DELTA XTEND 1307-62-038Recalls & reported problems
Depuy (Ireland) · DELTA XTEND Lateralized Glenosphere +2mm Eccentric 038mm · Class 2 · Implant
All models of DELTA XTEND (194)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Delta Xtend shoulder implant is a reverse shoulder replacement component used in shoulder arthroplasty. The 28 reports mostly describe surgeons having difficulty getting the glenosphere and metaglene parts to fit and lock together during surgery, sometimes with components found damaged before use, plus cases needing revision surgery for infection, loosening of the stem in the bone, pain, or an unstable shoulder. Reports include 18 injuries and no deaths. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 21
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 10603295502432), received between Feb 10, 2020 and Sep 28, 2026 — classified as 21 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem19
- Difficult to Insert9
- Device Damaged Prior to Use2
- Device Dislodged or Dislocated2
- Defective Device1
- Device-Device Incompatibility1
- Fracture1
- Off-Label Use1
Patient problems
- Insufficient Information7
- Pain7
- Unspecified Infection7
- Joint Laxity4
- No Code Available4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Clinical Signs, Symptoms or Conditions2
- Adhesion(s)1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50923253 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that patient presented with a painful right shoulder. The surgeon revised the glenosphere and poly.Affected Side: Unknown Shoulder.
Patient problems: Pain
MAUDE report 50881916 ↗InjurySep 14, 2026Device Dislodged or Dislocated
It was reported that the patient revised due to dislocation.DOI: (b)(6) 2020.DOR: (b)(6) 2026.Affected side: Left shoulder.
Patient problems: Joint Dislocation
MAUDE report 50752309 ↗InjuryMay 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT REVISED DUE TO INFECTION. EVERYTHING TAKEN OUT. DOI: (B)(6) 2024, DOR: (B)(6) 2026, AFFECTED SIDE: LEFT.
Patient problems: Unspecified Infection
MAUDE report 25303204 ↗InjuryMay 9, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
PATIENT WAS REVISED DUE TO LOOSENING OF THE HUMERAL STEM AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER WAS UNKNOWN. IT WAS ALSO REPORTED THAT THE PATIENT BROKE HER HUMERUS MID SHAFT, JUST DISTAL TO THE HUMERAL STEM. SURGEON REMOVED IMPLANTS TO REVISE SHOULDER USING ANOTHER VENDOR AFTER FIXING FRACTURE. NO KNOWN INFECTION PRESENT. NO DELAY IN SURGERY. STEM WAS A LITTLE LOOSE. GLENOSPHERE WAS NOT ENGAGED VIA THE CENTRAL SCREW TO METAGLENE. METAGLENE WAS NOT LOOSE, BUT REMOVED ANYWAYS. DOI: (B)(6) 2020, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Insufficient Information
MAUDE report 21996435 ↗InjuryFeb 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN UNSTABLE SHOULDER. THE SURGEON UP SIZED CUP COMPONENTS. DOI: UNKNOWN, DOR: (B)(6) 2025. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 21458499 ↗InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD INFECTED SHOULDER. ALL IMPLANTS REMOVED AND ANTIBIOTIC SPACER PLACED. NO DELAYS IN SURGERY. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 21080740 ↗InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, THAT PATIENT HAD REVISED FOR INSTABILITY. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Laxity
MAUDE report 20748014 ↗MalfunctionMar 13, 2024Difficult to Insert; Off-Label Use; Difficult to Insert; Off-Label Use
IT WAS REPORTED THAT THE GLENOSPHERE SCREW WOULD NOT LOCK INTO THE METAGLENE COLLET. EVERYTHING WAS IMPLANTED AND USED ACCORDING THE THE SURGICAL TECHNIQUE, BUT WHEN THE SURGEON WENT TO TIGHTENT THE SCREW INTO THE METAGLENE IT ACTED AS IF IT WAS STRIPPED. A GOOD BITE WAS NOT ATTAINABLE AND THE SCREW WOULD JUST KEEP SPINNING AND NOT ENGAGE INTO THE COLLET. ANY EXCESS BONE WAS REMOVED AROUND THE METAGLENE, SO THE SURGEON DID NOT SEE ANY INTERFERENCE WITH THE HEAD. THE GLENOSPHERE HAD GOOD FIXATION VIA THE MORSE TAPER ONTO THE METAGLENE, SO THE SURGEON LEFT IT AS THAT WITHOUT TIGHTENING THE SCREW.
Patient problems: Insufficient Information
MAUDE report 18892945 ↗MalfunctionMar 12, 2024Difficult to Insert; Difficult to Insert
ADDITIONAL INFORMATION WAS RECEIVED. PATIENT CONSEQUENCES EXTENDED, THE SURGERY TIME. BOTH GLENOSPHERE AND METAGLENE WERE ATTEMPTED TO BE USED IN SAME CASE AND IN BOTH METAGLENE. ENHANCE WAS USED IN COMBINATION WITH DELTA, AFTER THE DELTA EXTEND.
Patient problems: Insufficient Information
MAUDE report 18884493 ↗InjuryOct 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT HAD A REVISED REVERSE SHOULDER FOR PAIN. NO SURGICAL DELAY. DOI: UNKNOWN, DOR: (B)(6) 2023, AFFECTED SIDE: UNKNOWN.
Patient problems: Pain
MAUDE report 17873421 ↗MalfunctionFeb 21, 2023Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE GLENOSPHERE WOULD NOT THREAD INTO THE METAGLENE AND SURGEON FELT SOMETHING WRONG WITH IMPLANT. SURGICAL DELAY IS UNKNOWN. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16409534 ↗InjuryDec 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RE REVISION OF A RIGHT DELTA XTEND DUE TO INSTABILITY. ORIGINAL PRIMARY SURGERY DONE ON THE (B)(6) 2022 AND THEN A 1ST REVISION OF THIS PATIENT WAS DONE ON THE (B)(6) 2022. PATIENT WAS STILL EXPERIENCING INSTABILITY, COMPONENTS CHANGED TO PROVIDE ADDITIONAL STABILITY. DOI: (B)(6) 2022. DOR (1ST): (B)(6) 2022. DOR (2ND): (B)(6) 2022.
