PINNACLE 2220-00-130Recalls & reported problems

Depuy (Ireland) · PINNACLE SCREW SET CASE · Class 1

All models of PINNACLE (198)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Pinnacle screw set case is a plastic tray used to hold orthopedic screws. The 33 reports mostly describe the plastic peeling or delaminating from inside the tray, which some say left it no longer able to be sterilized; a few also mention discoloration, contamination, or a break. No deaths or injuries were reported. Reports rose recently, with 14 in the last 12 months versus 6 in the prior 12. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 10603295424208), received between Apr 26, 2023 and Apr 28, 2026 — classified as 32 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 5 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 9 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated32
  • Contamination /Decontamination Problem4
  • Material Discolored4
  • Break1
  • Illegible Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2026Peeled/Delaminated

    IT WAS REPORTED THAT PLASTIC WAS PEELING OFF THE INSIDE OF THE TRAY. THERE WAS NO DELAY TO A PROCEDURE. NO HARM TO PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25022431 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427591 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427590 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427585 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427584 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427580 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427569 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427568 ↗
  • MalfunctionFeb 24, 2026Peeled/Delaminated

    IT WAS REPORTED THE PINN SCREW TRAYS ARE DELAMINATING. THE EVENT DID NOT HAPPEN IN SURGERY. NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427567 ↗
  • MalfunctionFeb 10, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE COATING IN THE TRAY WAS PEELING AND THERE WAS PAINT PEELING FROM THE SIDE OF THE TRAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305965 ↗
  • MalfunctionNov 21, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE COATING ON DIVIDERS FOR INSTRUMENT SEATING WERE PEELING AND HOSPITALS WERE REJECTING THE DEVICES DUE TO CONTAMINATION LIABILITY. IT WAS REPORTED THAT THE EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23617862 ↗
  • MalfunctionNov 20, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT PAINT FALLING OFF PAN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23607369 ↗
  • MalfunctionNov 6, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC INSIDE THE SCREW SET PAN IS DELAMINATING AND FLAKING OFF. SPD SAYS THE TRAY NEEDS TO BE REPLACED. NO HARM TO PATIENT. NO DELAY. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23485458 ↗
  • MalfunctionAug 22, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS PORTED THAT PLASTIC COATING INSIDE THE PAN WAS DELAMINATED AND FLAKING OFF. THERE WAS NO HARM TO PATIENT AND THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22873743 ↗
  • MalfunctionAug 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC HAS DELAMINATED FROM THE INSIDE OF THE TRAYS AND SPD WILL NOT STERILIZE THEM. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689454 ↗
  • MalfunctionAug 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC HAS DELAMINATED FROM THE INSIDE OF THE TRAYS AND SPD WILL NOT STERILIZE THEM. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22689435 ↗
  • MalfunctionJul 9, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC IS COMING OFF THE INSIDE OF THE SCREW SET AND GRATER HEADS TRAYS. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22462490 ↗
  • MalfunctionJul 9, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC ON THE INSIDE OF THE SCREW AND GRATER PANS IS DELAMINATING AND SPD SAYS THEY HAVE TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22460491 ↗
  • MalfunctionJul 1, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT THE PAN WAS PEELING, THE PINNACLE SCREW SET CASE WAS PEELING, 1 UNKNOWN PRODUCT WAS CRACKED AND ANOTHER UNKNOWN PRODUCT WAS ALSO PEELING. THE EVENT DID NOT HAPPENED DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22378549 ↗
  • MalfunctionJun 6, 2024Material Discolored; Peeled/Delaminated; Material Discolored; Peeled/Delaminated

    IT WAS REPORTED THAT THE TRAYS NEED TO BE REPLACED DUE TO DISCOLORATION AND FLAKING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19477201 ↗
  • MalfunctionMay 14, 2024Material Discolored; Peeled/Delaminated; Material Discolored; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS NEED TO BE REPLACED DUE TO DISCOLORATION AND FLAKING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19312988 ↗
  • MalfunctionMay 14, 2024Material Discolored; Peeled/Delaminated; Material Discolored; Peeled/Delaminated

    IT WAS REPORTED THAT TRAYS NEED TO BE REPLACED DUE TO DISCOLORATION AND FLAKING. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19312986 ↗
  • MalfunctionMay 13, 2024Material Discolored; Peeled/Delaminated; Material Discolored; Peeled/Delaminated

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO CONSEQUENCES, TRAYS NEED TO BE REPLACED AS THEY HAVE BECOME OLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19298623 ↗
  • MalfunctionMar 20, 2024Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRAYS ARE CHIPPING FLAKES OFF. THE HOSPITAL IS DEEMING THEM CONTAMINATED AFTER STERILE PROCESS AND UNFIT FOR PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18942131 ↗
  • MalfunctionMar 20, 2024Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE TRAYS ARE CHIPPING FLAKES OFF. THE HOSPITAL IS DEEMING THEM CONTAMINATED AFTER STERILE PROCESS AND UNFIT FOR PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18942099 ↗