Skip to content

INHANCE 6201-01-141Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM CANNULATED IMPACTOR HANDLE · Class 1

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy INHANCE Shoulder System cannulated impactor handle has 41 adverse event reports that name its identifier, all describing malfunctions, with no injuries or deaths. Reports mostly describe stripped or broken threads, worn parts, and pieces not disconnecting after washing, sometimes noticed outside surgery. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10603295539339), received between Mar 21, 2023 and Jun 22, 2026 — classified as 40 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 4 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn12
  • Material Deformation9
  • Break8
  • Device-Device Incompatibility8
  • Material Integrity Problem2
  • No Apparent Adverse Event2
  • Corroded1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions30
  • Insufficient Information10

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2026Material Deformation

    IT WAS REPORTED THAT IT WAS NOTICED THAT THE THREADS OF THE DEVICE WERE DAMAGED, THEREFORE THE DEVICE WAS REMOVED FROM THE USE. THIS DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25577404 ↗
  • MalfunctionJun 5, 2026Break

    IT WAS REPORTED THAT BROKEN THREADS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414466 ↗
  • MalfunctionFeb 9, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT IT WAS NOTICED THAT THE TIP OF THE HANDLE LOOKS WORN. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24291906 ↗
  • MalfunctionJan 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLE AND IMPACTOR TIP WOULD NOT DISCONNECT FROM EACH OTHER AFTER WASHING. THE ISSUE DID NOT OCCUR DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24120276 ↗
  • MalfunctionJan 16, 2026Break

    INSTRUMENTS WERE REPORTED FOR UNKNOWN REASON. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24092403 ↗
  • MalfunctionJan 15, 2026Material Deformation

    IT WAS REPORTED THAT THE THREADS ON THE IMPACTOR HANDLE WERE STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24087341 ↗
  • MalfunctionJan 13, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT WHILE REASSEMBLING THE INSTRUMENT, THE BLAZER WAS COLD-WELDED TO THE IMPACTOR HANDLE. THE EVENT DID NOT OCCUR DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24054318 ↗
  • MalfunctionDec 26, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE IMPACTOR HANDLE AND HEAD DISTRACTOR BECAME FUSED TOGETHER DURING SURGERY AND COULD NOT BE SEPARATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23897773 ↗
  • MalfunctionDec 18, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE FOUND BROKEN IN TRAY PRIOR TO GOING TO SPD FOR NEXT CASE. THE CASE DID NOT HAPPENED DURING A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838485 ↗
  • MalfunctionDec 9, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE IMPACTOR HANDLE THREADS/ TIP DAMAGED. BACK UP INSTRUMENTS ON HAND, NO ADDITIONAL SURGERY TIME WAS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23757125 ↗
  • MalfunctionOct 31, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE THREADS WERE STRIPPED ON THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23437471 ↗
  • MalfunctionOct 22, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE STARDRIVER 20 WAS STRIPPED AND COULD NOT TIGHTEN THE SCREW FULLY. THERE WAS A MINOR DELAY OF 10 MINUTES TO THE TOTAL SHOULDER REPLACEMENT CASE TO FIND A REPLACEMENT DRIVER. UPON DECONTAMINATION OF AN IMPACTOR, IT WAS NOTED THAT THE THREADS ON IMPACTOR HAD BEEN DAMAGED. THERE WAS NO DELAY TO THE CASE AS IT WAS FOUND POST OPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23359253 ↗
  • MalfunctionAug 1, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENTS WERE BROKEN WHEN USED IN CONJUNCTION AS A BACK SLAP DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22690641 ↗
  • MalfunctionJul 3, 2025Corroded; Mechanical Jam; Corroded; Mechanical Jam

    IT WAS REPORTED THAT CANNULA WAS CORRODING IN CANNULATION. PIN WAS STUCK INSIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22394129 ↗
  • MalfunctionJun 26, 2025Break; Break

    IT WAS REPORTED THAT THE SCREWDRIVERS ARE STRIPPED, THE INSERTER HANDLE BROKE, AND THE GLENOID REAMER BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22341298 ↗
  • MalfunctionJun 4, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE IMPACTOR WAS STRIPPED. THERE WAS NO SURGICAL DELAY OR PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22142040 ↗
  • MalfunctionApr 30, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PRIOR TO THE CASE STARTING THE SCRUB TECH NOTICED THAT THE CANNULATED IMPACTOR HANDLE WAS NOT THREADING INTO THE BLAZERS. ANOTHER WAS AVAILABLE AND THE SURGERY PROCEEDED AS PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936009 ↗
  • MalfunctionApr 17, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING ROUTINE MEDICAL DEVICE REPROCESSING, MEDICAL DEVICE REPROCESSING STAFF IDENTIFIED THAT THE THREADS ON THE TIPS OF COMPLAINED ITEMS HAD BECOME DAMAGED AND STRIPPED. THESE COMPLAINED INSTRUMENTS WITH THE DAMAGED TIPS WERE DEEMED TO BE NO LONGER FUNCTIONAL, AND WERE SUBSEQUENTLY DISCARDED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21861802 ↗
  • MalfunctionApr 17, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING ROUTINE MEDICAL DEVICE REPROCESSING, MEDICAL DEVICE REPROCESSING STAFF IDENTIFIED THAT THE THREADS ON THE TIPS OF COMPLAINED ITEMS HAD BECOME DAMAGED AND STRIPPED. THESE COMPLAINED INSTRUMENTS WITH THE DAMAGED TIPS WERE DEEMED TO BE NO LONGER FUNCTIONAL, AND WERE SUBSEQUENTLY DISCARDED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21857390 ↗
  • MalfunctionApr 2, 2025Break; Break

    IT WAS REPORTED THAT THE CANNULATED IMPACTOR HANDLE TIP BROKE AT THE THREADS.

    Patient problems: Insufficient Information

    MAUDE report 21756614 ↗
  • MalfunctionMar 28, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT TWO HANDLES WERE FOUND STRIPPED IN THE SET.

    Patient problems: Insufficient Information

    MAUDE report 21723841 ↗
  • MalfunctionMar 28, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT TWO HANDLES WERE FOUND STRIPPED IN THE SET.

    Patient problems: Insufficient Information

    MAUDE report 21723827 ↗
  • MalfunctionMar 11, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE HANDLE THREADS WERE STRIPPED AND NEEDED TO BE REPLACED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 21575399 ↗
  • MalfunctionJan 15, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TIP OF THE LARGE BOWTIE GLENOID REAMER BROKE WHILE REAMING THE GLENOID. FRAGMENTS WERE GENERATED AND WERE RETRIEVED. THE LARGE BLAZER FUSED ONTO THE CANNULATED HANDLE. THERE WAS NO MISHANDLING OF THE PRODUCT BEFORE OR AFTER ASSEMBLING THE TWO ITEMS. THERE WAS ABOUT A 20 MINUTE TIME DELAY TO TRY TO REMOVE THE BLAZER FROM THE CANNULATED HANDLE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21158362 ↗
  • MalfunctionDec 5, 2024Break; Break

    ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20858680 ↗

Other IMPACTOR devices

All IMPACTOR devices →

Sources and updates