CADD 21-7061-24Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of CADD (123)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10610586020445), received between Apr 13, 2018 and Jul 2, 2025 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak4
- Device Alarm System3
- Infusion or Flow Problem2
- Obstruction of Flow2
- Air/Gas in Device1
- Complete Blockage1
- Excess Flow or Over-Infusion1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient2
- No Information2
- Pain1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DRUG LEAKED. THIS OCCURRED DURING USE IN (B)(6) 2025. THERE WAS PATIENT INVOLVEMENT AND NO REPORTED PATIENT HARM/ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22382977 ↗MalfunctionJun 16, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING A VISIT TO EXCHANGE THE CASSETTE, THERE WAS A LEAK DUE TO A CRACK IN THE TUBE, AND THE MEDICATION WAS NOT BEING INJECTED INTO THE PATIENT. THE EVENT OCCURRED DURING USE, AND THE SITUATION WAS ADDRESSED AS THE PATIENT WAS COMPLAINING OF SEVERE PAIN DURING THE VISIT.
Patient problems: Pain
MAUDE report 22211982 ↗MalfunctionMar 17, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AFTER FILLING THE CASSETTE, LIQUID LEAKAGE WAS DETECTED. IT WAS REPORTED THAT THERE WAS A DELAY IN SUPPLYING THE PATIENT, BUT NO THERAPY-RELEVANT DELAY. THE EVENT TOOK PLACE AT THE PHARMACY. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21627191 ↗MalfunctionSep 20, 2024Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT THE DEVICE HAD AN AIR IN LINE ALARM. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20268018 ↗MalfunctionMay 10, 2024Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BLOCKAGE ALARM SOUNDED AND UPON CHECKING, AND THE LINE WAS BLOCKED AND CRYSTALLIZED. THIS OCCURRED TWO SEPARATE TIMES. THIS OCCURRED DURING PATIENT USE, AND THERE WAS NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19286452 ↗MalfunctionApr 12, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED THE DISPOSABLE CAUSED THE DEVICE TO EXHIBIT AN OBSTRUCTION ALARM - WITH NO CAUSE. PER REPORTER THE ALARM SOUNDS EVEN AFTER A SUBCUTANEOUS PUNCTURE, BUT IT IS LESS LIKELY TO SOUND IF THE PUMP IS RAISED TO A HIGHER LEVEL. NO ADVERSE PATIENT EFFECTS WERE REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19101294 ↗MalfunctionJul 20, 2023Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THE DISPOSABLE EXHIBITED AN OBSTRUCTION DURING INFUSION. NO ADVERSE PATIENT EFFECTS WERE REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17367834 ↗MalfunctionDec 29, 2022Device Alarm System; Device Alarm System
IT WAS REPORTED THAT AN ERROR MESSAGE AND ALARM GO OFF. NO PATIENT INJURY WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL BY THE MANUFACTURER ON 08-DEC-2022 AND ATTACHED TO COMPLAINT OBJECT. A CASSETTE WAS CONNECTED TO AN EXTENSION HOSE. THE CASSETTE WAS FILLED WITH MORPHINE, KETAMINE, MIDAZOLAM AND SODIUM CHLORIDE. THE CASSETTE WAS FILLED AT THE USERS SIDE (B)(6) 2022 AND THE PALLIATIVE TEAM SHOULD CONNECT THE CASSETTE TO THE PUMP ON (B)(6) 2022. BY CONNECTING THE CASSETTE TO THE PUMP, ERRORS AND AN ALARM ACCRUED. THE CASSETTE WAS TRIED TO BE CONNECTED TO OTHER PUMPS, BUT WITH THE SAME RESULT. EVENTUALLY THE PUMP WAS STARTED, BUT WITH SO MANY ALARMS, THAT THEY HAD TO FILL A NEW CASSETTE THE NEXT DAY WITH A NEW BATCH INSTEAD AND THIS BATCH WORKED NORMALLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16069539 ↗MalfunctionFeb 21, 2022Obstruction of Flow; Obstruction of Flow
INFORMATION WAS RECEIVED INDICATING THAT DURING PRE-USE CHECK A PUMP USING A CADD EXTENSION SET WAS UNABLE TO PASS MEDICAL FLUID THROUGH THE PRODUCT. NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13568685 ↗MalfunctionMay 5, 2021Excess Flow or Over-Infusion; Infusion or Flow Problem; Excess Flow or Over-Infusion; Infusion or Flow Problem
INFORMATION WAS RECEIVED INDICATING THAT DURING USE OF THIS SMITHS MEDICAL CADD CASSETTE RESERVOIRS, OVER DELIVER OF MEDICAL FLUID WAS OBSERVED. IT WAS REPORTED THAT THE CUSTOMER ESTIMATED SOME AMOUNT OF MEDICAL FLUID WAS SUPPOSED TO REMAIN IN THE CASSETTE; HOWEVER, NONE WAS FOUND IN IT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11775056 ↗MalfunctionSep 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL ADMINISTRATION SET WAS LEAKING WHEN CONNECTED TO A THREE-WAY STOPCOCK AS THE CUSTOMER ATTEMPTED TO INFUSE MEDIAL FLUID (PHYSIOLOGICAL SALINE SOLUTION) USING A SYRINGE BEFORE USE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10482775 ↗MalfunctionOct 2, 2019Infusion or Flow Problem; Infusion or Flow Problem
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL FLEX TUBING WHEN USING A CADD LEGACY PUMP FOUND THAT THE TUBING WAS WET. IT WAS DISCOVERED THAT THE TUBE WAS CROOKED AND ONLY HELD AT THE END OF THE CONNECTOR. THERE WAS NO REPORTED ADVERSE EVENTS DUE TO THIS ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9145597 ↗MalfunctionAug 20, 2019Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED INDICATING THAT AFTER FIVE HOURS OF USE OF THE SMITHS MEDICAL CADD EXTENSION SET, IT WAS NOTED THAT THE CONNECTION POINT WAS LEAKING. NO ADVERSE EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 8912833 ↗MalfunctionApr 13, 2018Device Alarm System; Device Alarm System
INFORMATION WAS RECEIVED INDICATING THAT A CADD® EXTENSION SETS ATTACHED TO A SMITHS MEDICAL CADD® INFUSION PUMP WAS ALARMING. THERE WERE NO REPORTED ADVERSE EFFECTS.
Patient problems: No Information
MAUDE report 7430549 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page