ATTUNE 2504-40-551Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION BOSS REAMER · Class 1
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295439875), received between Sep 22, 2019 and Apr 10, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility4
- Naturally Worn2
- Break1
- Failure to Cut1
- Material Deformation1
- Material Integrity Problem1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT DURING THE PROCEDURE THE BOSS REAMER WOULD NOT PASS THROUGH THE REAMER GUIDE AS INTENDED. AFTER VISUAL INSPECTION IT WAS FOUND TO HAVE BURS ON THE REAMER SHAFT KEEPING IT FROM ADVANCING. IT WAS UNKNOWN, IF THERE WAS A SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 19081463 ↗MalfunctionMar 14, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE BOSS REAMER IS STRIPPED. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16541726 ↗MalfunctionMar 15, 2022Material Integrity Problem; Material Integrity Problem
TECHNICIAN REPORTS THAT DURING THE INSPECTION THERE WAS A DEFECTIVE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13771471 ↗MalfunctionJan 10, 2022Failure to Cut; Device-Device Incompatibility; Failure to Cut; Device-Device Incompatibility
IT WAS REPORTED THAT THE CUTTING FLUTES ARE DAMAGED CREATING JAGGED EDGES. WILL NOT FIT SMOOTHLY THROUGH REVISION REAMER TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13204507 ↗MalfunctionNov 12, 2021Scratched Material; Scratched Material
IT WAS REPORTED THAT DURING REAMING THE REAMER CAME IN CONTACT WITH PREVIOUS HARDWARE IN PATIENT AND GAUGED IT HEAVILY. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12803708 ↗MalfunctionNov 12, 2021Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn
IT WAS FOUND THAT D 250440551 FLUTES WERE WORN AND WOULD NOT FIT THROUGH THE REAMING TOWER/BARREL WHILE DEVICE WAS BEING CONSIGNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12802518 ↗MalfunctionSep 22, 2019Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
IT WAS REPORTED THAT THE BOSS REAMER WOULD NOT FIT DOWN BUSHING ADAPTERS OF THE CUT THROUGH TRIALS. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9102259 ↗
Other REAMER devices
- QUICKSET 2440-00-551Depuy (Ireland) · REAMER1,150 reports
- QUICKSET 2440-00-549Depuy (Ireland) · REAMER1,113 reports
- QUICKSET 2440-00-553Depuy (Ireland) · REAMER1,105 reports
- QUICKSET 2440-00-547Depuy (Ireland) · REAMER1,073 reports
- QUICKSET 2440-00-555Depuy (Ireland) · REAMER1,065 reports
- QUICKSET 2440-00-557Depuy (Ireland) · REAMER1,031 reports
- QUICKSET 2440-00-545Depuy (Ireland) · REAMER947 reports
- QUICKSET 2440-00-552Depuy (Ireland) · REAMER888 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page