Deltec 21-2865-24Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Deltec (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10610586025693), received between May 26, 2016 and May 5, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Protective Measures Problem” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K021999); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Protective Measures Problem3
- Fail-Safe Problem2
- Fluid/Blood Leak2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
IT WAS REPORTED THAT THE RUBBER STOPPER AT THE Y SITE BROKE IN HALF WHILE IN USE WITH A PATIENT. THE DEVICE LEAKED MEDICATION, SALINE AND BLOOD ONTO THE PATIENT AND DISPOSABLE CURTAIN. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21961661 ↗MalfunctionMay 5, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PRODUCT WAS LEAKING SALINE ON THE PATIENT. THE EVENT OCCURRED DURING USE. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21961651 ↗MalfunctionOct 15, 2024Fail-Safe Problem; Fail-Safe Problem
IT WAS REPORTED THAT A PATIENT WENT HOME WITH THE PUMP AND FLUID WAS LEAKING FROM PORT. PATIENT MENTIONED IT THEN LEAKED AGAIN AFTER REINSERTION. THERE WAS PATIENT INVOLVEMENT, BUT NO HUMAN HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20453450 ↗MalfunctionOct 15, 2024Fail-Safe Problem; Fail-Safe Problem
IT WAS REPORTED THAT WHEN THEY REMOVED THE NEEDLE THEY HEARD AND FELT THE CLICK OFF THE SAFETY NEEDLE. WHEN THEY WERE PUTTING THE NEEDLE IN THE SHARPS BIN THE NEEDLE CLICKED BACK OUT AND STABBED THEM IN THE MIDDLE FINGER. THERE WAS PATIENT INVOLVEMENT, BUT NO HUMAN HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20453234 ↗MalfunctionJun 7, 2017Protective Measures Problem; Protective Measures Problem
THE EVENT WAS CONSIDERED RESOLVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6619502 ↗MalfunctionJun 3, 2016Protective Measures Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING USE THE NEEDLE WOULD NOT ENGAGE INTO THE SAFETY MECHANISM. NO ADVERSE HEALTH OUTCOME RESULTED FROM THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5697687 ↗MalfunctionMay 26, 2016Protective Measures Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING USE THE NEEDLE WOULD NOT ENGAGE INTO THE SAFETY MECHANISM. NO ADVERSE HEALTH OUTCOME RESULTED FROM THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5683781 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K021999. One submission often covers many models, so check the product and lot details in each record.
- Z-0253-2020 · Jul 15, 2019 · TerminatedPort full Access Tray Product Code: DT19330
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.