Skip to content

INHANCE 5507-06-100Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM KICKSTAND Augment 6mm CEMENTLESS · Class 2 · Implant

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10603295546948), received between Feb 6, 2024 and Aug 27, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert6
  • Naturally Worn2

Patient problems

  • Insufficient Information4
  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE THREAD ON THE KICKSTAND WOULD NOT LOCK IN THE BASEPLATE. DOE: (B)(6) 2025. AFFECTED SIDE: UNKNOWN SHOULDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22906733 ↗
  • MalfunctionOct 23, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE LEFT REVERSE TSA FOR HUMERAL SIDED REVISION, GLENOID EROSION. THE ENHANCE GLENOID WAS PREPPED FOR A 6MM KICKSTAND VIA SURGICAL TECHNIQUE. KICKSTAND IMPLANT WOULD NOT SCREW INTO BASEPLATE. ANOTHER KICKSTAND IMPLANT WAS OPENED AND INSERTED AFTER REDRILLING FOR A 6MM AUGMENT.

    Patient problems: Insufficient Information

    MAUDE report 20513389 ↗
  • MalfunctionSep 6, 2024Difficult to Insert; Naturally Worn; Difficult to Insert; Naturally Worn

    ADDITIONAL INFORMATION WAS RECEIVED: A. WAS THERE ANY PATIENT HARM RELATED TO THE EVENT? NO PATIENT HARM DUE TO EVENT. B. WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? YES, THERE WAS A 15 MINUTE DELAY DUE TO EVENT. C. IS THERE ANY PATIENT CONSEQUENCE AGAINST THE REPORTED PRODUCT? YES, AUGMENT WAS UNABLE TO BE USED. AS A RESULT, SURGEON ESTIMATED THAT GLENOID COVERAGE WAS AT APPROXIMATELY 80%, THUS WARRANTING THE NEED FOR AUGMENTATION THAT WAS UNFORTUNATELY UNABLE TO BE PERFORMED. D. WAS THERE ANY ALLEGATION WITH THE BASEPLATE/METAGLENE? NO ALLEGATION WITH THE BASEPLATE. THE CENTRAL SCREW AND PERIMETER LOCKING SCREWS THAT WERE USED WENT IN WITH NO ISSUES AND APPEARED TO ENGAGE PROPERLY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20155645 ↗
  • MalfunctionAug 15, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE AUGMENT ON DEMAND WAS PREPPED VIA SURGICAL PROTOCOL. KICKSTAND IMPLANT WOULD NOT ENGAGE THREADS OF IMPLANT. SECOND IMPLANT WAS PUT IN SUCCESSFULLY.

    Patient problems: Insufficient Information

    MAUDE report 19990086 ↗
  • MalfunctionJul 2, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE SCREW STRIPPED OUT. AFTER 2 ATTEMPTS, A NEW SCREW WAS USED AND PROCEDURE WAS COMPLETED SUCCESSFULLY. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER.

    Patient problems: Insufficient Information

    MAUDE report 19656299 ↗
  • MalfunctionMay 2, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE PATIENT HAD A RIGHT RTSA, SURGEON COULD NOT GET KICKSTAND IMPLANT TO DEFINITIVELY LOCK IN BASEPLATE. REPORTED A 5 MINUTE DELAY AS THEY ATTEMPTED TO RE-DRILL, RE INSERT AND FINALLY OPENED ANOTHER IMPLANT WHICH LOCKED INTO THE BASE PLATE ON FIRST ATTEMPT. DOE: (B)(6) 2024 AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228391 ↗
  • MalfunctionFeb 6, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE AUGMENT WOULD NOT SCREW INTO THE BASEPLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions