FORE-SIGHT ELITE® 01-06-3000Recalls & reported problems

Cas Medical Systems, Inc. · Tissue Oximeter · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10609538630009), received between Oct 22, 2024 and Jun 30, 2025 — classified as 1 death, 1 injury.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2022Open, Classified

    The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment

    Z-0834-2022 · Root cause: Software in the Use Environment · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • Low Oxygen Saturation1
  • Stroke/CVA1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING USE OF A FORESIGHT ELITE MONITOR, A 15-YEAR-OLD MALE PATIENT WAS DECLARED BRAIN DEAD FOLLOWING AN OPEN-HEART SURGERY FOR REPLACEMENT OF PULMONARY VALVE REPLACEMENT AND PFO CLOSURE. PER THE LOGS, THE FORESIGHT ELITE MONITOR DISPLAYED SIGNIFICANT DROPS IN TISSUE OXYGEN SATURATION. IT IS SUSPECTED THAT THAT THE ALARM VOLUME WAS SET TO WHISPER DURING THE CASE. THE FACILITY STATED THAT THE SURGEON WAS UNAWARE OF THE LOW TISSUE OXYGEN SATURATION (STO2) VALUES DURING THE CASE. THE FACILITY DOES NOT SUSPECT A PRODUCT MALFUNCTION.

    Patient problems: Low Oxygen Saturation

    MAUDE report 22366674 ↗
  • InjuryOct 22, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE CUSTOMER HAS BEEN HAVING SOME ISSUES WITH FORESIGHT. THE CUSTOMER STATED THAT THEY HAVE BEEN HAVING ISSUES WITH THE SENSOR STAYING ON THE PATIENT, SEVERAL TIMES THEY HAVE NOT BEEN ABLE TO GET READINGS ON PATIENTS AND THEY HAVE ALSO SAID THAT THEY HAVE HAD ISSUES WITH NUMBERS THAT DO NOT MATCH THE PATIENTS CONDITION. THE CUSTOMER MENTIONED AN INSTANCE OF A PATIENT A FEW MONTHS AGO THAT WAS SHOWING NORMAL NUMBERS ON THE FORESIGHT BUT HAD A STROKE. IT IS UNKNOWN IF THE STROKE OCCURRED IN THE OR OR THE ICU AFTER SURGERY.

    Patient problems: Stroke/CVA

    MAUDE report 20510744 ↗