INHANCE 5500-11-290Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM MODULAR UNITI PLATFORM BASEPLATE LARGE Ø29mm CEMENTLESS · Class 2 · Implant

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295553205), received between Mar 11, 2025 and Mar 11, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE SURGEON COULD NOT ADVANCE THE 35MM CENTRAL SCREW INTO THE LARGE PLATFORM BASEPLATE. THE SCREW WOULD NOT ADVANCE PAST THE FACE OF THE BASEPLATE AFTER SEVERAL ATTEMPTS AND RE-DRILLING. FINALLY, A 7X15MM CENTRAL POST WAS PLACED INSTEAD OF THE CENTRAL SCREW. THE CENTRAL POST WAS ABLE TO BE FULLY SEATED. ONLY ONE CENTRAL SCREW WAS SCARRED UP JUST BELOW THE HEAD, THE ORIGINAL 35MM. THE SCARRING WAS DUE TO THE SCREW SOMEHOW INTERFERING WITH THE INNER BOSS OF THE BASEPLATE. THE SCREW WAS CLEAN OUT OF THE BOX. THERE WAS APPROXIMATELY 10 MINUTES OF SURGICAL DELAY TO DETERMINE THE COURSE OF ACTION AND EXECUTE IT WHICH WAS TO SWITCH TO A CENTRAL POST IMPLANT. THERE WAS NO PATIENT HARM. AFFECTED SIDE: RIGHT SHOULDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21578122 ↗

Other prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented devices

All prosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page