PINNACLE 2220-00-110Recalls & reported problems

Depuy (Ireland) · PINNACLE CORE CASE · Class 1

All models of PINNACLE (198)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle surgical instrument tray is used to hold and carry instruments during surgery. The 27 reports all describe the black plastic coating peeling or delaminating from the metal framing, with flakes sometimes coming loose inside the tray. None report a death, an injury, or a patient consequence, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10603295424185), received between Mar 23, 2021 and May 28, 2026 — classified as 27 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 8 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated27

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313481 ↗
  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313292 ↗
  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313290 ↗
  • MalfunctionMay 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS ARE STARTING TO PEEL IN THE INSIDES DUE TO WEAR AND TEAR. NO DELAYS NOTHING LEFT INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25313289 ↗
  • MalfunctionApr 27, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE BLACK PLASTIC COATING THAT COVERS SOME OF THE METAL FRAMING ON THE TRAY HAS PEELED OFF AND FLAKES INTO THE TRAY WHICH COMPROMISES ITS STERILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24997542 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427290 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427286 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427283 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427282 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427279 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427277 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427269 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN CORE TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427262 ↗
  • MalfunctionJul 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENTS WERE PEELING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22378517 ↗
  • MalfunctionJul 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENTS WERE PEELING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22378516 ↗
  • MalfunctionJul 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENTS WERE PEELING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22378510 ↗
  • MalfunctionJul 8, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT A CORE CASE IS DAMAGED. THE PLASTIC COATING INSIDE THE CASE HAS BEGUN TO LOOSEN. NO IMPACT ON SURGERY OR DELAY FOR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19689166 ↗
  • MalfunctionJun 21, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC COATING INSIDE THE CASE HAS BEGUN TO LOOSEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19582925 ↗
  • MalfunctionJun 7, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT A CORE CASE IS DAMAGED. THE PLASTIC COATING INSIDE THE CASE HAS BEGUN TO LOOSEN. NO IMPACT ON SURGERY OR DELAY FOR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19484816 ↗
  • MalfunctionMar 27, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS WERE PEELING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18987168 ↗
  • MalfunctionMar 26, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS WERE PEELING. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18984319 ↗
  • MalfunctionNov 29, 2023Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS PRESENTED WERE CHIPPING. CALLED FOR NEW SETS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18226850 ↗
  • MalfunctionApr 27, 2023Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT PAINT FLAKES OFF CREATING CONTAMINATION ISSUE FOR HOSPITAL. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16824201 ↗
  • MalfunctionSep 21, 2021Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAY HAS COLOR FLAKES, AND CANNOT BE STERILIZED PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12510893 ↗
  • MalfunctionMay 11, 2021Peeled/Delaminated; Peeled/Delaminated

    PLASTIC TORN SEE PICTURE. NO REPORTED SURGICAL DELAYS OR PATIENT CONSEQUENCES. NO MORE INFO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11805492 ↗

FDA clearance and classification

The FDA regulates product code FSM (tray, surgical, instrument) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the PINNACLE 2220-00-110

Has the PINNACLE 2220-00-110 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295424185), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PINNACLE 2220-00-110?

The FDA's MAUDE database lists 27 adverse event reports for this device, received between Mar 23, 2021 and May 28, 2026. In those reports, the most common specific device problem is “Peeled/Delaminated” (27). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PINNACLE 2220-00-110?

None of the reports are classified as deaths, 0 as injuries and 27 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PINNACLE 2220-00-110 FDA approved?

Devices of this type (product code FSM) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the PINNACLE 2220-00-110, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand PINNACLE, model 2220-00-110). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.