Skip to content

ACTIS 2010-01-010Recalls & reported problems

Depuy (Ireland) · ACTIS BROACH SIZE 1 · Class 1

All models of ACTIS (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ACTIS broach is a surgical tool used to shape bone for a hip implant. Many of its 88 reports describe broaches breaking, and some describe a broken piece stuck in the patient or failure to cut. The reports include 1 injury and no deaths. Reports rose from 7 in the prior 12 months to 21 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
87
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 87 adverse event reports that list this exact device (DI 10603295393634), received between May 3, 2018 and Sep 1, 2026 — classified as 1 injury, 86 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (35 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 15 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility35
  • Break29
  • Failure to Cut12
  • Difficult to Advance9
  • Material Twisted/Bent7
  • Scratched Material7
  • No Apparent Adverse Event6
  • Fitting Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions58
  • No Consequences Or Impact To Patient24
  • Insufficient Information3
  • Limb Fracture1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 87 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2026Break; Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50544147 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24732460 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24731807 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24731782 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE BROACH UNABLE TO EXTRACT WITH BROKEN TRUNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638268 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE BROACH UNABLE TO EXTRACT WITH BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638242 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE BROACH UNABLE TO EXTRACT WITH BROKEN TRUNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638207 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE BROACH UNABLE TO EXTRACT WITH BROKEN TRUNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638119 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE BROACH UNABLE TO EXTRACT WITH BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638086 ↗
  • MalfunctionMar 13, 2026Break; Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24598813 ↗
  • MalfunctionMar 13, 2026Break; Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THAT THE BROACH WAS UNABLE TO EXTRACT WITH BROKEN TRUNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24598613 ↗
  • MalfunctionMar 13, 2026Break; Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24593308 ↗
  • MalfunctionMar 13, 2026Break; Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24593198 ↗
  • MalfunctionMar 13, 2026Break; Device-Device Incompatibility; No Apparent Adverse Event

    IT WAS REPORTED THE BROACH IS UNABLE TO EXTRACT DUE TO A BROKEN TRUNNION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24593093 ↗
  • MalfunctionMar 12, 2026Break

    IT WAS REPORTED THAT BROACHES ARE BROKEN INTO TWO PIECES. NO ADDITIONAL FRAGMENTS. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24583535 ↗
  • MalfunctionMar 11, 2026Break

    IT WAS REPORTED THE BROACH WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24575576 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT THE HAD QTY:1 SIZE 6 BROACH TRUNION SNAP DURING EXTRACTION. IT BROKE DURING REMOVAL NO TIME ADDED. ALL PARTS WERE RETRIEVED. SWAPPED ALL BROACHES FOR SLOTTED BROACHES FOR EASE OF EXTRACTION PER SURGEON REQUEST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24051533 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT THE HAD QTY:1 SIZE 6 BROACH TRUNION SNAP DURING EXTRACTION. IT BROKE DURING REMOVAL NO TIME ADDED. ALL PARTS WERE RETRIEVED. SWAPPED ALL BROACHES FOR SLOTTED BROACHES FOR EASE OF EXTRACTION PER SURGEON REQUEST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24051532 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT THE HAD QTY:1 SIZE 6 BROACH TRUNION SNAP DURING EXTRACTION. IT BROKE DURING REMOVAL NO TIME ADDED. ALL PARTS WERE RETRIEVED. SWAPPED ALL BROACHES FOR SLOTTED BROACHES FOR EASE OF EXTRACTION PER SURGEON REQUEST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24051530 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT THE HAD QTY:1 SIZE 6 BROACH TRUNION SNAP DURING EXTRACTION. IT BROKE DURING REMOVAL NO TIME ADDED. ALL PARTS WERE RETRIEVED. SWAPPED ALL BROACHES FOR SLOTTED BROACHES FOR EASE OF EXTRACTION PER SURGEON REQUEST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24051523 ↗
  • MalfunctionOct 15, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS BROACHES NEEDED TO BE REPLACED. THERE WERE SEVERE STRESS RINGS ON THEM WHERE THEY ATTACH TO THE BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM. WHEN THEY BREAK, IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23283381 ↗
  • MalfunctionOct 10, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THERE ARE SEVERE STRESS RINGS ON ACTIS BROACHES WHERE THEY ATTACH TO BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM AND WHEN THEY BREAK IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23272686 ↗
  • MalfunctionMay 30, 2025Break; Break

    IT WAS REPORTED THAT THE ACTIS HIP BROACHES POSTS SHEARED OFF AND DO NOT HAVE THE T-SLOT FOR EXTRACTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22115223 ↗
  • MalfunctionMay 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE ACTIS BROACH WON¿T SEPARATE FROM THE KINCISE ANTERIOR BROACH ADAPTOR. PATIENT NOT AFFECTED. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084397 ↗
  • MalfunctionJan 2, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED BROACHES DID NOT BREAK BUT EXTRACTOR WILL NOT FIT INTO THESE. THERE WAS NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21061750 ↗

Other BROACH devices

All BROACH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page