Skip to content

INHANCE 6500-11-102Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM BASEPLATE INSERTER ROD · Class 1

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy INHANCE shoulder baseplate inserter rod is a surgical tool used in shoulder surgery. Of its 565 reports, all describe malfunctions, most often deformed material, wear and poor fit with other devices. Reports describe tools stripped during shoulder surgery, needing replacement, and cracked handles. No deaths or injuries are recorded, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
567
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 567 adverse event reports that list this exact device (DI 10603295547235), received between Sep 27, 2022 and Sep 18, 2026 — classified as 567 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (251 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 19 reportsDec 2023: 14 reportsJan 2024: 10 reportsFeb 2024: 19 reportsMar 2024: 17 reportsApr 2024: 34 reportsMay 2024: 13 reportsJun 2024: 12 reportsJul 2024: 9 reportsAug 2024: 14 reportsSep 2024: 19 reportsOct 2024: 22 reportsNov 2024: 16 reportsDec 2024: 7 reportsJan 2025: 22 reportsFeb 2025: 15 reportsMar 2025: 22 reportsApr 2025: 20 reportsMay 2025: 11 reportsJun 2025: 16 reportsJul 2025: 10 reportsAug 2025: 17 reportsSep 2025: 3 reportsOct 2025: 10 reportsNov 2025: 13 reportsDec 2025: 20 reportsJan 2026: 17 reportsFeb 2026: 15 reportsMar 2026: 19 reportsApr 2026: 16 reportsMay 2026: 11 reportsJun 2026: 12 reportsJul 2026: 4 reportsAug 2026: 0 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202634
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation251
  • Naturally Worn218
  • Device-Device Incompatibility208
  • Break43
  • No Apparent Adverse Event24
  • Material Integrity Problem12
  • Mechanical Jam5
  • Illegible Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions465
  • Insufficient Information117

Recent adverse event reports

The 25 most recent of 567 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Device-Device Incompatibility

    It was reported that upon inspection of the 32 +0 trial poly, one of the locking tabs was missing. Another set was opened so the surgeon could trial the range of motion. No additional time was added to the surgery, and the broken trial was removed. Another instrument in the tray still had a baseplate attached from a removal surgery performed the week prior. The implant could not be removed, so SPD processed the instrument assembled. The tray was replaced with another tray that was available at the account. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819597 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Material Deformation

    It was reported that while doing a total shoulder replacement with the Inhance instruments, the glenosphere inserter and tip would not screw together. Surgeon believed the threads on each instrument were stripped so he opened another set of instruments and the same thing happened so he need two each of the inserter handle and the inserter tips.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721057 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50720996 ↗
  • MalfunctionSep 1, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549113 ↗
  • MalfunctionSep 1, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50548356 ↗
  • MalfunctionSep 1, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50543706 ↗
  • MalfunctionJul 13, 2026Device-Device Incompatibility; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25803289 ↗
  • MalfunctionJul 10, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25785475 ↗
  • MalfunctionJul 6, 2026Material Deformation

    IT WAS REPORTED THAT THE DEVICE WAS STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716682 ↗
  • MalfunctionJul 2, 2026Material Deformation

    IT WAS REPORTED THAT DURING A SHOULDER REPAIR, BOTH DEVICES WERE STRIPPED AND NEED TO BE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689971 ↗
  • MalfunctionJun 19, 2026Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE THREADS WERE MESSED UP. PIECES WOULD NOT SCREW TOGETHER, POSSIBLY CROSS THREADED. THE ISSUE WAS CAUGHT BEFORE THE CASE STARTED SO THERE WERE NO DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565896 ↗
  • MalfunctionJun 18, 2026Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT THE DEVICE WAS CROSS THREADED AND DAMAGED PIECES WERE OBSERVED. IT WAS REPORTED THAT THE EVENT DID NOT OCCUR IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25555695 ↗
  • MalfunctionJun 17, 2026Material Deformation

    IT WAS REPORTED THAT THE INSTRUMENT HAD AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN IN SURGERY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE TWO PIECES WOULD NOT THREAD TOGETHER CROSSTHREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25536000 ↗
  • MalfunctionJun 17, 2026Material Deformation

    IT WAS REPORTED THAT THE INSTRUMENT HAD AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN IN SURGERY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE TWO PIECES WOULD NOT THREAD TOGETHER CROSSTHREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25535994 ↗
  • MalfunctionJun 11, 2026Material Deformation

    IT WAS REPORTED THAT THE T20 SCREWDRIVER BROKEN. T30 SCREWDRIVER STRIPPED. IMPACTOR HANDLE STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477961 ↗
  • MalfunctionJun 8, 2026No Apparent Adverse Event

    IT WAS REPORTED IT WAS NOTICED CONSIDERABLE WEAR ON THE INSTRUMENTS. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25429994 ↗
  • MalfunctionJun 5, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE INHANCE T-HANDLE, THREADS ARE BAD. BASEPLATE INSERTER HANDLE, PRIOR TO SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414015 ↗
  • MalfunctionJun 2, 2026Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT DRIVER SHAFT WAS CRACKED AT THE DISTAL TIP NEAR COMPRESSION RETAINER CLIPS. IT WAS FURTHER OBSERVED THAT THE T-HANDLE AND GLENO INSERTER TIP CROSS THREADED, WAS ABLE TO SEPARATE, BUT WOULD NOT RE-SEAT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365351 ↗
  • MalfunctionJun 2, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREWDRIVER WAS STRIPPED AND THE ADAPTOR AND IMPACTOR HANDLE WERE STUCK TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365163 ↗
  • MalfunctionJun 1, 2026Material Deformation

    IT WAS REPORTED THAT THE GLENOSPHERE INSERTER TIP THREADS STRIPPED T-HANDLE MODULAR GLENOIND BASE PLATE AND GLENOSHERE INSERTER THREADS STRIPPED. ADDITIONALLY, 2X STAR 20 SCREW DRIVERS WERE STRIPPED. ALL DEVICES ARE INTACT. NO DELAYS AS THE REPORTER HAVE BACKUP. NO HARM TO PATIENT OR STAFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351105 ↗
  • MalfunctionJun 1, 2026Material Deformation

    IT WAS REPORTED THAT THREADS ARE STRIPPED ON INSERTER. REAMER SHAFT WILL NOT HOLD REAMER. CUT GUIDE PRODUCES METAL SHAVING WHEN PLACING PINS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25350708 ↗
  • MalfunctionJun 1, 2026Material Deformation

    IT WAS REPORTED THAT THREADS ARE STRIPPED ON INSERTER. REAMER SHAFT WILL NOT HOLD REAMER. CUT GUIDE PRODUCES METAL SHAVING WHEN PLACING PINS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25350707 ↗
  • MalfunctionMay 27, 2026Material Deformation

    IT WAS REPORTED THAT THE SCREWDRIVERS HANDLE AND THE INSERTER TIP WERE STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25303577 ↗
  • MalfunctionMay 25, 2026Break; Material Deformation

    IT WAS REPORTED THAT THREADS WERE DAMAGED/CROSS THREADED, AND THE PLASTIC PIECE BROKE OFF. THIS EVENT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25270801 ↗
  • MalfunctionMay 22, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE INSTRUMENT HAD UNKNOWN ALLEGATION. IT WAS FURTHER REPORTED THAT WHILE CHECKING TRAYS AT THE OFFICE, THE TWO PIECES WOULD NOT THREAD TOGETHER. THE FAILURE DID NOT OCCUR DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256019 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates