Skip to content

Delta Xtend 1309-60-315Recalls & reported problems

Depuy (Ireland) · Delta Xtend REVERSE SHOULDER SYSTEM CENTRAL METAGLENE SCREW 6mm x 15mm · Class 2 · Implant

All models of DELTA XTEND (194)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295525202), received between Sep 4, 2025 and Apr 29, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Material Deformation2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Bone Fracture(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 29, 2026Adverse Event Without Identified Device or Use Problem

    SUBJECT ID: (B)(6). STUDY NO: (B)(6). CLINICAL ADVERSE EVENT RECEIVED FOR DISPLACED TYPE 2B SCAPULAR SPINE FRACTURE DEVICE (RELATEDNESS): PROBABLE PROCEDURE (RELATEDNESS): CAUSAL RELATIONSHIP. DATE OF EVENT: (B)(6) 2023. DATE OF IMPLANT: NO INFORMATION PROVIDED. DATE OF REVISION: (B)(6) 2025. DEVICE LOCATION: NO INFORMATION PROVIDED. TREATMENT/IMPACT: SURGICAL INTERVENTION FOR THE REGISTRY SHOULDER: YES DATE OF SURGICAL INTERVENTION: (B)(6) 2026, OPEN REDUCTION INTERNAL FIXATION (ORIF): YES, COMPONENTS REMOVED (MARK NONE OR ALL THAT APPLY.):(* PLEASE COMPLETE A PATIENT COMPLETION/DISCONTINUATION FORM.) NONE: YES . DEPUY SYNTHES PRODUCTS USED: TYPE: SCREW. CATALOG NUMBER: 1309603.15. LOT NUMBER: RL295828 DESCRIPTION: CENTRAL METAGLENE SCREW 6MM X 15MM. TYPE: SCREW. CATALOG NUMBER: 130790024. LOT NUMBER: 5547930. DESCRIPTION: DELTA XTEND LOCKING SCREW 4.5 X 24. TYPE: SCREW. CATALOG NUMBER: 130770018. LOT NUMBER: 5819790. DESCRIPTION: DELTA XTEND NON-LOCKING SCREW 4.5 X 18. TYPE: SCREW. CATALOG NUMBER: 130790024. LOT NUMBER: 5552950. DESCRIPTION: DELTA XTEND LOCKING SCREW 4.5 X 24. TYPE: SCREW. CATALOG NUMBER: 130770018. LOT NUMBER: 5829670. DESCRIPTION: DELTA XTEND NON-LOCKING S

    Patient problems: Bone Fracture(s)

    MAUDE report 25033879 ↗
  • MalfunctionSep 4, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED THAT THERE WERE MAJOR ISSUES WITH THE COLLET NOT ALLOWING GLENOSPHERE SCREW TO THREAD. THE COLLET EXTRACTION TOOL ALSO WOULDN'T FIT INTO THE HOLE. HAD TO USE A DENTAL PICK TO RIP IT OUT. SURGEON WAS FRUSTRATED AND ENDED UP WASTING THE COLLETT, STD CSM BASEPLATE, AND 15 CENTRAL SCREW AND PUTTING IN LEGACY BASEPLATE +10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22961550 ↗
  • MalfunctionSep 4, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED THAT THERE WERE MAJOR ISSUES WITH THE COLLET NOT ALLOWING GLENOSPHERE SCREW TO THREAD. THE COLLET EXTRACTION TOOL ALSO WOULDN'T FIT INTO THE HOLE. HAD TO USE A DENTAL PICK TO RIP IT OUT. SURGEON WAS FRUSTRATED AND ENDED UP WASTING THE COLLETT, STD CSM BASEPLATE, AND 15 CENTRAL SCREW AND PUTTING IN LEGACY BASEPLATE +10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22961547 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page