Skip to content

Sensation Plus 7.5Fr. 40cc IAB W/ Accessories (W/O Statlock)(China) 0684-00-0568-05Recalls & reported problems

Datascope Corp. · Sensation Plus 7.5Fr. 40cc IAB W/ Accessories (W/O Statlock)(China) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10607567109442), received between Sep 2, 2025 and Jan 21, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 23, 2023Class IIOpen, Classified

    During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Z-2221-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem3
  • Gas/Air Leak1
  • Inability to Auto-Fill1
  • Optical Problem1
  • Restricted Flow rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2026Device Sensing Problem

    IT WAS REPORTED BY THE CUSTOMER THROUGH EMERGENCY SUPPORT PROGRAM THAT THE SENSATION PLUS 40CC HAD FIBER-OPTIC SENSOR FAILURE ALARM ON A CARDIOSAVE DURING USE. NURSE DENIED ANY KINK OR FOLD IN THE FO CABLE AND STATED THAT HE HAD ALREADY UNPLUGGED AND RE PLUGGED THE FO SENSOR CABLE ONCE BEFORE. NO HARM OR INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24128063 ↗
  • MalfunctionNov 7, 2025Restricted Flow rate; Device Sensing Problem; Restricted Flow rate; Device Sensing Problem

    THE COMPLAINT WAS RECEIVED THROUGH A DIRECT CALL FROM THE CUSTOMER TO THE EMERGENCY SUPPORT PROGRAM. NO HARM OR INJURY TO THE PATIENT WAS REPORTED. IT WAS REPORTED BY CUSTOMER THAT DURING USE THE SENSATION PLUS 40CC INTRA-AORTIC BALLOON (IAB) HAD A FIBER OPTIC SENSOR FAILURE ALARM AND AN IAB CATHETER RESTRICTION ALARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23497356 ↗
  • MalfunctionSep 2, 2025Inability to Auto-Fill; Device Sensing Problem; Gas/Air Leak; Optical Problem; Inability to Auto-Fill; Device Sensing Problem; Gas/Air Leak; Optical Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22939567 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page