Patient problems: Joint Laxity
MAUDE report 16013721 ↗InjuryOct 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: UNKNOWN SHOULDER.
Patient problems: Pain
MAUDE report 15615703 ↗InjurySep 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT REVISED FOR SUSPECTED INFECTION. NO INFECTION IDENTIFIED. GLENOSPHERE AND POLY WERE EXCHANGED. ALL COMPLIMENTS WERE WELL FIXED, NO LOOSENING. NO SURGICAL DELAY. NOTHING FURTHER TO REPORT. DOI: UNKNOWN; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 15510530 ↗InjuryJul 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE HEAD AND LINER WERE EXCHANGED FOR PAIN AND POOR RANGE OF MOTION. SHOULDER WAS DEBRIDED OF SCAR TISSUE AND THE POLY AND GLENOSPHERE WERE EXCHANGED FOR DIFFERENT SIZES TO ACHIEVE STABILITY. THERE WAS NO REPORTED LOOSENING. THERE WAS NO REPORTED SURGICAL DELAY. DOI: (B)(6) 2021. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Adhesion(s); Joint Laxity; Loss of Range of Motion; Pain
MAUDE report 15122286 ↗InjuryJul 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED TO A HEMI DUE TO PAIN. NO SURGICAL DELAY. DOI: UNKNOWN, DOR: (B)(6) 2022, AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Pain
MAUDE report 14948611 ↗MalfunctionJun 30, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT WHEN THE SURGEON PLACING THE GLENOSPHERE ON TO THE STANDARD BASEPLATE, THEY COULD NOT GET THE THREADS OF THE LOCKING POST TO SEAT PROPERLY AFTER. AFTER MULTIPLE ATTEMPTS ADD TRYING ALL THE TRICKS THEY KNEW, LIGHT IMPACT WAS USED. THE THREADS STILL WOULDN'T SEAT AND THEY DECIDED TO REMOVE IT AND START OVER. HOWEVER, WHEN THEY UNSCREWED TO LOCKING POST, IT WOULD NOT COME OFF, MULTIPLE THINGS WERE TRIED TO REMOVE IT SO THEY DECIDED TO CHANGE PATHS AND TRY RESEATING THE THREADS AGAIN. AFTER LIGHT IMPACTION, THE THREADS FINALLY SEATED AND SURGEON ABLE TO COMPLETE THE SURGERY. THERE WAS A SURGICAL DELAY OF 5 MINUTES.
Patient problems: Insufficient Information
MAUDE report 14877604 ↗InjuryJun 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO AN UNEXPLAINED PAIN. NO SIGN OF INFECTION. NO LOOSENING OF COMPONENTS. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Pain
MAUDE report 14828735 ↗InjuryNov 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO PRESENT OF INFECTION. I&D WASHOUT WITH IMPLANT EXCHANGE WERE PERFORMED. ALL COMPONENTS HAD SOLID FIXATION. ORIGINAL IMPLANTATION WAS (B)(6) 2020. SHE HAD ANOTHER WASHOUT IN (B)(6) 2021. DOI: (B)(6) 2020. (B)(6) 2021 (GLENOSPHERE, METAGLENE AND PE CUP DOR: (B)(6) 2021. RIGHT SHOULDER.
Patient problems: Unspecified Infection
MAUDE report 12861781 ↗MalfunctionNov 17, 2021Difficult to Insert; Device Damaged Prior to Use; Difficult to Insert; Device Damaged Prior to Use
AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12829059 ↗MalfunctionNov 16, 2021Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DR. STROUD WAS DID A DELTA REVERSE SHOULDER ON THIS PATIENT AND SELECTED A ECENTRIC 38MM +2 GLENOSPHERE. ONE HE HAD THE IMPLANT POSITIONED OVER THE BASE PLATE, HE STATED HE COULD NOT GET THE THREADS TO ENGAGE. HE MADE SEVERAL ATTEMPTS AND ASKED IF I WOULD OPEN A STD 38MM + 2 GLENOSPHERE. THAT GLENOSPHERE SEATED AND THREADED INTO THE BASE PLATE. THE CASE WAS DELAYED BY 4-5MINUTES DO THE ISSUE. THE 130762038 WAS CLEANED, PUT BACK INTO BOX AND RETURNED TO MY OFFICE.
Patient problems: Insufficient Information
MAUDE report 12821895 ↗InjuryMay 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS REVISED DUE TO BONE GROWTH. THE PATIENT HAD HETEROTOPIC BONE GROWTH THAT HAD GOWN OVER IMPLANTS AND CAUSED IMPINGEMENT. DOI: (B)(6) 2019, DOR: (B)(6) 2021, RIGHT SHOULDER.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Extraskeletal ossification
MAUDE report 11891684 ↗InjuryApr 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION ORIGINAL IMPLANT DATE IS UNKNOWN. DOI: UNKNOWN DOR: (B)(6) 2021 UNKNOWN SHOULDER SIDE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11632085 ↗MalfunctionDec 28, 2020Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT LATERALIZED GLENOSPHERE WOULD NOT TIGHTEN DOWN ON METAGLENE. SURGEON WAS HAVING TROUBLE TIGHTENING THE PRODUCT. NO SURGICAL DELAY. DOI: (B)(6) 2020.
Patient problems: Insufficient Information
MAUDE report 11080201 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page