Skip to content

ATTUNE 2545-01-714Recalls & reported problems

Depuy (Ireland) · ATTUNE PATELLA PREP (GEN 2) · Class 1

All models of ATTUNE (1471)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This is a DePuy ATTUNE Patella Prep tray, a reusable surgical tray for instruments used in kneecap preparation. Most of the 32 reports describe the tray's plastic peeling or delaminating, plus cracks, breaks, and paint flaking off; in several cases sterile processing would not accept the tray for reuse. No deaths or injuries were reported, and no recalls name this device. Reports fell from 22 in the earlier 12-month period to 6 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10603295442936), received between Feb 21, 2023 and Sep 30, 2026 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 7 reportsFeb 2025: 11 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 9 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated38
  • Crack11
  • Break8
  • Contamination /Decontamination Problem1
  • Device-Device Incompatibility1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Peeled/Delaminated

    It was reported that the black plastic on the inside of the trays is delaminating.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944314 ↗
  • MalfunctionSep 30, 2026Peeled/Delaminated

    It was reported that the black plastic on the inside of the trays is delaminating.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944299 ↗
  • MalfunctionSep 30, 2026Peeled/Delaminated

    It was reported that the black plastic on the inside of the trays is delaminating.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944282 ↗
  • MalfunctionSep 30, 2026Peeled/Delaminated

    It was reported that the black plastic on the inside of the trays is delaminating.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944280 ↗
  • MalfunctionSep 30, 2026Peeled/Delaminated

    It was reported that the black plastic on the inside of the trays is delaminating.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944270 ↗
  • MalfunctionSep 28, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921060 ↗
  • MalfunctionSep 28, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921031 ↗
  • MalfunctionSep 28, 2026Peeled/Delaminated

    It was reported that the black plastic inside the trays was delaminating. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920700 ↗
  • MalfunctionSep 28, 2026Peeled/Delaminated

    It was reported that the black plastic inside the trays was delaminating. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920523 ↗
  • MalfunctionMay 6, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC WAS COMING OFF THE I SIDE OF THE TRAY AND STERILE PROCESSING (SPD) WOULD NOT PROCESS IT. NO DELAY, NO HARM TO PATIENT, NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25091938 ↗
  • MalfunctionMay 5, 2026Peeled/Delaminated

    IT WAS REPORTED THAT THE PLASTIC PEELED OFF FROM INSIDE THE PAN AND STERILE PROCESSING DEPARTMENT WOULD NOT STERILIZE IT. NO DELAY. NO HARM TO THE PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25086300 ↗
  • MalfunctionJan 28, 2026Peeled/Delaminated

    PATELLA PREP TRAYS HAVE BURNS ON THE INSIDE OF THE TRAY AND PAINT WAS FLAKING OFF ON THE EXTERIOR OF THE TRAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24200441 ↗
  • MalfunctionJan 28, 2026Peeled/Delaminated

    PATELLA PREP TRAYS HAVE BURNS ON THE INSIDE OF THE TRAY AND PAINT WAS FLAKING OFF ON THE EXTERIOR OF THE TRAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24200424 ↗
  • MalfunctionJan 28, 2026Peeled/Delaminated

    PATELLA PREP TRAYS HAVE BURNS ON THE INSIDE OF THE TRAY AND PAINT WAS FLAKING OFF ON THE EXTERIOR OF THE TRAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24200423 ↗
  • MalfunctionJul 18, 2025Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE DEVICE WAS PEELING AND SCRATCHED. IT DIDN'T HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22543965 ↗
  • MalfunctionJul 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT INSTRUMENTS WERE PEELING. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22398466 ↗
  • MalfunctionJul 2, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT TRAY CHIPPING. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED FOR REPORTING. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22386762 ↗
  • MalfunctionApr 28, 2025Peeled/Delaminated; Device-Device Incompatibility; Peeled/Delaminated; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL KNEE REPLACEMENT ON (B)(6) 2025. THE TRAY DIVIDER COATING HAD DELAMINATED AND CREATED A RISK OF LOOSE CONTAMINATES/ PARTICULATES IN THE TRAY. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21918457 ↗
  • MalfunctionFeb 5, 2025Crack; Peeled/Delaminated; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL IS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21312271 ↗
  • MalfunctionFeb 5, 2025Crack; Peeled/Delaminated; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL IS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21312257 ↗
  • MalfunctionFeb 4, 2025Crack; Peeled/Delaminated; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL IS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21306087 ↗
  • MalfunctionFeb 4, 2025Crack; Peeled/Delaminated; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL IS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21305903 ↗
  • MalfunctionFeb 4, 2025Crack; Peeled/Delaminated; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL IS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21305887 ↗
  • MalfunctionFeb 4, 2025Break; Crack; Peeled/Delaminated; Break; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL WAS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21305192 ↗
  • MalfunctionFeb 4, 2025Break; Crack; Peeled/Delaminated; Break; Crack; Peeled/Delaminated

    IT WAS REPORTED THAT THE HOSPITAL WAS COMPLAINING ABOUT TRAYS DUE TO CHIPPING, PEELING, CRACKING AND OUTER RUBBER PEELING OFF. THE EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21305146 ↗

FDA clearance and classification

The FDA regulates product code FSM (tray, surgical, instrument) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ATTUNE 2545-01-714

Has the ATTUNE 2545-01-714 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295442936), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ATTUNE 2545-01-714?

The FDA's MAUDE database lists 40 adverse event reports for this device, received between Feb 21, 2023 and Sep 30, 2026. In those reports, the most common specific device problems are “Peeled/Delaminated” (38), “Crack” (11) and “Break” (8). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ATTUNE 2545-01-714?

None of the reports are classified as deaths, 0 as injuries and 40 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ATTUNE 2545-01-714 FDA approved?

Devices of this type (product code FSM) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Depuy (Ireland).

Who makes the ATTUNE 2545-01-714, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand ATTUNE, model 2545-01-714). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page