CS100 Intra-Aortic Balloon Pump, Japanese, 110V 0998-00-3013-65Recalls & reported problems
Datascope Corp. · CS100 Intra-Aortic Balloon Pump, Japanese, 110V · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datascope CS100 intra-aortic balloon pump supports the heart during critical care. The 60 reports mostly describe equipment breaking, air or gas leaks, screen noise, and pumps stopping or losing pressure on their own during use or routine checks. There is 1 death and no injuries. One open recall, started in June 2023, covers a dilator that can crack at the hub during insertion and leave a fragment behind, risking artery damage.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 60
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 60 adverse event reports that list this exact device (DI 10607567107684), received between Sep 25, 2020 and Mar 19, 2026 — classified as 1 death, 59 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
8 other recalls cover devices cleared under the same FDA submission (K031636); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2218-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break11
- Appropriate Device Problem Term/Code Not Available6
- Device Contamination with Body Fluid6
- Backflow4
- Display or Visual Feedback Problem4
- Erratic or Intermittent Display4
- Gas/Air Leak4
- Inability to Auto-Fill3
Patient problems
- No Clinical Signs, Symptoms or Conditions53
- Insufficient Information15
- No Consequences Or Impact To Patient3
- Low Blood Pressure/ Hypotension1
- No Known Impact Or Consequence To Patient1
- Shock1
Recent adverse event reports
The 25 most recent of 60 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2026Device Contamination with Body Fluid
MalfunctionMar 6, 2026Decrease in Pressure
IT WAS REPORTED THAT THE CS100 INTRA AORTIC BALLOON PUMP (IABP) EXPERIENCED AN ISSUE OF INSUFFICIENT COMPRESSOR NEGATIVE PRESSURE. THE EVENT WAS IDENTIFIED DURING OPERATION OF THE DEVICE. THERE WERE NO PATIENT HARM OR INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24530482 ↗MalfunctionDec 14, 2025Gas/Air Leak; Gas/Air Leak
MalfunctionAug 19, 2025Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED BY THE CUSTOMER THAT DURING ROUTINE INSPECTION, CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD AUTOFILL FAILURE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22837550 ↗MalfunctionAug 7, 2025No Display/Image; Failure to Power Up; No Display/Image; Failure to Power Up
IT WAS REPORTED BY THE CUSTOMER THAT BEFORE USE ON A PATIENT CS100 INTRA-AORTIC BALLOON PUMP (IABP) WHEN THE POWER WAS TURNED ON, A SCREEN DISPLAY FAILURE (NOISE) OCCURRED. NO PATIENT HARM AND INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22735319 ↗MalfunctionJun 23, 2025Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT DURING USE, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD STOPPED UNINTENTIONALLY. NO HARM OR INJURY REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22308930 ↗MalfunctionJun 5, 2025Gas/Air Leak; Infusion or Flow Problem; Gas/Air Leak; Infusion or Flow Problem
IT WAS REPORTED THAT DURING USE, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD RAPID GAS LEAK ALARM. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22148803 ↗MalfunctionMay 15, 2025Gas/Air Leak; Gas/Air Leak
MalfunctionMar 3, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED IMMEDIATELY AFTER THE START OF CLINICAL USE THAT CS100 INTRA-AORTIC BALLOON PUMP SAID OPTICAL SENSOR CALIBRATION NOT POSSIBLE. NO HARM OR INJURY REPORTED.
Patient problems: Insufficient Information
MAUDE report 21505464 ↗MalfunctionMar 3, 2025Inability to Auto-Fill; Inability to Auto-Fill
IT WAS REPORTED BY THE CUSTOMER THAT THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) HAD AUTO-FILL ERROR DURING USER CHECKING. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21503086 ↗MalfunctionJan 20, 2025Failure to Charge; Charging Problem; Failure to Charge; Charging Problem
IT WAS REPORTED THAT THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) CANNOT BE CHARGED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21190446 ↗MalfunctionDec 13, 2024Inability to Auto-Fill; Electrical /Electronic Property Problem; Inability to Auto-Fill; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE CS100 INTRA AORTIC BALLOON PUMP (IABP) HAD FREQUENT AUTOMATIC REFILL ERRORS. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20929836 ↗MalfunctionNov 19, 2024Failure to Power Up; Failure to Power Up
MalfunctionJul 5, 2024Break; Break
IT WAS REPORTED THAT DURING CUSTOMER INSPECTION, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNIT'S TWO REAR CART CASTERS ARE BROKEN. THERE WAS NO PATIENTS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19679709 ↗MalfunctionJun 14, 2024Break; Break
MalfunctionJun 3, 2024Break; Break
MalfunctionApr 16, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) THERE WAS NO ARTERIAL WAVEFORM ON DISPLAY. OPERATED WITH ECG TRIGGER AND TRANSPORTED TO MIE UNIVERSITY HOSPITAL. TRANSFERRED TO CARDIOSAVE AT MIE UNIVERSITY HOSPITAL. THE CARDIOSAVE DISPLAYED ARTERIAL PRESSURE WAVEFORM AND FUNCTIONED NORMALLY. THERE WAS NO HEALTH HAZARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19115245 ↗MalfunctionDec 29, 2023Output Problem; Output Problem
N/A
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18419264 ↗MalfunctionDec 19, 2023Gas/Air Leak; Gas/Air Leak
MalfunctionDec 6, 2023Positioning Problem; Positioning Problem
MalfunctionNov 9, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
MalfunctionNov 2, 2023Noise, Audible; Noise, Audible
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE BY GETINGE FIELD SERVICE ENGINEER, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD ABNORMAL NOISE DUE TO FAN ABNORMALITY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18057448 ↗MalfunctionNov 1, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED DURING AN INSPECTION THE CS100 INTRA-AORTIC BALLOON PUMP (IABP)CASTER IS DAMAGE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18049361 ↗MalfunctionOct 31, 2023Break; Break
IT WAS REPORTED THAT DURING REGULAR INSPECTION, THE CS100 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD CASTER DAMAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18040103 ↗MalfunctionOct 25, 2023Break; Break
Recalls under the same FDA submission
These recalls involve devices cleared under K031636. One submission often covers many models, so check the product and lot details in each record.
- Z-1486-2026 · Jan 23, 2026 · Open, ClassifiedCS100 IABP. Software Version CS100 IABP Q.01.
- Z-3085-2017 · Jul 17, 2017 · Open, ClassifiedCS 100i Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by a
- Z-3086-2017 · Jul 17, 2017 · Open, ClassifiedCS 100 Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by au
- Z-2736-2017 · Jun 16, 2017 · TerminatedCS 100i Intra-Aortic Balloon Pump
- Z-2737-2017 · Jun 16, 2017 · TerminatedCS 100 Intra-Aortic Balloon Pump
- Z-1523-2014 · Mar 21, 2014 · TerminatedDatascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate in
- Z-2414-2011 · Mar 16, 2011 · TerminatedDatascope/ Maquet IABPs. System 98/98XT Intra-Aortic Balloon Pump; CS100/CS100i Intra-Aortic Balloon Pump; CS300 Intra-Aortic Balloon Pump; Manufactured by Datascope Corp., 130
- Z-1811-2010 · May 10, 2010 · TerminatedCS100 Intra-Aortic Balloon Pump.
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page