Skip to content

ATTUNE 2545-06-017Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM ROTATING PLATFORM TIBIAL BASE IMPACTOR · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10603295514312), received between Feb 9, 2022 and May 20, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem6
  • Scratched Material6
  • Break5
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2026Contamination /Decontamination Problem; Material Integrity Problem

    IT WAS REPORTED THAT IMPACTOR WAS DAMAGED, SO CEMENT COULD NOT BE REMOVED FROM THE GROOVES ANYMORE FROM OVER USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25234077 ↗
  • MalfunctionDec 17, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT ARTIFICIAL KNEE REPLACEMENT FOR KNEE OSTEOARTHRITIS USING RP IMPACTORS. DURING THE TIBIA IMPACTION, THE IMPACTORS BROKE. THE DEVICES WERE REMOVED FROM USE AND THE FRAGMENT RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23826842 ↗
  • MalfunctionOct 24, 2025Break; Break

    IT WAS REPORTED THAT A RP IMPACTOR WAS CHIPPED AND NEEDED REPLACING, FROM GENERAL USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379290 ↗
  • MalfunctionFeb 24, 2025Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025. THE INSTRUMENTS WERE BROKEN/SCRATCHED UP.

    Patient problems: Insufficient Information

    MAUDE report 21445474 ↗
  • MalfunctionAug 30, 2024Break; Scratched Material; Break; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS WERE SCRATCHED, DAMAGED, AND BROKEN. THE HOSPITAL REFUSED TO PROCESS DUE TO STERILITY CONCERNS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20106634 ↗
  • MalfunctionJul 24, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE SPD DEPARTMENT/STAFF SAYS THE INSTRUMENTS ARE UNABLE TO BE PROPERLY CLEANED. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 19819171 ↗
  • MalfunctionApr 16, 2024Contamination /Decontamination Problem; Scratched Material; Contamination /Decontamination Problem; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS ARE SCRATCHED AND UNABLE TO BE CLEANED BY SPD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19115664 ↗
  • MalfunctionApr 16, 2024Contamination /Decontamination Problem; Scratched Material; Contamination /Decontamination Problem; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS ARE SCRATCHED AND UNABLE TO BE CLEANED BY SPD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19115644 ↗
  • MalfunctionApr 16, 2024Contamination /Decontamination Problem; Scratched Material; Contamination /Decontamination Problem; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS ARE SCRATCHED AND UNABLE TO BE CLEANED BY SPD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19115611 ↗
  • MalfunctionApr 16, 2024Contamination /Decontamination Problem; Scratched Material; Contamination /Decontamination Problem; Scratched Material

    IT WAS REPORTED THAT INSTRUMENTS ARE SCRATCHED AND UNABLE TO BE CLEANED BY SPD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19115605 ↗
  • MalfunctionFeb 9, 2022Break; Break

    IT WAS REPORTED THAT THE POST ON BUTTON HAS SHEARED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13498120 ↗

Other IMPACTOR devices

All IMPACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page