ATTUNE 2545-01-702Recalls & reported problems

Depuy (Ireland) · ATTUNE TIBIA PREP & IMPACT · Class 1

All models of ATTUNE (1471)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy ATTUNE tibia preparation and impact surgical tray is used to prepare the bone during knee replacement surgery. Most of the 46 reports describe the tray's surface coating or paint peeling and delaminating, along with trays cracking, chipping, warping, or wearing down from repeated use and sterilization. No deaths or injuries are reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 10603295423522), received between Jul 16, 2020 and Sep 1, 2026 — classified as 42 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated32
  • Break5
  • Naturally Worn5
  • Material Twisted/Bent4
  • Crack1
  • Illegible Information1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • No Consequences Or Impact To Patient4
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549313 ↗
  • MalfunctionMay 7, 2026Peeled/Delaminated

    IT WAS REPORTED THAT AFTER REPEATED USAGE THE PAN HAD BEGAN PEELING AND WAS NO LONGER UP TO STANDARD FOR THE HOSPITAL. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25109677 ↗
  • MalfunctionMar 31, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE PAINT IS PEELING FORM TRAYS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24744937 ↗
  • MalfunctionMar 13, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAY'S PLASTIC COATING IS DELAMINATING. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24597852 ↗
  • MalfunctionFeb 2, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TWO ATTUNE PANS ARE PEELING FROM REPEATED USE. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24237845 ↗
  • MalfunctionNov 18, 2025Crack; Peeled/Delaminated; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENT TRAY WAS CRACKING DURING THE CLEANING PROCESS. IT WAS REPORTED THAT THERE WAS NO SURGICAL DELAY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586446 ↗
  • MalfunctionJul 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENTS WERE PEELING. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22398465 ↗
  • MalfunctionJul 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENTS WERE PEELING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22378513 ↗
  • MalfunctionMar 4, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED BY CENTRAL STERILE SERVICES DEPARTMENT (CSSD) THAT THE PLASTIC COATING IS DELAMINATING FROM METAL TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21518938 ↗
  • MalfunctionMar 4, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED BY CENTRAL STERILE SERVICES DEPARTMENT (CSSD) THAT THE PLASTIC COATING IS DELAMINATING FROM METAL TRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21518937 ↗
  • MalfunctionFeb 7, 2025Peeled/Delaminated; Illegible Information; Peeled/Delaminated; Illegible Information

    IT WAS REPORTED BY STERILIZATION DEPARTMENT THAT THE PLASTIC ISOLATION IS FALLING OFF OF THE INSTRUMENT TRAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21336526 ↗
  • MalfunctionJan 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS HAD PLASTIC PEELING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074232 ↗
  • MalfunctionJan 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS HAD PLASTIC PEELING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074231 ↗
  • MalfunctionJan 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS HAD PLASTIC PEELING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074230 ↗
  • MalfunctionJan 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS HAD PLASTIC PEELING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074229 ↗
  • MalfunctionJan 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS HAD PLASTIC PEELING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074225 ↗
  • MalfunctionJan 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS HAD PLASTIC PEELING OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21074222 ↗
  • MalfunctionDec 19, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING STERILE PROCESSING IT WAS NOTICED THAT THE PANS WERE DELAMINATING IN SOME PLACES IN THE PAN. THE PLASTIC OR PAINT LAMINATE WAS COMING OFF IN SEVERAL PLACES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20982044 ↗
  • MalfunctionDec 19, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING STERILE PROCESSING IT WAS NOTICED THAT THE PANS WERE DELAMINATING IN SOME PLACES IN THE PAN. THE PLASTIC OR PAINT LAMINATE WAS COMING OFF IN SEVERAL PLACES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20982040 ↗
  • MalfunctionAug 20, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE INNER BLACK LINING IS PEELING OFF. HOSPITAL WILL NOT STERILIZE TRAY DUE TO THIS REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20007001 ↗
  • MalfunctionJul 1, 2024Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19650951 ↗
  • MalfunctionApr 9, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE GRAPHIC CASE FOR INSTRUMENTS IS FLAKING AND PEELING, DEEMING IT CONTAMINATED FOR PROCEDURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075723 ↗
  • MalfunctionMar 27, 2024Peeled/Delaminated; Scratched Material; Peeled/Delaminated; Scratched Material

    ADDITIONAL INFORMATION N RECEIVED: A. WAS THERE ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT? NO. B. WAS THERE A SURGICAL DELAY? IF YES, WHAT IS THE DURATION OF THE DELAY? NO. C. CAN YOU PLEASE CLARIFY WHAT DO YOU MEAN BY ¿DAMAGED¿? WAS IT WORN, CRACKED, BROKEN INTO PIECES, BENT, STRIPPED, CROSS THREADED OR ANY DEVICE INTERACTION. GOUGE IN BLOCK. D. EVENT DESCRIPTION STATES THAT "CUT BLOCK UNABLE TO USE DUE TO SURFACE DAMAGE." PLEASE CLARIFY WHAT YOU MEAN BY DAMAGED. WAS IT BROKE, CRACKED, BENT OR ANY OTHER DEFICIENCY WITH THE DEVICE THAT WAS NOT FUNCTION AS INTENDED? DEEP LACERATIONS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985636 ↗
  • MalfunctionFeb 1, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PAINT WAS PEELING ON TRAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18622686 ↗
  • MalfunctionNov 28, 2023Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THERE WAS A CHIPPING OF THE TRAY. UNKNOWN SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18218545 ↗

FDA clearance and classification

The FDA regulates product code FSM (tray, surgical, instrument) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ATTUNE 2545-01-702

Has the ATTUNE 2545-01-702 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295423522), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ATTUNE 2545-01-702?

The FDA's MAUDE database lists 42 adverse event reports for this device, received between Jul 16, 2020 and Sep 1, 2026. In those reports, the most common specific device problems are “Peeled/Delaminated” (32), “Break” (5) and “Naturally Worn” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ATTUNE 2545-01-702?

None of the reports are classified as deaths, 0 as injuries and 42 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ATTUNE 2545-01-702 FDA approved?

Devices of this type (product code FSM) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the ATTUNE 2545-01-702, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand ATTUNE, model 2545-01-702). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page