PINNACLE 2999-66-321Recalls & reported problems
Depuy Orthopaedics, Inc. · PINNACLE MAKE TO ORDER Straight Cup Impactor SIZE: 56MM RIGHT · Class 1
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295495048), received between Jan 5, 2018 and Oct 19, 2018 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available3
- Break1
- Dull, Blunt1
- Fracture1
Patient problems
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2018Break; Break
THE STRIKE PLATE ON END OF CUP IMPACTOR BROKE OFF THE SHAFT. ALL PARTS RETRIEVED. PATIENT CONSEQUENCE? NO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7984440 ↗MalfunctionJan 8, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE MTO PINNACLE CUP IMPACTOR PN 299966321 HAD THE THREADED TIP COME LOOSE AND REMAIN IN THE CUP AFTER IMPLANTATION. THE TIP WAS REMOVED FROM THE CUP WITH MINOR DELAY LESS THAN 5 MINUTES AND NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7171384 ↗MalfunctionJan 8, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE MTO PINNACLE CUP IMPACTOR PN 299966321 HAD THE THREADED TIP COME LOOSE AND REMAIN IN THE CUP AFTER IMPLANTATION. THE TIP WAS REMOVED FROM THE CUP WITH MINOR DELAY LESS THAN 5 MINUTES AND NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7171381 ↗MalfunctionJan 5, 2018Dull, Blunt; Appropriate Device Problem Term/Code Not Available; Dull, Blunt; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE MTO PINNACLE CUP IMPACTOR WAS DISCOVERED AT WASH/STERILIZATION TO HAVE THE THREADED TIP INSERT EXTENDED FROM THE FACE OF THE HANDLE BY APPROXIMATELY 1MM. A SECOND HANDLE WAS FOUND WITH DAMAGED THREADS. BOTH OF THESE OCCURRED AFTER ON USE. INSERT RETAINING PIN APPEARS TO HAVE SHEARED. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7167643 ↗MalfunctionJan 5, 2018Fracture; Fracture
IT WAS REPORTED THAT THE MTO PINNACLE CUP IMPACTOR WAS DISCOVERED AT WASH/STERILIZATION TO HAVE THE THREADED TIP INSERT EXTENDED FROM THE FACE OF THE HANDLE BY APPROXIMATELY 1MM. A SECOND HANDLE WAS FOUND WITH DAMAGED THREADS. BOTH OF THESE OCCURRED AFTER ON USE. INSERT RETAINING PIN APPEARS TO HAVE SHEARED. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7167624 ↗
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,295 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,539 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,498 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,456 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall875 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR716 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR704 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR337 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.