&D AND IMPLANTED A ANTIBIOTIC CEMENT SPACER. DOI: (B)(6) 2020 DOR: (B)(6) 2021 LEFT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12296662 ↗
  • InjuryJun 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS INFECTION. A WASHOUT WAS DONE AND A NEW INSERT OF THE SAME SIZE WAS REIMPLANTED. NO OTHER INFORMATION REGARDING THE PATIENT IS AVAILABLE. DOI: (B)(6) 2015, DOR: (B)(6) 2021, LEFT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11986298 ↗
  • InjuryMay 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PAIN AND POSSIBLE ROTATION OF IMPLANTS. NO LOOSENING WAS REPORTED. DOI: (B)(6) 2019. DOR: (B)(6) 2021. LEFT KNEE.

    Patient problems: Pain

    MAUDE report 11804837 ↗
  • InjuryFeb 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PAIN. DOI: (B)(6) 2021, DOR: (B)(6) 2021, RIGHT KNEE.

    Patient problems: Pain

    MAUDE report 11365705 ↗
  • InjuryJan 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS INFECTION. POLY SWAP WAS DONE. NO SURGICAL DELAY. DOI: UNKNOWN, DOR: (B)(6) 2020, LEFT KNEE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11131485 ↗
  • InjurySep 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2020, THE PATIENT UNDERWENT A LEFT KNEE REVISION DUE TO INSTABILITY, EFFUSION, PAIN, DISCOMFORT, FEMORAL COMPONENT MIGRATION AND LOOSENING AT AN UNKNOWN INTERFACE, TIBIAL TRAY LOOSENING AT AN UNKNOWN INTERFACE, AND NO INGROWTH OF THE TIBIAL OR FEMORAL SLEEVES. THE UNKNOWN MANUFACTURER PATELLAR COMPONENT WAS RETAINED. ALL OTHER COMPONENTS WERE REVISED. THERE WERE NO INDICATED INTRA-OPERATIVE COMPLICATIONS.

    Patient problems: Discomfort; Edema; No Code Available; Pain

    MAUDE report 10600347 ↗
  • InjuryJul 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED TO ADDRESS PAIN AND DISLOCATION OF THE PATELLA. THE SURGEON DEBATED BETWEEN PUTTING A THICKER POLY OR REVISING THE FEMUR TO CHANGE THE JOINT LINE. THE SURGEON INCREASED THE POLY BY 2 SIZES AND WAS HAPPY WITH THE STABILITY AND TRACKING OF THE PATELLA WITH THE THICKER INSERT. ORIGINAL IMPLANTATION DATE IS UNKNOWN. DOI: UNKNOWN, DOR: (B)(6) 2020.

    Patient problems: No Code Available; Pain

    MAUDE report 10296758 ↗
  • InjuryJun 29, 2020Malposition of Device; Malposition of Device

    CLINICAL ADVERSE EVENT RECEIVED FOR LOOSENING OF BOTH FEMORAL AND TIBIAL COMPONENTS : ABNORMAL RADIOGRAPHIC EVALUATION. EVENT IS SERIOUS AND IS CONSIDERED MODERATE. EVENT IS DEFINITELY RELATED TO DEVICE AND POSSIBLY RELATED TO PROCEDURE.

    Patient problems: Bone Fracture(s); Injury; Muscular Rigidity; No Code Available; Pain; Swelling; Synovitis

    MAUDE report 10207362 ↗
  • InjuryNov 25, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CAME IN WITH AN UNSTABLE KNEE. THE SURGEON REMOVED THE 4X20 TC3 INSERT AND REPLACED WITH A 4X25 MM POLY TO HELP CORRECT PATIENT INSTABILITY AND GIVE THEM BETTER RANGE OF MOTION.

    Patient problems: Joint Disorder; Limited Mobility Of The Implanted Joint; No Code Available

    MAUDE report 9369422 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.