Skip to content

ACTIS 2010-01-100Recalls & reported problems

Depuy (Ireland) · ACTIS BROACH SIZE 10 · Class 1

All models of ACTIS (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ActIS broach is a surgical instrument used to shape the femur bone during hip replacement. The 61 reports mostly describe broaches breaking during surgery, sometimes into multiple pieces that had to be removed, along with difficulty advancing the instrument, failure to cut, and compatibility problems with other devices. No deaths or injuries are reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
62
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 10603295393726), received between May 4, 2018 and Sep 9, 2026 — classified as 62 malfunctions.

The most frequently reported specific device problem is “Break” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 6 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break20
  • Device-Device Incompatibility18
  • Failure to Cut12
  • Difficult to Advance10
  • Scratched Material4
  • Appropriate Device Problem Term/Code Not Available3
  • Fitting Problem3
  • Material Deformation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions41
  • No Consequences Or Impact To Patient15
  • Insufficient Information5
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Break; Device-Device Incompatibility

    It was reported that the Broaches were broke during a case. Surgeon needed to order all new sets of broaches with the new Design in case this happens again.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50689155 ↗
  • MalfunctionAug 20, 2026Break; Entrapment of Device

    It was reported that during a Total Hip Replacement surgical procedure it was observed that the broaches broke and could not be extracted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50346130 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742029 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742004 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS REPORTED THE OLD STYLE ACTIS BROACHES WERE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24741906 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THAT BROACHES ARE BROKEN INTO TWO PIECES. NO ADDITIONAL FRAGMENTS. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24591477 ↗
  • MalfunctionMar 12, 2026Break

    IT WAS REPORTED THAT THIS FACILITY HAD MULTIPLE VERSION 1 AND 2 BROACHES BREAK, AND WERE REQUESTING ALL OF THEIR CONSIGNED BROACHES BE SWAPPED OUT FOR VERSION 3 TO AVOID FURTHER BREAKAGES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579464 ↗
  • MalfunctionMar 12, 2026Break

    IT WAS REPORTED THAT THIS FACILITY HAD MULTIPLE VERSION 1 AND 2 BROACHES BREAK AND WERE REQUESTING ALL OF THEIR CONSIGNED BROACHES BE SWAPPED OUT FOR VERSION 3 TO AVOID FURTHER BREAKAGES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579455 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT A SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION AND BROKE INTO MULTIPLE PIECES DURING REMOVAL. ALL PARTS WERE RETRIEVED, AND THE SURGEON REQUESTED SWAPPING ALL BROACHES FOR SLOTTED BROACHES TO EASE FUTURE EXTRACTIONS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24060276 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT A SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION AND BROKE INTO MULTIPLE PIECES DURING REMOVAL. ALL PARTS WERE RETRIEVED, AND THE SURGEON REQUESTED SWAPPING ALL BROACHES FOR SLOTTED BROACHES TO EASE FUTURE EXTRACTIONS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24060275 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT A SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION AND BROKE INTO MULTIPLE PIECES DURING REMOVAL. ALL PARTS WERE RETRIEVED, AND THE SURGEON REQUESTED SWAPPING ALL BROACHES FOR SLOTTED BROACHES TO EASE FUTURE EXTRACTIONS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24060272 ↗
  • MalfunctionJan 13, 2026Break

