INHANCE 5503-00-020Recalls & reported problems
Depuy (Ireland) · INHANCE SHOULDER SYSTEM NON-LOCKING SCREW 20mm · Class 2 · Implant
All models of INHANCE (202)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295546849), received between Oct 31, 2022 and Mar 17, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Insert4
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Joint Laxity1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem
SUBJECT ID: (B)(6). STUDY NO: (B)(6). CLINICAL ADVERSE EVENT RECEIVED THE PATIENT EXPERIENCED PERSISTENT INSTABILITY IN THE LEFT SHOULDER FOLLOWING A REVERSE TOTAL SHOULDER ARTHROPLASTY, DESPITE INITIAL POSTOPERATIVE IMPROVEMENT. IMAGING CONFIRMED STABLE ALIGNMENT OF THE IMPLANT, AND THE INSTABILITY DID NOT RESULT IN A TRUE GLENOHUMERAL DISLOCATION. PERSISTENT SYMPTOMS LED TO A DIAGNOSTIC CT SCAN TO ASSESS PROSTHESIS STABILITY AND POSITION. DEVICE RELATED: POSSIBLE (HUMERAL STEMLESS IMPLANT: YES); PROCEDURE RELATED: POSSIBLE; DATE OF EVENT: (B)(6) 2026; DATE OF IMPLANT: (B)(6) 2025; DATE OF REVISION: NO INFORMATION PROVIDED; DEVICE LOCATION: LEFT. TREATMENT/IMPACT: DIAGNOSTIC INTERVENTION (CT). DEPUY SYNTHES PRODUCTS USED: CATALOG NUMBER: 520000036; LOT NUMBER: 662988; COMPONENT TYPE: HUMERAL; DESCRIPTION: INHANCE SHOULDER SYSTEM STEMLESS MEDIUM DIAMETER 36MM CEMENTLESS. CATALOG NUMBER: 550011240; LOT NUMBER: 664496; COMPONENT TYPE: : BASEPLATE; DESCRIPTION: INHANCE SHOULDER SYSTEM MODULAR UNITI PLATFORM BASEPLATE SMALL Ø24MM CEMENTLESS. CATALOG NUMBER: 550035600; LOT NUMBER: 206439; COMPONENT TYPE: SCREW; DESCRIPTION: INHANCE SHOULDER SYSTEM CENTRAL SCREW Ø6.0 X 35MM. CATALOG NUMB
Patient problems: Joint Laxity; Pain
MAUDE report 24616505 ↗MalfunctionDec 28, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE SCREW PENETRATED THE MODULAR BASEPLATE DURING REVERSE SHOULDER INHANCE PROCEDURE. 15MIN SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16062306 ↗MalfunctionDec 20, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE SURGEON HAD ANOTHER SCREW GO THROUGH A BASEPLATE. THE INCIDENT INVOLVED A SELF-DRILLING SCREW AND A LARGE BASEPLATE. HE DID NOT USE A NON-LOCKING SCREW PRIOR TO USING THE SELF-DRILLING SCREW. WHEN THINKING ABOUT WHY THE SURGEON IS THE ONLY ONE SEEING THIS ISSUE, IT WAS HYPOTHESIZED THAT HE IS DAMAGING THE THREADS WITH THE BASEPLATE IMPACTOR HANDLE WHEN IT IS USED WITHOUT THE INSERTER ROD. THE SURGEON'S TECHNIQUE IS AS FOLLOWS: INSERT THE BASEPLATE WITH THE BASEPLATE INSERTER ROD, WITHOUT THE IMPACTOR HANDLE ASSEMBLED REMOVE THE BASEPLATE INSERTER ROD AND USE THE IMPACTOR HANDLE TO SEAT THE BASEPLATE. OTHER SURGEONS REPORT THAT THEY USE THE ENTIRE CONSTRUCT (IMPACTOR HANDLE AND ROD) TO SEAT THE GLENOID. ANOTHER SURGEON ONLY USES THE IMPACTOR ROD TO SEAT THE GLENOID. THIS IS A SHOULDER RECONSTRUCTION INHANCE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16013998 ↗MalfunctionOct 31, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE SURGEON HAD IMPLANTED A SMALL BASEPLATE WITH A 25MM CENTRAL SCREW. HE WAS LOOKING TO GAIN SOME ADDITIONAL COMPRESSION, SO HE ASKED FOR THE PERIPHERAL DRILL SO HE COULD DRILL FOR SOME NON LOCKING SCREWS. THE DRILL GUIDE WAS USED, AND THEN THE SURGEON MEASURED FOR A 20MM SCREW. WHEN IMPLANTING THE SCREW, THE SURGEON SAID THAT