Skip to content

Battery Charging Station recallZ-1587-2026 · Datascope Corp. · initiated Feb 6, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.

FDA root cause category: Under Investigation by firm

What the company is doing

On February 6, 2026 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken by the customer: Our records indicate that you may have one or more Battery Charging Station(s) in your facility. Please forward this information to all current Battery Charging Station users within your facility. " If you are currently using one of the affected Battery Charging Stations, please follow the instructions in the Battery Charging Instructions section of this letter. " Complete and sign the attached Response Form to acknowledge that you have received and understand this notification. Return the completed form to [email protected]. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action.

Recall details

Recall number
Z-1587-2026
Status
Open, Classified
Initiated by the company
Feb 6, 2026
Posted by the FDA
Mar 19, 2026
Recalling firm
Datascope Corp.
Quantity
28 units (20 US, 8 OUS)
Product code
DSP
FDA submissions
K112372
Distribution
Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.

Product description

Battery Charging Station; Model: 0998-00-0802;

Code information (models, lots, serial numbers)

Model: 0998-00-0802; UDI: 10607567111964; Serial Numbers: 200019347EAI, 200019847EAI, 200019447EAI, 210018625EAI, 210017925EAI, 210017525EAI, 210018425EAI, 210020144EAI, 210020944EAI, 210022544EAI, 210020244EAI, 210022144EAI, 210020044EAI, 210020544EAI, 210021344EAI, 210023244EAI, 210018025EAI, 210019025EAI, 210021144EAI, 210021044EAI, 210022344EAI, 210019225EAI, 210019125EAI, 210018525EAI, 210022844EAI, 200017647EAI, 210019725EAI, 210022744EAI;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page