Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10610586025556), received between Jun 12, 2017 and Sep 10, 2022 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K021999); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Pressure Problem2
- Fluid/Blood Leak1
- Leak/Splash1
- Mechanical Jam1
- Obstruction of Flow1
- Protective Measures Problem1
- Unstable1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Edema1
- Extravasation1
- Needle Stick/Puncture1
- Sepsis1
- Tingling1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2022Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE NEEDLE WAS NOT CONTINUOUS AND COMPLETELY LOCKED. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15397567 ↗MalfunctionSep 9, 2022Pressure Problem; Pressure Problem
IT WAS REPORTED THAT WHEN THE USER INJECT THROUGH THE POWER PORT TUBING THE PRESSURE WILL RISE UP; BUT AT 300 PSI, IT WILL BE CONSIDERED PRESSURE LIMITING AND THE INJECTOR ITSELF WILL SUDDENLY DROP THE RATE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15397441 ↗MalfunctionSep 9, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT THE NEEDLE WAS SHOWING OBSTRUCTION. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15397429 ↗MalfunctionSep 9, 2022Pressure Problem; Pressure Problem
IT WAS REPORTED THAT WHEN THE USER INJECT THROUGH THE POWER PORT TUBING THE PRESSURE WILL RISE UP; BUT AT 300 PSI, IT WILL BE CONSIDERED PRESSURE LIMITING AND THE INJECTOR ITSELF WILL SUDDENLY DROP THE RATE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15397413 ↗InjuryDec 27, 2017Fluid/Blood Leak; Fluid/Blood Leak
INFORMATION WAS RECEIVED INDICATING THAT DURING A BOLUS OF MEDICATION THROUGH A SMITHS MEDICAL DELTEC® GRIPPER PLUS® SAFETY NEEDLE THE PATIENT REPORTED TINGLING AND NOTED EDEMA TO THE SIDE OF THE PERFUSION. THE INFUSION WAS STOPPED, DOCTOR NOTIFIED AND EXTRAVASATION PROTOCOL INITIATED. THE PATIENT WAS SUBSEQUENTLY ADMITTED TO THE HOSPITAL OVERNIGHT FOR OBSERVATION AND DISCHARGED HOME THE FOLLOWING DAY WITH NO ADVERSE EFFECTS.
Patient problems: Edema; Extravasation; Tingling
MAUDE report 7149806 ↗InjuryOct 5, 2017Leak/Splash; Unstable; Leak/Splash; Unstable
IT WAS ADDITIONALLY REPORTED THAT THE CURRENT STATE OF THE PATIENT WAS "WELL AND CONTROLLED". ANTIBIOTICS WERE ADMINISTERED.
Patient problems: Sepsis
MAUDE report 6918357 ↗InjuryJun 12, 2017Protective Measures Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING A CHEMOTHERAPY TREATMENT, THE SAFETY DEVICE OF A DELTEC® GRIPPER PLUS® SAFETY NEEDLE DID NOT ENGAGE AND THE NEEDLE PRICKED THE NURSE. THE NURSE WENT TO THE EMERGENCY ROOM AND ACCIDENTAL BLOOD EXPOSURE PROTOCOL HAS BEEN FOLLOWED. NO PERMANENT INJURY WAS REPORTED.
Patient problems: Needle Stick/Puncture
MAUDE report 6634846 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K021999. One submission often covers many models, so check the product and lot details in each record.
- Z-0253-2020 · Jul 15, 2019 · TerminatedPort full Access Tray Product Code: DT19330
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page