CADD 21-7095-24Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of CADD (123)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10610586025273), received between Aug 28, 2019 and Oct 15, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K933390); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak3
- Insufficient Flow or Under Infusion2
Patient problems
- No Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2019Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL CADD ADMINISTRATION SET WITH AIR-ELIMINATING FILTER WAS LEAKING AT THE FILTER. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9194991 ↗MalfunctionAug 29, 2019Fluid/Blood Leak; Insufficient Flow or Under Infusion; Fluid/Blood Leak; Insufficient Flow or Under Infusion
INFORMATION WAS RECEIVED INDICATING THAT THE CADD ADMINISTRATION SET LEAKED. THE KIT WAS USE TO ADMINISTER 133.8 UG POUCH OF BLINCYTO WITH THE SOLIS VIP 2120 CADD PUMP FOR A 96-HOUR HOME DELIVERY. AFTER MORE OR LESS THAN 42 HOURS OF THE CONNECTION, THE PATIENT NOTICED THAT THERE WAS A LEAK IN THE 0.22 UM FILTER. THE INFUSION HAD TO BE PREPARED AGAIN DUE TO THE AMOUNT OF MEDICATION THAT WAS LEAKED. THERE WAS NO REPORTED INJURY TO THE PATIENT.
Patient problems: No Information
MAUDE report 8946518 ↗MalfunctionAug 28, 2019Fluid/Blood Leak; Insufficient Flow or Under Infusion; Fluid/Blood Leak; Insufficient Flow or Under Infusion
INFORMATION WAS RECEIVED INDICATING THAT THE CADD ADMINISTRATION SET LEAKED. THE KIT WAS USE TO ADMINISTER 133.8 UG POUCH OF BLINCYTO WITH THE SOLIS VIP 2120 CADD PUMP FOR A 96-HOUR HOME DELIVERY. AFTER MORE OR LESS THAN 42 HOURS OF THE CONNECTION, THE PATIENT NOTICED THAT THERE WAS A LEAK IN THE 0.22 UM FILTER. IT WAS REPORTED THAT THE LEAK DID NOT HAVE AN IMPACT ON THE TREATMENT. THERE WAS NO REPORTED INJURY TO THE PATIENT.
Patient problems: No Information
MAUDE report 8942330 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K933390. One submission often covers many models, so check the product and lot details in each record.
- Z-2759-2017 · Dec 9, 2016 · TerminatedSmiths Medical CADD Blue-Striped IV Administration Set, REF 21-7036-01, Sterilized using ethylene oxide.
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page