SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and… recallZ-2221-2023 · Datascope Corp. · initiated Jun 23, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
FDA root cause category: Under Investigation by firm
What the company is doing
Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Recall details
- Recall number
- Z-2221-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 23, 2023
- Posted by the FDA
- Jul 21, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 321,609 total kits
- Product code
- DSP
- FDA submissions
- K122628
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
Product description
SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES D684-00-0568-01 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) D684-00-0568-03 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE D684-00-0568-05 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) D684-00-0568-09 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY D684-00-0568-01U PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB D884-00-0019-22
Code information (models, lots, serial numbers)
Product Description UDI SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES 10607567108063 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (W/O STATLOCK) 10607567109428 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (WITH PRESSURE 10607567109442 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES (TURKEY) 10607567113166 SENSATION PLUS 7.5Fr. 40cc WITH ACCESSORIES, US ONLY 10607567109565 PACKAGED INSERTION KIT - SENSATION PLUS 7.5 Fr. 40 cc IAB 10607567108612
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- SENSATION PLUS 7.5Fr. 40cc IAB & Accessories 0684-00-0568-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls10 deaths407 reports
- Sensation Plus 7.5Fr. 40cc IAB with Accessories, US Only 0684-00-0568-01UDatascope Corp. · System, Balloon, Intra-Aortic And Control1 recall24 reports
- Sensation Plus 7.5Fr. 40cc IAB W/ Accessories (W/O Statlock)(China) 0684-00-0568-05Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall3 reports
- Insertion Kit with 7.5Fr. 6” (15cm) Introducer for Use with the SENSATION PLUS 7 0884-00-0019-22Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall0 reports
- Sensation Plus 7.5Fr. 40cc IAB (W/O Statlock) 0684-00-0568-03Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall0 reports
- Sensation Plus 7.5Fr 40cc IAB with Accessories (Turkey) 0684-00-0568-09Datascope Corp. · System, Balloon, Intra-Aortic And Control1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92527).
- Z-2216-2023 · Open, ClassifiedYAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description …
- Z-2217-2023 · Open, ClassifiedSENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2218-2023 · Open, ClassifiedTRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2219-2023 · Open, ClassifiedMEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2220-2023 · Open, ClassifiedMEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2222-2023 · Open, ClassifiedSENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2223-2023 · Open, ClassifiedLINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2224-2023 · Open, ClassifiedREINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page