    IT WAS REPORTED THAT A SIZE 6 BROACH TRUNION SNAPPED DURING EXTRACTION AND BROKE INTO MULTIPLE PIECES DURING REMOVAL. ALL PARTS WERE RETRIEVED, AND THE SURGEON REQUESTED SWAPPING ALL BROACHES FOR SLOTTED BROACHES TO EASE FUTURE EXTRACTIONS. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24060264 ↗
  • MalfunctionOct 15, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS BROACHES NEED TO BE REPLACED. THERE WERE SEVERE STRESS RINGS ON THEM WHERE THEY ATTACH TO BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM. WHEN THEY BREAK IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT. (B)(4) IS RELATED TO (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23277850 ↗
  • MalfunctionOct 15, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS BROACHES NEED TO BE REPLACED. THERE WERE SEVERE STRESS RINGS ON THEM WHERE THEY ATTACH TO BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM. WHEN THEY BREAK IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT. (B)(4) IS RELATED TO(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23277824 ↗
  • MalfunctionOct 15, 2025Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ACTIS BROACHES NEED TO BE REPLACED. THERE WERE SEVERE STRESS RINGS ON THEM WHERE THEY ATTACH TO BROACH HANDLE. SURGEON FREQUENTLY BREAKS THEM. WHEN THEY BREAK IT IS EXTREMELY HARD TO REMOVE FROM THE PATIENT. (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23277807 ↗
  • MalfunctionJun 23, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT BROACH WON'T SEAT ON THE HANDLE COMPLETELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22308275 ↗
  • MalfunctionMay 30, 2025Break; Break

    IT WAS REPORTED THAT THE ACTIS HIP BROACHES POSTS SHEARED OFF AND DO NOT HAVE THE T-SLOT FOR EXTRACTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22114695 ↗
  • MalfunctionMay 30, 2025Break; Break

    IT WAS REPORTED THAT THE ACTIS HIP BROACHES POSTS SHEARED OFF AND DO NOT HAVE THE T-SLOT FOR EXTRACTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22114550 ↗
  • MalfunctionJan 2, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THE EXTRACTOR WOULD NOT FIT INTO THE OLD BROACHES SO THEY SWAPPED OUT FOR NEW ONES.

    Patient problems: Insufficient Information

    MAUDE report 21056603 ↗
  • MalfunctionOct 21, 2024Break; Naturally Worn; No Apparent Adverse Event; Break; Naturally Worn; No Apparent Adverse Event

    IT WAS REPORTED THAT THE ACTIS BROACHES ARE DAMAGED AND REQUIRE REPLACEMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20492967 ↗
  • MalfunctionAug 20, 2024Device-Device Incompatibility; Device-Device Incompatibility

    BROACHES DO NOT FIT SMOOTHLY INTO BRASH HANDLES OR KINCISE BROACH ADAPTERS. DIFFERENT BROACH HANDLES AND ADAPTERS WERE ATTEMPTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20006074 ↗
  • MalfunctionJan 18, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE TRAIL NECKS DO NOT GO ALL THE WAY DOWN. IT APPEARS THE ATTEMPT PIECE IS BENT.

    Patient problems: Insufficient Information

    MAUDE report 18534006 ↗
  • MalfunctionOct 13, 2023Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE ACTIS SIZE 10 BROACH WAS INSERTED INTO THE FEMUR VIA KINCISE. REMOVAL OF THE BROACH FROM FEMUR WAS ATTEMPTED USING KINCISE, AN EXTRA CURVED BROACH HANDLE, AND A DOUBLE OFFSET BROACH HANDLE. AFTER SUCCESSFUL EXTRACTION OF BROACH, IMPERFECTIONS IN THE NECK/ TRUNION AREA OF BROACH WERE PRESENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17930064 ↗
  • MalfunctionOct 12, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED BY THE SURGEON THAT THERE WAS A MISMATCH BETWEEN OUR BROACH TO IMPLANT SIZES, AND ALSO A MISMATCH BETWEEN OUR TRIAL HEADS TO OUR REAL HEADS. PARTS WERE SENT IN AND INSPECTED. UNKNOWN SURGICAL DELAY. DOE: (B)(6) 2023. AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 17917546 ↗
  • MalfunctionOct 11, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED BY THE SURGEON THAT THERE WAS A MISMATCH BETWEEN OUR BROACH TO IMPLANT SIZES, AND ALSO A MISMATCH BETWEEN OUR TRIAL HEADS TO OUR REAL HEADS. PARTS WERE SENT IN AND INSPECTED. UNKNOWN SURGICAL DELAY. DOE: (B)(6) 2023. AFFECTED SIDE: LEFT HIP.

    Patient problems: Insufficient Information

    MAUDE report 17917539 ↗

Other BROACH devices

All BROACH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page