Skip to content

RECLAIM 34-185-0014Recalls & reported problems

Depuy (Ireland) · RECLAIM MONOBLOC – REVISION HIP SYSTEM MONOBLOC DISTAL REAMER STANDARD 14 · Class 1

All models of RECLAIM (149)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295536093), received between Aug 2, 2024 and Apr 27, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Entrapment of Device1
  • Improper or Incorrect Procedure or Method1
  • Material Integrity Problem1
  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2026Material Integrity Problem

    IT WAS REPORTED THAT DURING INSPECTION, THE REPORTER FOUND REAMERS TO BE DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24999385 ↗
  • InjuryJan 3, 2025Entrapment of Device; Off-Label Use; Improper or Incorrect Procedure or Method; Device-Device Incompatibility; Entrapment of Device; Off-Label Use; Improper or Incorrect Procedure or Method; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A HIP FX CASE, THE SURGEON BEGAN USING RECLAIM MONOBLOC INSTRUMENTS PER TECHNIQUE. THE SURGEON USED THE ENTRY REAMER FOLLOWED BY THE 12MM STANDARD DISTAL REAMER, 13MM STANDARD DISTAL REAMER, AND 14MM STANDARD DISTAL REAMER. THE 14MM STANDARD DISTAL REAMER HAD GOOD CORTICAL ENGAGEMENT AND WAS USED FOR TRIALING. SURGEON PROCEEDED TO USE THE PROXIMAL REAMERS SEQUENTIALLY STARTING WITH THE 12MM STANDARD PROXIMAL REAMER, THE 13MM STANDARD PROXIMAL REAMER, AND FINALLY THE 14MM STANDARD PROXIMAL REAMER. SURGEON ATTACHED THE STANDARD TRIAL NECK AND THE TRIAL 36MM X 1.5MM HEAD. SURGEON REDUCED THE HIP AND PERFORMED RANGE OF MOTION. THE SURGEON REQUESTED IMPLANTS TO MATCH THE TRIALS. SURGEON REMOVED THE HEAD TRIAL AND THE NECK TRIAL. SURGEON BEGAN THE REMOVAL OF THE 14MM STANDARD DISTAL REAMER AND WAS NOT ABLE TO REMOVE IT. SURGEON PROCEEDED TO USE THE POWER REAMER AND WAS NOT ABLE TO REMOVE THE 14MM STANDARD DISTAL REAMER. SURGEON TRIED THE SLAP HAMMER AND WAS UNSUCCESSFUL. SURGEON THEN TRIED THE T-HANDLE FROM THE RECLAIM MONOBLOC INSTRUMENT SET UNSUCCESSFULLY. THE SURGEON DECIDED TO LEAVE THE 14MM STANDARD DISTAL REAMER IN PATIENT, REATTACH THE STANDARD NECK TRI

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 21067572 ↗
  • MalfunctionAug 2, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SURGEON BENT SCREWDRIVER FAILING TO DISCONNECT PROXIMAL GUIDE POST FROM DISTAL REAMER USING RECLAIM MONOBLOC. SHORT DELAY IN SURGERY CAUSING SURGEON TO SWITCH SYSTEMS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19891092 ↗

Other REAMER devices

All REAMER devices →

Sources and updates