FORE-SIGHT ELITE® 01-06-3000Recalls & reported problems
Cas Medical Systems, Inc. · Tissue Oximeter · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10609538630009), received between Oct 22, 2024 and Jun 30, 2025 — classified as 1 death, 1 injury.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 1, 2022Open, Classified
The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment
Z-0834-2022 · Root cause: Software in the Use Environment · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Low Oxygen Saturation1
- Stroke/CVA1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathJun 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING USE OF A FORESIGHT ELITE MONITOR, A 15-YEAR-OLD MALE PATIENT WAS DECLARED BRAIN DEAD FOLLOWING AN OPEN-HEART SURGERY FOR REPLACEMENT OF PULMONARY VALVE REPLACEMENT AND PFO CLOSURE. PER THE LOGS, THE FORESIGHT ELITE MONITOR DISPLAYED SIGNIFICANT DROPS IN TISSUE OXYGEN SATURATION. IT IS SUSPECTED THAT THAT THE ALARM VOLUME WAS SET TO WHISPER DURING THE CASE. THE FACILITY STATED THAT THE SURGEON WAS UNAWARE OF THE LOW TISSUE OXYGEN SATURATION (STO2) VALUES DURING THE CASE. THE FACILITY DOES NOT SUSPECT A PRODUCT MALFUNCTION.
Patient problems: Low Oxygen Saturation
MAUDE report 22366674 ↗InjuryOct 22, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT THE CUSTOMER HAS BEEN HAVING SOME ISSUES WITH FORESIGHT. THE CUSTOMER STATED THAT THEY HAVE BEEN HAVING ISSUES WITH THE SENSOR STAYING ON THE PATIENT, SEVERAL TIMES THEY HAVE NOT BEEN ABLE TO GET READINGS ON PATIENTS AND THEY HAVE ALSO SAID THAT THEY HAVE HAD ISSUES WITH NUMBERS THAT DO NOT MATCH THE PATIENTS CONDITION. THE CUSTOMER MENTIONED AN INSTANCE OF A PATIENT A FEW MONTHS AGO THAT WAS SHOWING NORMAL NUMBERS ON THE FORESIGHT BUT HAD A STROKE. IT IS UNKNOWN IF THE STROKE OCCURRED IN THE OR OR THE ICU AFTER SURGERY.
Patient problems: Stroke/CVA
MAUDE report 20510744 ↗
Other Oximeter, Tissue Saturation devices
- INVOS PMAC71RSCCovidien LP · Oximeter, tissue saturation1 recall67 reports
- HEMOSPHERE FORESIGHT MODULE HEMFSM10Edwards Lifesciences LLC · Oximeter, Tissue Saturation1 recall12 reports
- INVOS PMAC71RICCovidien LP · Oximeter, tissue saturation1 recall4 reports
- Invos 5100C-EUCovidien LP · Oximeter, tissue saturation22 reports
- INVOS PM7100Covidien LP · Oximeter, tissue saturation21 reports
- INVOS PM7100Covidien LP · Oximeter, tissue saturation18 reports
- INVOS 5100C-PACovidien LP · Oximeter, tissue saturation15 reports
- O3 4384Masimo Corporation · Oximeter, tissue saturation15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page