Skip to content

INHANCE 6500-11-104Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM GLENOSPHERE INSERTER TIP · Class 1

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Inhance inserter tip is a surgical tool used to place parts during shoulder surgery. Its 826 reports describe malfunctions, with no injuries reported. Many describe threads that become cross-threaded or stripped, and parts that wear, deform or break. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
831
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 831 adverse event reports that list this exact device (DI 10603295547266), received between Aug 10, 2022 and Sep 18, 2026 — classified as 831 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (343 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 28 reportsDec 2023: 24 reportsJan 2024: 25 reportsFeb 2024: 25 reportsMar 2024: 26 reportsApr 2024: 34 reportsMay 2024: 22 reportsJun 2024: 32 reportsJul 2024: 16 reportsAug 2024: 15 reportsSep 2024: 22 reportsOct 2024: 25 reportsNov 2024: 23 reportsDec 2024: 7 reportsJan 2025: 20 reportsFeb 2025: 15 reportsMar 2025: 25 reportsApr 2025: 18 reportsMay 2025: 24 reportsJun 2025: 26 reportsJul 2025: 13 reportsAug 2025: 18 reportsSep 2025: 10 reportsOct 2025: 11 reportsNov 2025: 16 reportsDec 2025: 33 reportsJan 2026: 30 reportsFeb 2026: 26 reportsMar 2026: 23 reportsApr 2026: 27 reportsMay 2026: 14 reportsJun 2026: 15 reportsJul 2026: 3 reportsAug 2026: 4 reportsSep 2026: 10 reportsOct 2026: 0 reports20242025202634
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn343
  • Device-Device Incompatibility312
  • Material Deformation312
  • Break92
  • No Apparent Adverse Event29
  • Material Integrity Problem12
  • Entrapment of Device3
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions688
  • Insufficient Information158
  • Foreign Body In Patient2
  • Bone Notching1

Recent adverse event reports

The 25 most recent of 831 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Break; Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819968 ↗
  • MalfunctionSep 17, 2026Device-Device Incompatibility; Material Deformation

    It was reported that during a Total Shoulder Replacement surgical procedure, after the insertion of the glenosphere, the baseplate rod and glenosphere inserter tip were cross threaded, and unable to be separated. There was no harm to the patient, no delay in the case.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50800536 ↗
  • MalfunctionSep 16, 2026No Apparent Adverse Event

    It was reported that the instrument had unknown allegation. It did not happen in surgery.During manufacturer's investigation of the returned device, it was identified that the distal threaded tip broken.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50789647 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721032 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50720855 ↗
  • MalfunctionSep 1, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50549123 ↗
  • MalfunctionSep 1, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50548573 ↗
  • MalfunctionSep 1, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50543511 ↗
  • MalfunctionSep 1, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50543433 ↗
  • MalfunctionSep 1, 2026No Apparent Adverse Event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50542498 ↗
  • MalfunctionAug 13, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50241404 ↗
  • MalfunctionAug 13, 2026Material Deformation

    It was reported that the items have non functioning threads. Additional information received:Cross threaded and/or stripped.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50233508 ↗
  • MalfunctionAug 13, 2026Material Deformation

    It was reported that the items have non functioning threads. Additional information received:Cross threaded and/or stripped.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50233505 ↗
  • MalfunctionAug 11, 2026Material Deformation

    It was reported that during a surgical procedure for shoulder repair it was observed that the threads on the glenoid inserter and the Glenosphere inserter became cross threaded and had stripped.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50212378 ↗
  • MalfunctionJul 13, 2026Device-Device Incompatibility; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25803288 ↗
  • MalfunctionJul 10, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25785467 ↗
  • MalfunctionJul 6, 2026Material Deformation

    IT WAS REPORTED THAT THE DEVICE WAS STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716676 ↗
  • MalfunctionJun 30, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE GLENOSPHERE INSERTOR BOLT WOULD NOT THREAD INTO GLENOSPHERE. THEY TRIED TWO DIFFERENT BOLTS BUT IT JUST KEPT SPINNING AND WOULD NOT THREAD INTO GLENOSPHERE. THEY WERE ABLE TO PLACE IT IN BY HAND AND TIGHTEN IT WITH THE SCREW DRIVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25664003 ↗
  • MalfunctionJun 19, 2026Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE THREADS WERE MESSED UP. PIECES WOULD NOT SCREW TOGETHER, POSSIBLY CROSS THREADED. THE ISSUE WAS CAUGHT BEFORE THE CASE STARTED SO THERE WERE NO DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565727 ↗
  • MalfunctionJun 18, 2026Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT THE DEVICE WAS CROSS THREADED AND DAMAGED PIECES WERE OBSERVED. IT WAS REPORTED THAT THE EVENT DID NOT OCCUR IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25555681 ↗
  • MalfunctionJun 17, 2026Material Deformation

    IT WAS REPORTED THAT THE INSTRUMENT HAD AN UNKNOWN ALLEGATION. THE EVENT DID NOT HAPPEN IN SURGERY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE TWO PIECES WOULD NOT THREAD TOGETHER CROSSTHREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25535997 ↗
  • MalfunctionJun 11, 2026Material Deformation

    IT WAS REPORTED THAT THE GLENOSPHERE INSERTER TIPS WERE STRIPPED. THE EVENT DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25473835 ↗
  • MalfunctionJun 11, 2026Material Deformation

    IT WAS REPORTED THAT THE GLENOSPHERE INSERTER TIPS WERE STRIPPED. THE EVENT DID NOT HAPPEN DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25473692 ↗
  • MalfunctionJun 10, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A LEFT SHOULDER REPAIR PROCEDURE, WHEN IMPLANTING THE GLENOSPHERE COMPONENT, IT WAS OBSERVED THAT THE GLENOSPHERE INSERTER TIP BROKE OFF IN THE GLENOSPHERE IMPLANT. IT WAS REPORTED THAT THE SURGEON REMOVED THE GLENOSPHERE AND PUT A NEW GLENOSPHERE ON AND USED A NEW TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25456705 ↗
  • MalfunctionJun 10, 2026Material Deformation

    IT WAS REPORTED6 THAT DURING A SHOULDER REPAIR PROCEDURE IT WAS OBSERVED THAT THE GLENOSPHERE INSERTER TIP WAS CROSS-THREADED WHEN TRYING TO THREAD THE TIP TO THE GLENOSPHERE IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25456413 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates