Skip to content

CARDIOSAVE BATTERY CHARGING STATION 0998-00-0802Recalls & reported problems

Datascope Corp. · CARDIOSAVE BATTERY CHARGING STATION · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10607567111964), received between Feb 25, 2021 and Aug 26, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Accessory Incompatible” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 6, 2026Class IIIOpen, Classified

    The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.

    Z-1587-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Accessory Incompatible1
  • Charging Problem1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Charging Problem

    It was reported that during installation the CARDIOSAVE Intra-Aortic Balloon Pump (IABP) left side battery charger is not charging the batteries . No patient involvement. No harm was reported. The FSE reported that there are no servicable parts for charger. Could not complete repair on this device as no service parts are available for this device. The charging station still charges batteries in right bay. The FSE reported that repair/replacement of the charging station is NOT going to take place.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50445646 ↗
  • MalfunctionSep 27, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) UNIT WAS OVERHEATING ON A PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 17825060 ↗
  • MalfunctionFeb 25, 2021Accessory Incompatible; Accessory Incompatible

    IT WAS REPORTED THAT THE CARDIOSAVE BATTERY CHARGING STATION NEEDS TO BE RETURNED BECAUSE IS DIFFICULT FOR THE CUSTOMER TO REMOVE THE BATTERY PACK FROM ONE OF THE SLOTS. SINCE THIS IS AN ACCESSORY FOR THE IABP, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 11381772 ↗

Other System, Balloon, Intra-Aortic And Control devices

All System, Balloon, Intra-Aortic And Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page