Skip to content

INHANCE 5200-01-044Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM SHORT STEM X-LARGE diameter 44mm X 82mm · Class 2 · Implant

All models of INHANCE (200)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295538226), received between Dec 9, 2022 and Jun 12, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Loss of or Failure to Bond1
  • Osseointegration Problem1
  • Use of Device Problem1

Patient problems

  • Bone Fracture(s)4
  • Inadequate Osseointegration1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2026Use of Device Problem

    IT WAS REPORTED THAT UPON THE PATIENT LEAVING THE OPERATING ROOM, A POST-OPERATIVE X-RAY REVEALED THE HUMERAL STEM WAS POSITIONED IN VARUS AND HAD PERFORATED THE LATERAL HUMERUS. DOE: (B)(6) 2026. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Bone Fracture(s)

    MAUDE report 25488461 ↗
  • InjuryMay 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT COMES IN TODAY HAVING PROBLEMS WITH CHRONIC DISLOCATIONS. THE PATIENT LIFTED A HEAVY BOX 4 WEEKS POST OPERATION AND DISLOCATED CAUSING MORE PROBLEMS. THE DOCTOR DECIDED IT BEST TO REMOVE HIS GLENOSPHERE AND POLY TO REBALANCE THE SHOULDER. WHILE DOING THIS THE PROXIMAL HUMERUS FRACTURED SO HE REMOVED THE STEM AND SHELL AS WELL. AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Bone Fracture(s)

    MAUDE report 25056097 ↗
  • InjuryJan 15, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT HUMERAL FRACTURE AND LONGER STEM WAS REIMPLEMENTED. DOI: UNKNOWN. DOR: (B)(6) 2026. AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Bone Fracture(s)

    MAUDE report 24083027 ↗
  • InjuryJan 30, 2024Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT PATIENT HAD INITIAL SURGERY DONE BY THE SURGEON AT THE HOSPITAL ON (B)(6) 2023. PATIENT WAS IMPLANTED WITH DEPUY HUMERAL COMPONENTS, AND BIOMET AUGMENTED GLENOSPHERE. THE GLENOSPHERE BECAME LOSE AND CAME COMPLETELY DISLODGED FROM THE GLENOID. SURGEON WANTED TO CONVERT THE SHOULDER TO A HEMI AND FELT THE HUMERAL SIDE WAS LOSE, SO SURGEON COMPLETELY REMOVED THE DEPUY HUMERAL SIDE AND ENDED UP IMPLANTING A ZIMMER BIOMET HEMI SHOULDER IMPLANT. NO ADVERSE EVENTS TOOK PLACE. DOI: (B)(6) 2023 .DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18606171 ↗
  • InjuryDec 9, 2022Osseointegration Problem; Osseointegration Problem

    PATIENT WAS REVISED DUE TO PERIPROSTHETIC FRACTURE. NEEDED LONGER STEM. NO ADDITIONAL INFORMATION. THERE WAS A LOOSENING AND LOOSENING INTERFACE WAS NON-CEMENTED. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: Bone Fracture(s); Inadequate Osseointegration

    MAUDE report 15952967 ↗

FDA clearance and classification

The FDA regulates product code MBF (prosthesis, shoulder, semi-constrained, metal/polymer, uncemented) under 21 CFR 888.3670 as a Class II device. 21 CFR 888.3670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the INHANCE 5200-01-044

Has the INHANCE 5200-01-044 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10603295538226), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the INHANCE 5200-01-044?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Dec 9, 2022 and Jun 12, 2026. In those reports, the most common specific device problems are “Loss of or Failure to Bond” (1), “Osseointegration Problem” (1) and “Use of Device Problem” (1); the most common patient problems are “Bone Fracture(s)” (4) and “Inadequate Osseointegration” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the INHANCE 5200-01-044?

None of the reports are classified as deaths, 5 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the INHANCE 5200-01-044 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the INHANCE 5200-01-044, and is it still sold?

The FDA's device database lists Depuy (Ireland) as the company responsible for it (brand INHANCE, model 5200-01-044). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page