THE SCREW WENT DIRECTLY THROUGH THE BASEPLATE. HE ASKED FOR A DIFFERENT 20MM NON-LOCKING SCREW, AND THAT SCREW ALSO WENT THROUGH THE BASEPLATE. AT THIS POINT, THE SURGEON OPTED TO IMPLANT LOCKING SCREWS, SO HE ASKED FOR THE SELF DRILLING & LOCKING SCREW PACK. WHEN IMPLANTING ONE OF THE LOCKING SCREWS, THAT ALSO WENT THROUGH THE BASEPLATE. WHEN THE SURGEON ATTEMPTED TO BACK THE LOCKING SCREW OUT, HE INFORMED ME THAT IT MUST HAVE CROSS THREADED ITSELF INTO THE PLATE, WHERE IT EVENTUALLY LOCKED INTO THE PLATE. HE ASKED THAT I WRITE THESE IMPLANTS UP AND HAVE THE SCREWS RETURNED TO BE ANALYZED. HE ALSO MENTIONED TO ME HIS DISAPPOINTMENT IN THE ¿PICK & STICK¿ ASPECT OF THE PERIPHERAL SCREWS. HE SAID THAT THEY JUST DON¿T SEEM TO LOCK ONTO THE T20 SCREWDRIVER THE WAY THAT THEY SHOULD. THERE WAS NO REPORTED LOOSENING. THERE WAS A REPORTED SURGIC
Patient problems: Insufficient Information
MAUDE report 15698614 ↗MalfunctionOct 31, 2022Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THE SURGEON HAD IMPLANTED A SMALL BASEPLATE WITH A 25MM CENTRAL SCREW. HE WAS LOOKING TO GAIN SOME ADDITIONAL COMPRESSION, SO HE ASKED FOR THE PERIPHERAL DRILL SO HE COULD DRILL FOR SOME NON LOCKING SCREWS. THE DRILL GUIDE WAS USED, AND THEN THE SURGEON MEASURED FOR A 20MM SCREW. WHEN IMPLANTING THE SCREW, THE SURGEON SAID THAT THE SCREW WENT DIRECTLY THROUGH THE BASEPLATE. HE ASKED FOR A DIFFERENT 20MM NON-LOCKING SCREW, AND THAT SCREW ALSO WENT THROUGH THE BASEPLATE. AT THIS POINT, THE SURGEON OPTED TO IMPLANT LOCKING SCREWS, SO HE ASKED FOR THE SELF DRILLING & LOCKING SCREW PACK. WHEN IMPLANTING ONE OF THE LOCKING SCREWS, THAT ALSO WENT THROUGH THE BASEPLATE. WHEN THE SURGEON ATTEMPTED TO BACK THE LOCKING SCREW OUT, HE INFORMED ME THAT IT MUST HAVE CROSS THREADED ITSELF INTO THE PLATE, WHERE IT EVENTUALLY LOCKED INTO THE PLATE. HE ASKED THAT I WRITE THESE IMPLANTS UP AND HAVE THE SCREWS RETURNED TO BE ANALYZED. HE ALSO MENTIONED TO ME HIS DISAPPOINTMENT IN THE ¿PICK & STICK¿ ASPECT OF THE PERIPHERAL SCREWS. HE SAID THAT THEY JUST DON¿T SEEM TO LOCK ONTO THE T20 SCREWDRIVER THE WAY THAT THEY SHOULD. THERE WAS NO REPORTED LOOSENING. THERE WAS A REPORTED SURGIC
Patient problems: Insufficient Information
MAUDE report 15698612 ↗
Other prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented devices
- Comprehensive Reverse Shoulder 115310Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 death99 reports
- INHANCE 5120-00-110Depuy (Ireland) · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented75 reports
- Comprehensive® Reverse Shoulder XL-115364Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 recall7 reports
- Comprehensive® Reverse Shoulder XL-115366Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 recall4 reports
- Comprehensive® Reverse Shoulder 180550Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented2 deaths43 reports
- Comprehensive® Reverse Shoulder 180551Biomet Orthopedics, LLC · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented42 reports
- DJO SURGICAL 521-07-246Encore Medical, L.P. · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented42 reports
- Comprehensive® 110031399Zimmer, Inc. · prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented1 death36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page