TRANS-RAY IAB, Intra-Aortic Balloon Catheter and… recallZ-2218-2023 · Datascope Corp. · initiated Jun 23, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
FDA root cause category: Under Investigation by firm
What the company is doing
Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Recall details
- Recall number
- Z-2218-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 23, 2023
- Posted by the FDA
- Jul 21, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 321,609 total kits
- Product code
- DSP
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
Product description
TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D884-00-0019-18 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs D884-00-0019-26
Code information (models, lots, serial numbers)
Product Description UDI 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES 10607567107656 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES 10607567107663 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES 10607567111711 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES 10607567111728 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs 10607567107684 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs 10607567111810
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES 0684-00-0607Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall1 death157 reports
- TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES 0684-00-0608Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall2 deaths152 reports
- CS100 Intra-Aortic Balloon Pump, Japanese, 110V 0998-00-3013-65Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall1 death60 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92527).
- Z-2216-2023 · Open, ClassifiedYAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description …
- Z-2217-2023 · Open, ClassifiedSENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2219-2023 · Open, ClassifiedMEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2220-2023 · Open, ClassifiedMEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2221-2023 · Open, ClassifiedSENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2222-2023 · Open, ClassifiedSENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2223-2023 · Open, ClassifiedLINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2224-2023 · Open, ClassifiedREINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page