CORAIL L20416Recalls & reported problems

Depuy (Ireland) · CORAIL AMT BROACH 16 · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10603295258186), received between Nov 7, 2017 and Jun 17, 2025 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility8
  • Scratched Material3
  • Contamination /Decontamination Problem2
  • Material Deformation2
  • Material Twisted/Bent2
  • Break1
  • Difficult to Insert1
  • Illegible Information1

Patient problems

  • No Consequences Or Impact To Patient9
  • No Clinical Signs, Symptoms or Conditions5
  • No Code Available1
  • No Information1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 17, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE SIZE 16 BROACH DAMAGED DURING TRIALLING. THE TIP OF BROACH WAS HIT WITH A MALLET ACCIDENTALLY, DEFORMING THE TIP OF BROACH, RESULTING IN A TRIAL NECK NOT FITTING. THERE WAS A 3-4 MINUTES OF SURGICAL DELAY. A NEW TRAY OF BROACHES WAS OPENED, THE DAMAGES BROACH REMOVED AND REPLACED WITH ANOTHER OF THE SAME SIZE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22236204 ↗
  • MalfunctionApr 5, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SURGEON BROACHED USING CORAIL 8-16 CALCAR REAMED OVER A 16 WAS THEN UNABLE TO FIT THE NECK TO TRIAL AS REPORTED BY SCRUB NURSE. THERE WAS A SURGICAL DELAY OF UNKNOWN LENGTH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14013466 ↗
  • MalfunctionFeb 4, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    PROSTHESES / INSTRUMENTS FROM THE DPS JOINTS FRANCHISE ARE SHOWING (COMING OUT) SOOT AFTER THE STERILIZATION PROCESS AND THAT THE BOXES ALSO HAVE DARK SPOTS ON THE TRAYSWHICH, ACCORDING TO THE HOSPITAL, REPRESENTS THAT THE MATERIAL IS CONTAMINATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13454619 ↗
  • MalfunctionFeb 4, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    PROSTHESES / INSTRUMENTS FROM THE DPS JOINTS FRANCHISE ARE SHOWING (COMING OUT) SOOT AFTER THE STERILIZATION PROCESS AND THAT THE BOXES ALSO HAVE DARK SPOTS ON THE TRAYS WHICH, ACCORDING TO THE HOSPITAL, REPRESENTS THAT THE MATERIAL IS CONTAMINATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13454342 ↗
  • MalfunctionDec 17, 2020Material Deformation; Illegible Information; Material Deformation; Illegible Information

    METAL CONNECTION OF BROACH WAS BENT. NOT USEABLE. BROACH NUMBER 16 CORAIL. NECK PIECE GETTING STUCK DUE TO ANGLE OF BENT METAL. DIFFICULT TO REMOVE. UNABLE TO TRIAL. NO ADVERSE EVENTS. NO SURGICAL DELAY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11028665 ↗
  • MalfunctionFeb 21, 2020Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THAT THE SURGEON PERFORMED A DIRECT ANTERIOR APPROACH TOTAL HIP REPLACEMENT ON THE LEFT SIDE WITH A PINNACLE CUP AND A CORAIL STEM. THE SURGEON BROACHED UP TO A SIZE 16 CORAIL BROACH. WHEN HE WANTED TO CALCAR PLANE, HE REALIZED THE CALCAR PLANER DID NOT SEAT FULLY OVER THE NUB ON THE BROACH. AFTER OPENING ANOTHER TRAY WITH ANOTHER CALCAR PLANER, IT WAS DETERMINED THAT SOMETHING MUST BE WRONG WITH THE SIZE 16 CORAIL BROACH.

    Patient problems: No Code Available

    MAUDE report 9741990 ↗
  • MalfunctionNov 12, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENTS DON'T ATTACH TO BROACH HANDLE WELL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310079 ↗
  • MalfunctionNov 12, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENTS DON'T ATTACH TO BROACH HANDLE WELL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310063 ↗
  • MalfunctionNov 12, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENTS DON'T ATTACH TO BROACH HANDLE WELL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310059 ↗
  • MalfunctionNov 12, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENTS DON'T ATTACH TO BROACH HANDLE WELL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310056 ↗
  • MalfunctionOct 25, 2019Device-Device Incompatibility; Device-Device Incompatibility

    WON'T GO IN BROACH HANDLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9241353 ↗
  • MalfunctionOct 17, 2019Material Twisted/Bent; Material Twisted/Bent

    THE POST ON A SIZE 16 CORAIL BROACH (L20416) HAS BECOME DEFORMED AS A RESULT OF CALCAR PLANING. NECK TRIALS ARE NO LONGER FITTING ON THE BROACH. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? NO. WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. WERE FRAGMENTS GENERATED? NO. PATIENT STATUS/ OUTCOME / CONSEQUENCES: NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9203792 ↗
  • MalfunctionMay 30, 2019Material Twisted/Bent; Scratched Material; Material Twisted/Bent; Scratched Material

    IT WAS REPORTED THAT AFTER OUR INSTRUMENTS CAME OUT OF THE WASH, IT WAS FOUND THAT THE 25MM DRILL BIT WAS STUCK IN THE DRILL GUIDE. AS WELL, THE FIRST LEVEL OF A CORAIL TRAY WAS DAMAGED IN STERILE PROCESSING. FINALLY, THE POST ON THE SIZE 16 CORAIL BROACH WAS BENT WHILE STERILE PROCESSING. SURGERY WAS NOT DELAYED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8655159 ↗
  • MalfunctionFeb 14, 2019Difficult to Insert; Scratched Material; Difficult to Insert; Scratched Material

    IT WAS REPORTED THAT THE INSTRUMENT OFFICE SETS HAD AN UNSPECIFIED MALFUNCTION. IT DID NOT AFFECT SURGERY OR PATIENT. IN-HOUSE ENGINEERING EVALUATION DETERMINED THE FOLLOWING: A FUNCTIONAL CHECK WITH A MATING BROACH HANDLE FOUND THE CIRCULAR SCRATCHING ON THE TAPER PREVENTED FROM SEATING CORRECTLY INTO THE BROACH HANDLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8339794 ↗
  • MalfunctionJul 5, 2018Break; Break

    AFTER REMOVAL OF O-RING AS PER FSCA CORAIL (B)(4). THE TRIAL NECK AND THE TRIAL STEM ARE DISASSEMBLING/SEPARATING WITHOUT ANY FORCES. THIS DID NOT HAPPEN INTAOPERATIVELY, NO ADVERSE CONSEQUENCES FOR ANYBODY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7663864 ↗
  • MalfunctionNov 7, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE (B)(6) BROACH HAS SLIGHT WEAR AND TEAR ON THE TRUNION. IT¿S CAUSING DIFFICULTY ACCEPTING FEMORAL NECK TRIAL AS WELL AS A CALCAR PLANER.

    Patient problems: No Information

    MAUDE report 7007194 ↗

Other BROACH devices

All BROACH devices →

Sources and updates