CARDIOSAVE BATTERY CHARGING STATION 0998-00-0802Recalls & reported problems
Datascope Corp. · CARDIOSAVE BATTERY CHARGING STATION · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10607567111964), received between Feb 25, 2021 and Aug 26, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Accessory Incompatible” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 6, 2026Class IIIOpen, Classified
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
Z-1587-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Accessory Incompatible1
- Charging Problem1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2026Charging Problem
It was reported that during installation the CARDIOSAVE Intra-Aortic Balloon Pump (IABP) left side battery charger is not charging the batteries . No patient involvement. No harm was reported. The FSE reported that there are no servicable parts for charger. Could not complete repair on this device as no service parts are available for this device. The charging station still charges batteries in right bay. The FSE reported that repair/replacement of the charging station is NOT going to take place.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50445646 ↗MalfunctionSep 27, 2023Overheating of Device; Overheating of Device
IT WAS REPORTED THAT, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNIT WAS OVERHEATING ON A PATIENT.
Patient problems: Insufficient Information
MAUDE report 17825060 ↗MalfunctionFeb 25, 2021Accessory Incompatible; Accessory Incompatible
IT WAS REPORTED THAT THE CARDIOSAVE BATTERY CHARGING STATION NEEDS TO BE RETURNED BECAUSE IS DIFFICULT FOR THE CUSTOMER TO REMOVE THE BATTERY PACK FROM ONE OF THE SLOTS. SINCE THIS IS AN ACCESSORY FOR THE IABP, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 11381772 ↗
Other System, Balloon, Intra-Aortic And Control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page