Sensation Plus 7.5Fr. 40cc IAB W/ Accessories (W/O Statlock)(China) 0684-00-0568-05Recalls & reported problems
Datascope Corp. · Sensation Plus 7.5Fr. 40cc IAB W/ Accessories (W/O Statlock)(China) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10607567109442), received between Sep 2, 2025 and Jan 21, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 23, 2023Class IIOpen, Classified
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Z-2221-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Sensing Problem3
- Gas/Air Leak1
- Inability to Auto-Fill1
- Optical Problem1
- Restricted Flow rate1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2026Device Sensing Problem
IT WAS REPORTED BY THE CUSTOMER THROUGH EMERGENCY SUPPORT PROGRAM THAT THE SENSATION PLUS 40CC HAD FIBER-OPTIC SENSOR FAILURE ALARM ON A CARDIOSAVE DURING USE. NURSE DENIED ANY KINK OR FOLD IN THE FO CABLE AND STATED THAT HE HAD ALREADY UNPLUGGED AND RE PLUGGED THE FO SENSOR CABLE ONCE BEFORE. NO HARM OR INJURY TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128063 ↗MalfunctionNov 7, 2025Restricted Flow rate; Device Sensing Problem; Restricted Flow rate; Device Sensing Problem
THE COMPLAINT WAS RECEIVED THROUGH A DIRECT CALL FROM THE CUSTOMER TO THE EMERGENCY SUPPORT PROGRAM. NO HARM OR INJURY TO THE PATIENT WAS REPORTED. IT WAS REPORTED BY CUSTOMER THAT DURING USE THE SENSATION PLUS 40CC INTRA-AORTIC BALLOON (IAB) HAD A FIBER OPTIC SENSOR FAILURE ALARM AND AN IAB CATHETER RESTRICTION ALARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23497356 ↗MalfunctionSep 2, 2025Inability to Auto-Fill; Device Sensing Problem; Gas/Air Leak; Optical Problem; Inability to Auto-Fill; Device Sensing Problem; Gas/Air Leak; Optical Problem
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page